Trackmaster recallZ-0366-2023 · Full Vision Inc · initiated Aug 31, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Full Vision Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.
FDA root cause category: Other
What the company is doing
The firm began sending URGENT MEDICAL DEVICE CORRECTION notices its consignees by email beginning 09/13/2022. The distributors were provided with a letter to distribute to their customers (end users) The notice explained the issue, risk to health, and requested the following actions be taken: "Actions to be taken by the Customer/User: A simple test can be performed to determine if the unit has the incorrect component installed on the PCB Assembly. The user should turn the device on. Then cycle the power with the power switch by turning it off, waiting 5 seconds, then turn it back on. Start the belt at 2 mph and pull the Magnetic Tether. If the assembly has the incorrect component the device should fail to stop. If the assembly has the correct component the device should stop. The assemblies that fail to stop should be replaced with the c-clip style safety tethers, provided by Full Vision to the distributors and end users. To continue using the treadmill with the Magnetic Tether assemblies until they can be replaced, you can prevent this condition from occurring by cycling the power for 30 seconds before turning the device back on. Product Correction: Full Vision has scrapped all the non-conforming components (Magnetic Safety Tether and PCB Board Assembly) from the inventory. All nonconforming components will be replaced."
Recall details
- Recall number
- Z-0366-2023
- Status
- Open, Classified
- Initiated by the company
- Aug 31, 2022
- Posted by the FDA
- Dec 3, 2022
- Recalling firm
- Full Vision Inc
- Quantity
- 4 units
- Product code
- IOL
- Distribution
- Wi, GA, NY, Mexico
Product description
FULL VISION INC. TRACKMASTER, TREADMILL TMX428CP 110V, Model #317-07928
Code information (models, lots, serial numbers)
UDI/DI 00860176000620, Serial Numbers: FVDC-7644, FVDC-7645, FVDC-7646, FVDC-7686
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Trackmaster TMX428CP 110VFull Vision, Inc. · Treadmill, Powered2 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 90995).
- Z-0364-2023 · Open, ClassifiedFULL VISION INC. TRACKMASTER, TREADMILL TMX428 110V, Model #317-07926
- Z-0365-2023 · Open, ClassifiedFULL VISION INC. TRACKMASTER,TREADMILL TMX428 220V, Model #317-07927
- Z-0367-2023 · Open, ClassifiedFULL VISION INC. TRACKMASTER, TREADMILL TMX428CP 220V, Model #317-07929
- Z-0368-2023 · Open, ClassifiedFULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST1 110V, Model #317-07926GE
- Z-0369-2023 · Open, ClassifiedFULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST2 220V, Model #317-07927GE
- Z-0370-2023 · Open, ClassifiedFULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST2 220V, Model #317-07927GE CHINA
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.