V. Mueller recallZ-0377-2023 · Carefusion 2200 Inc · initiated Nov 2, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Carefusion 2200 Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
BD has determined that the Instructions For Use (IFU) provided by BD is missing certain content that is contained in the manufacturer's IFU. This missing content is related to the interface with the device power supply, as well as cleaning and maintenance instructions. Refer to the attached V. Mueller" Bipolar Forcep IFU.
FDA root cause category: Under Investigation by firm
What the company is doing
The firm issued an URGENT MEDICAL DEVICE RECALL NOTIFICATION to its consignees on 11/02/2022 by FedEx 2nd day mail. The notice explained the problem and the hazard and requested the following actions from medical professionals and hospitals: "Please Take the Following Actions: 1. Please check all inventory locations within your facility for the affected product listed in Appendix I of this recall notification. 2. The content that is missing from the BD IFU is provided in the attached V. Mueller" Bipolar Forceps IFU for your records; maintain the IFU content in the appropriate location as determined by your facility. 3. Share this recall notice with any users of the product, within your facilities, to ensure they are also aware of this recall notification. 4. Complete and return the attached Customer Response Form to the BD contact noted on the form confirming acknowledgement of the recall notification, whether or not you have any affected product, so that BD may acknowledge your receipt of this notification." "Contact Information: If you require further assistance please contact: North American Regional Complaint Center Phone: 1-844-8BD- LIFE (1-844-823-5433) Say Recall when prompted M-F 8am - 5pm CT; Email: [email protected]" Actions to be taken by the distributors: "Please Take the Following Actions: 1. Please check all inventory locations within your facility for the affected product listed in Appendix I of this recall notification. 2. The content that is missing from the BD IFU is provided in the attached V. Mueller" Bipolar Forceps IFU; attach this IFU to all affected product within your facility in order to ensure the IFU is supplied with the product to your customers. 3. Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this field action notification on BD s behalf. 4. Complete and retur
Recall details
- Recall number
- Z-0377-2023
- Status
- Open, Classified
- Initiated by the company
- Nov 2, 2022
- Posted by the FDA
- Dec 3, 2022
- Recalling firm
- Carefusion 2200 Inc
- Quantity
- 84 units
- Product code
- GEI
- FDA submissions
- K213386
- Distribution
- Nationwide, Austria, Canada, Egypt, Germany, Hong Kong, Israel, Philippines, Saudi Arabia, South Africa, South Korea, Thailand
Product description
Carefusion V. Mueller" HARDY-STYLE BAYONET BIPOLAR FORCEPS, 0.5MM TIP, INSULATED OVERALL LENGTH 8-1/2" (215MM) , REF F-1035
Code information (models, lots, serial numbers)
UDI/DI 10885403040740, ALL LOTS
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- V. Mueller F-1035Steris Corporation · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 90967).
- Z-0373-2023 · Open, ClassifiedCareFusion V. Mueller CUSHING BAYONET BIPOLAR INSULATED FORCEPS, 0.7MM TIP OVERALL LENGTH 7-1/2" (190MM) , REF F-1000
- Z-0374-2023 · Open, ClassifiedCareFusion V. Mueller SCOVILLE-GREENWOOD BAYONET BIPOLAR FORCEPS 1.5MM TIP OVERALL LENGTH 7-3/4" (195MM) , REF F-1002
- Z-0375-2023 · Open, ClassifiedCarefusion V. Mueller" SCOVILLE-GREENWOOD BAYONET BIPOLAR FORCEPS 1.5MM TIP INSULATED, OVERALL LENGTH 7-3/4" (195MM), REF F-1003
- Z-0376-2023 · Open, ClassifiedCarefusion V. Mueller HARDY BAYONET BIPOLAR FORCEPS WITH STOP, 0.5MM TIP OVERALL LENGTH 8-1/2" (215MM), REF F-1030
