Skip to content

Nexxis recallZ-1854-2022 · Barco NV · initiated Jul 27, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Barco NV or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

installation issues

FDA root cause category: Labeling design

What the company is doing

Barco NV began notifying its consignees on about August 29, 2022, via email, about the device malfunction and stated that they would be providing an updated instructions for use (IFU). The firm followed with a Field Safety Action, via email on the same day, which described the issue, potential risks and a copy of the updated IFU. The letter requested the following: 1. Fill out and send back Response Form 1 to Barco Key Account Manager after receipt of this Notice. 2. Read the updated installation instructions, including the potential risk of wrong installation. 3. Inform relevant partners and ask them to evaluate if their installations must be adapted. 4. Fill out and send back Response Form 2 to Barco Key Account Manager after completion of the actions.

Recall details

Recall number
Z-1854-2022
Status
Open, Classified
Initiated by the company
Jul 27, 2022
Posted by the FDA
Sep 29, 2022
Recalling firm
Barco NV
Quantity
169 devices
Product code
DXJ
FDA submissions
K173381
Distribution
US Nationwide distribution in the states of FL, IL, NJ.

Product description

Barco Medical Network Adapter, MNA-240, part of the NexxisOR system, Model Numbers: a) K9303078A, b) K9303097A

Code information (models, lots, serial numbers)

a) K9303078A, UDI 05415334000056, all serial numbers; b) K9303097A, UDI 05415334001657, all serial numbers

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K173381 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page