The Randox Serology I Positive Control contains the… recallZ-1772-2022 · Randox Laboratories Ltd. · initiated Aug 1, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Randox Laboratories Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient results due to the positive control O36SR testing negative (Non-Reactive) for Marker HBsAg

FDA root cause category: Process change control

What the company is doing

****UPDATE*** Randox issued Updated Urgent Medical Device Correction REC611 Rev 2 dated 10/7/22. Letter states: The Randox Serology I Positive Control SR10352 lot 036SR will test negative (Non-Reactive) for Marker HBsAg using the Beckman Coulter Dxl and Ortho Vitros methods. Randox can confirm the Positive control will produce a Reactive result on the Abbott Alinity, Siemens Atellica, Biomerieux Vidas and Roche e411 methods. Randox cannot guarantee that the control will produce a Reactive result on methods not included on this notice. The distributors and customers will be instructed to: " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to [email protected] within five working days. Randox issued Urgent Medical Correction Notification (REC611) dated August 2, 2022 to the distribution center within the USA. The distributor will then contact the customer directly. The initial contact and the follow up contact will be via email. Letter states reason for recall, health risk and action to take: Randox Serology I Positive Control SR10352 lot 036SR will test negative (Non-Reactive) for Marker HBsAg using the Beckman Coulter Dxl method. The Positive control will produce a Reactive result on the Abbott Alinity, Siemens Atellica and Biomerieux Vidas methods. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to tech [email protected] within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organization.

Recall details

Recall number
Z-1772-2022
Status
Terminated
Initiated by the company
Aug 1, 2022
Posted by the FDA
Sep 19, 2022
Recalling firm
Randox Laboratories Ltd.
Quantity
2 units
Product code
OHQ
Distribution
US Nationwide distribution in the states of WV, NC.

Product description

The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg, Anti HBc, and Anti Treponema pallidum IgG. The product is delivered as a liquid in a kit of 3 identical ready-to-use vials of 5 mL each.

Code information (models, lots, serial numbers)

GTIN: 05055273216509 Lot number: 036SR Exp Date: 28th March 2023