AcuTwist Acutrak Compression Screws recallZ-0769-2023 · Acumed LLC · initiated Nov 29, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Acumed LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Distribution without Pre-Market Clearance
FDA root cause category: No Marketing Application
What the company is doing
On 11/29-30/2022, Acumed LLC sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via USPS First Class with signature required to customers informing them that the Acutrak Headless Compression Screws and the Acutrak 2 Headless Compression Screw Systems were commercialized with a Pre-Market Clearance (510(k)). Customer are instructed to: Product within the scope of this correction does not need to be returned to Acumed. No actions are required as this is a compliance issue and not a patient safety issue. This notification is solely for your awareness. However, physicians are urged to notify Acumed of any adverse events or product safety issues. Additionally, adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax. " Complete and submit the report Online " Regular Mail or Fax: Download form or call 1- 800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form or submit by fax to 1-800-FDA-0178. Please complete the survey on Page 8 of this letter to indicate your receipt of this notice and to acknowledge you have read the content of this notice. Return the survey via email to [email protected]. For questions - contact Acumed Business Services via email at [email protected] or via phone at (888) 627-9957 Option 4 (US and Canada) or +001 (503) 627-9957 and choose Option 4 (outside of US and Canada)
Recall details
- Recall number
- Z-0769-2023
- Status
- Open, Classified
- Initiated by the company
- Nov 29, 2022
- Posted by the FDA
- Dec 23, 2022
- Recalling firm
- Acumed LLC
- Quantity
- 80,805 units
- Product code
- HWC
- FDA submissions
- K221333
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, and WY. The countries of AE, AR, AU, BE, BR, CA, CH, CL, CN, CO, CR, DE, DK, EC, ES, FI, FR, GB, GT, HK, IE, IL, IT, JP, KR, KW, MX, MY, PA, PE, PR, PT, QA, SA, SE, SG, TH, TR, TW, and ZA.
Product description
AcuTwist Acutrak Compression Screws Part Number/Part Description: AI-0010-S 10mm AcuTwist Acutrak Compression Screw AI-0012-S 12mm AcuTwist Acutrak Compression Screw AI-0014-S 14mm AcuTwist Acutrak Compression Screw AI-0016-S 16mm AcuTwist Acutrak Compression Screw AI-0018-S 18mm AcuTwist Acutrak Compression Screw AI-0020 20mm AcuTwist Acutrak Compression Screw AI-0020-S 20mm AcuTwist Acutrak Compression Screw AI-0022-S 22mm AcuTwist Acutrak Compression Screw AI-0024-S 24mm AcuTwist Acutrak Compression Screw AI-0026 26mm AcuTwist Acutrak Compression Screw AI-0026-S 26mm AcuTwist Acutrak Compression Screw AI-0028-S 28mm AcuTwist Acutrak Compression Screw AI-0030 30mm AcuTwist Acutrak Compression Screw AI-0030-S 30mm AcuTwist Acutrak Compression Screw
Code information (models, lots, serial numbers)
Show the full code information (49k characters)
AcuTwist Acutrak Compression Device Batch Number/Part Number/Part Description/UDI Code: 204010 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-204010000000160531 219050 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-219050000000170130 226652 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-226652000000170531 236972 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-236972000000171231 239079 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-239079000000180131 260403 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-260403000000181231 265390 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-265390000000190228 270950 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-270950000000190331 273603 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-273603000000190731 278305 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-278305000000190930 358993 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-358993000000220731 358079 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-358079150903220902 404026 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-404026170131240131 411204 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-411204170418240417 412910 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-412910170728240727 417672 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-417672170721240720 433231 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-433231180125250124 440500 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-440500180430250429 427818 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-427818180604250603 444129 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-444129181005251004 453386 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-453386181018251017 453539 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-453539181130251129 459459 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-459459190114260113 459372 AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-459372190319260318 459459A AI-0010-S 10mm AcuTwist¿ Acutrak Compression Screw 10806378054501-459459A190114260113 207543 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-207543000000160630 221773 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-221773000000170330 232103 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-232103000000170930 239090 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-239090000000180228 252663 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-252663000000180731 261805 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-261805000000190131 265385 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-265385000000190331 270951 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-270951000000190331 273604 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-273604000000190731 276600 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-276600000000190831 283677 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-283677000000191231 301351 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-301351000000200831 334669 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-334669000000211130 355218 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-355218150903220902 369265 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-369265160209230208 376423 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-376423160303230303 387265 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-387265161003231003 411897 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-411897170724240723 417673 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-417673170721240720 433232 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-433232180125250124 440501 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-440501180430250429 426684 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-426684180531250530 444012 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-444012180612250611 453387 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-453387181018251017 453540 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-453540181205251204 459460 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-459460190114260113 459373 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-459373190111260110 468010 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-468010190411260410 492287 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-492287200217270216 509415 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-509415200922270922 554972 AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-554972220207290206 554972A AI-0012-S 12mm AcuTwist¿ Acutrak Compression Screw 10806378054525-554972A220207290206 204011 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-204011000000160630 218592 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-218592000000161231 224771 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-224771000000170531 233064 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-233064000000171030 239085 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-239085000000180228 252664 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-252664000000180731 260404 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-260404000000181231 264920 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-264920000000190228 270952 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-270952000000190331 273605 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-273605000000190731 276601 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-276601000000190831 283678 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-283678000000191231 310758 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-310758000000201231 341293 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-341293000000220228 358994 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-358994000000220731 