- Z-0378-2023 · Open, ClassifiedCarefusion V. Mueller" SEMKIN BIPOLAR FORCEPS WITH STOP, 0.7MM TIP OVERALL LENGTH 5-1/2" (140MM), REF F-1040
- Z-0379-2023 · Open, ClassifiedCarefusion V. Mueller SEMKIN BIPOLAR FORCEPS 1.5MM TIP, INSULATED OVERALL LENGTH 6" (150MM), REF F-1045
- Z-0380-2023 · Open, ClassifiedCarefusion V. Mueller GERALD BIPOLAR FORCEPS 1.0MM WIDE TIP OVERALL LENGTH 7" (17.8CM) , REF F-1110
- Z-0381-2023 · Open, ClassifiedCarefusion V. Mueller GERALD BAYONET BIPOLAR FORCEPS 1.0MM TIP, OVERALL LENGTH 7-1/2" (190MM), REF F-1111
- Z-0382-2023 · Open, ClassifiedCarefusion V. Mueller"GERALD BIPOLAR FORCEPS INSULATED, 1.0MM TIP OVERALL LENGTH 7" (180MM), REF F-1112
- Z-0383-2023 · Open, ClassifiedCarefusion V. Mueller GERALD BAYONET BIPOLAR FORCEPS 1.0MM TIP, INSULATED OVERALL LENGTH 7-1/2" (190MM), REF F-1113
- Z-0384-2023 · Open, ClassifiedCarefusion V. Mueller CUSHING BAYONET BIPOLAR FORCEPS, 0.7MM TIP, INSULATED IRRIGATING OVERALL LENGTH 7-1/2" (190MM), REF F-1300
- Z-0385-2023 · Open, ClassifiedCarefusion V. Mueller GERALD BAYONET IRRIGATING BIPOLAR FORCEPS 1.0MM TIP OVERALL LENGTH 7-1/2" (190MM), REF F-1301
- Z-0386-2023 · Open, ClassifiedCarefusion V. Mueller GERALD BAYONET BIPOLAR FORCEPS 1.0MM TIP, INSULATED IRRIGATING OVERALL LENGTH 7-1/2" (190MM), REF F-1303
- Z-0387-2023 · Open, ClassifiedCarefusion V. Mueller SCOVILLE-GREENWOOD BAYONET BIPOLAR FORCEPS 1.5MM TIPS, IRRIGATING OVERALL LENGTH 7-3/4" (19.7CM), REF F-1304
- Z-0388-2023 · Open, ClassifiedCarefusion V. Mueller SCOVILLE-GREENWOOD BAYONET IRRIGATING BIPOLAR FORCEPS INSULATED, 1.5MM TIP OVERALL LENGTH 7-3/4" (195MM), REF F-1305
- Z-0389-2023 · Open, ClassifiedCarefusion V. Mueller HARDY BAYONET IRRIGATING BIPOLAR FORCEPS INSULATED, 0.5MM TIP OVERALL LENGTH 8-1/2" (215MM), REF F-1335
- Z-0390-2023 · Open, ClassifiedCarefusion V. Mueller CUSHING BIPOLAR FORCEPS 1.5MM TIP, INSULATED OVERALL LENGTH 7" (180MM), REF F-2002
- Z-0391-2023 · Open, ClassifiedCarefusion V. Mueller" CUSHING BAYONET BIPOLAR FORCEPS, 2.0MM TIP, INSULATED OVERALL LENGTH 7-1/2" (190MM), REF F-2004
- Z-0392-2023 · Open, ClassifiedCarefusion V. Mueller JEWELERS BIPOLAR FORCEPS 0.4MM TIP, STRAIGHT OVERALL LENGTH 4" (100MM), REF F-3000
- Z-0393-2023 · Open, ClassifiedCarefusion V. Mueller Neuro/Spine JEWELERS BIPOLAR FORCEPS 0.4MM TIP, CURVED, # 7 OVERALL LENGTH 4-3/4" (120MM), REF F-3007
- Z-0394-2023 · Open, ClassifiedCarefusion V. Mueller ADSON BIPOLAR FORCEPS 1.0MM TIP OVERALL LENGTH 4-3/4" (120MM), REF F-3010
- Z-0395-2023 · Open, ClassifiedCareFusion V. Mueller ADSON BIPOLAR FORCEPS 1.0MM TIP, INSULATED OVERALL LENGTH 4-3/4" (120MM), REF F-3015
- Z-0396-2023 · Open, ClassifiedCareFusion V. Mueller TITANIUM ROUND HANDLE BAYONET BIPOLAR FORCEPS, 1.0MM TIP OVERALL LENGTH 8-1/2" (215MM), REF F-5002
- Z-0397-2023 · Open, ClassifiedCareFusion V. Mueller euro/Spine TITANIUM ROUND HANDLE BAYONET BIPOLAR FORCEPS 1.5MM TIP OVERALL LENGTH 8-1/2" (215MM), REF F-5006
- Z-0398-2023 · Open, ClassifiedCareFusion V. Mueller TITANIUM ROUND HANDLE BAYONET BIPOLAR FORCEPS, 0.7MM TIP OVERALL LENGTH 9-1/2" (240MM) , REF F-5008
- Z-0399-2023 · Open, ClassifiedCareFusion V. Mueller Neuro/Spine TITANIUM ROUND HANDLE BAYONET BIPOLAR FORCEPS, 1.0MM TIP OVERALL LENGTH 9-1/2" (240MM), REF F-5010
- Z-0400-2023 · Open, ClassifiedCareFusion V. Mueller TITANIUM ROUND HANDLE BAYONET BIPOLAR FORCEPS 1.5MM TIP OVERALL LENGTH 9-1/2" (240MM), REF F-5014
- Z-0401-2023 · Open, ClassifiedCareFusion V. Mueller TITANIUM ROUND HANDLE BAYONET BIPOLAR FORCEPS 0.7MM TIP OVERALL LENGTH 10-1/2" (265MM), REF F-5016
- Z-0402-2023 · Open, ClassifiedCareFusion V. Mueller TITANIUM ROUND HANDLE BAYONET BIPOLAR FORCEPS,1.0MM TIP OVERALL LENGTH 10-1/2" (265MM), REF F-5018
- Z-0403-2023 · Open, ClassifiedCareFusion V. Mueller TITANIUM ROUND HANDLE BAYONET BIPOLAR FORCEPS 1.5MM TIP OVERALL LENGTH 10-1/2" (265MM), REF F-5022
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page