365605 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-365605151204221203 395810 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-395810161014231014 403328 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-403328170306240305 422718 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-422718180110250109 430350 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-430350171228241227 433234 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-433234180125250124 440502 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-440502180430250429 428302 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-428302180529250528 453388 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-453388181018251017 453541 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-453541181205251204 459461 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-459461190114260113 459374 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-459374190111260110 492288 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-492288200217270216 453541A AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-453541A181205251204 541180 AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-541180210907280906 541180A AI-0014-S 14mm AcuTwist¿ Acutrak Compression Screw 10806378054549-541180A210907280906 202127 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-202127000000160331 207544 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-207544000000160730 219051 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-219051000000170130 221774 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-221774000000170330 228560 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-228560000000170630 236973 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-236973000000171231 239086 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-239086000000180228 255718 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-255718000000180930 263764 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-263764130131190131 270953 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-270953000000190331 267149 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-267149000000190430 278306 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-278306130930190930 293298 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-293298000000200430 326969 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-326969000000210731 349764 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-349764000000220630 358995 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-358995000000220731 369262 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-369262160218230217 398310 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-398310170106240106 423537 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-423537180109250108 430351 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-430351171227241226 433233 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-433233180125250124 440503 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-440503180430250429 453390 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-453390181018251017 453542 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-453542181130251129 430022 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-430022171213241212 455496 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-455496181116251115 459375 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-459375190111260110 468011 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-468011190411260410 468011A AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-468011A190411260410 430022A AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-430022A171213241212 537368 AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-537368210806280805 537368A AI-0016-S 16mm AcuTwist¿ Acutrak Compression Screw 10806378054563-537368A210806280805 200695 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-200695000000160331 208829 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-208829000000160730 219175 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-219175000000170131 220153 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-220153000000170131 225031 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-225031000000170531 236002 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-236002000000171231 239089 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-239089000000180228 252665 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-252665000000180731 260405 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-260405000000181231 266045 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-266045000000190331 270954 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-270954000000190331 273606 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-273606000000190731 278307 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-278307000000190930 310759 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-310759000000201231 334670 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-334670000000211130 347466 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-347466000000220430 369263 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-369263160209230208 376424 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-376424160303230303 403329 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-403329170320240319 419309 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-419309171010241009 433235 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-433235180125250124 427819 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-427819180503250502 440504 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-440504180430250429 453391 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-453391181018251017 453416 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-453416181130251129 453415 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-453415181130251129 430023 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-430023181121251120 430023A AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-430023A181121251120 540349 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-540349210825280824 561463 AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-561463220418290417 540349A AI-0018-S 18mm AcuTwist¿ Acutrak Compression Screw 10806378054587-540349A210825280824 347469 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-347469000000 352881 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-352881000000 357435 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-357435000000 360895 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-360895000000 360896 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-360896000000 360983 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-360983000000 380559 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-380559170717 383861 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-383861000000 385662 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-385662000000 383861A AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-383861A170717 428469 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-428469180521 453559 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-453559000000 458446 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-458446181231 524631 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-524631210331 528174 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-528174210510 540216 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-540216210824 540215 AI-0020 20mm AcuTwist¿ Acutrak Compression Screw 10806378054594-540215210824 202198 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-202198000000160430 207545 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-207545000000160731 216619 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-216619000000161231 220154 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-220154000000170130 224772 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-224772000000170531 226654 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-226654000000170531 236305 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-236305000000171230 239087 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-239087000000180228 250209 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-250209000000180630 254691 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-254691000000180930 260406 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-260406000000181231 265386 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-265386000000190331 270955 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-270955000000190331 273607 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-273607000000190731 276602 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-276602000000190831 278308 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-278308000000190930 305622 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-305622000000201031 318862 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-318862000000210430 326031 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-326031000000210831 332358 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-332358000000211031 343091 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-343091000000220331 357447 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-357447000000220630 355062 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-355062150828220827 365606 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-365606151204221203 371676 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-371676160119230118 371680 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-371680160323230323 398311 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-398311170106240106 410449 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-410449170607240606 424155 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-424155180110250109 430352 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-430352171228241227 433236 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-433236180125250124 427820 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-427820180503250502 440505 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-440505180430250429 453392 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-453392181018251017 453543 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-453543181205251204 453544 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-453544181205251204 456005 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-456005190110260109 444130 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-444130181005251004 459376 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-459376190111260110 430024 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-430024171213241212 456005A AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-456005A190110260109 542922 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-542922211110281109 553927 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-553927220129290128 558561 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-558561220317290316 560526 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-560526220407290406 560526A AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-560526A220407290406 560526B AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-560526B220407290406 561464 AI-0020-S 20mm AcuTwist¿ Acutrak Compression Screw 10806378054600-561464220418290417 204013 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-204013000000160630 211408 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-211408000000160930 219176 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-219176000000170131 221775 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-221775000000170330 226660 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-226660000000170531 236003 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-236003000000171230 239091 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-239091000000180228 252667 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-252667000000180731 254692 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-254692000000180930 263577 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-263577000000190228 266046 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-266046000000190331 270956 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-270956000000190331 267150 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-267150000000190430 293299 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-293299000000200430 315879 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-315879000000210331 337999 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-337999000000220131 349765 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-349765000000220630 358996 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-358996000000220731 363002 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-363002151214221213 371678 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-371678160225230224 387262 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-387262161003231003 408710 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-408710170317240316 419310 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-419310171013241012 371678A AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-371678A171004241003 425008 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-425008171107241106 424156 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-424156180118250117 433237 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-433237180125250124 440506 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-440506180430250429 427964 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-427964180604250603 453394 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-453394181018251017 453545 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-453545190110260109 453546 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-453546190123260122 514061 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-514061201119271119 554973 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-554973220207290206 561003 AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-561003220413290412 554973A AI-0022-S 22mm AcuTwist¿ Acutrak Compression Screw 10806378054624-554973A220207290206 211409 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-211409000000160930 218426 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-218426000000161231 219177 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-219177000000161231 220155 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-220155000000170131 226661 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-226661000000170531 234909 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-234909000000171031 236306 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-236306000000171230 239080 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-239080000000180130 250210 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-250210000000180630 253343 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-253343000000180831 255719 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-255719000000180930 264921 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-264921000000190228 270957 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-270957000000190331 267151 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-267151000000190430 276603 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-276603000000190831 278309 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-278309000000190930 301352 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-301352000000200831 315880 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-315880000000210331 318863 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-318863000000210531 341295 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-341295000000220228 358997 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-358997000000220731 353311 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-353311150928220927 363003 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-363003151123221122 369261 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-369261160212230211 379200 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-379200160528230528 382492 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-382492160516230516 387263 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-387263161003231003 398312 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-398312170110240110 408712 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-408712170317240316 410450 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-410450170607240606 427550 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-427550171205241204 423538 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-423538180116250115 433238 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-433238180125250124 427816 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-427816180430250429 440507 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-440507180430250429 428303 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-428303171128250603 453395 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-453395181018251017 453547 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-453547181130251129 453548 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-453548181130251129 453549 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-453549181204251203 459377 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-459377190111260110 471402 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-471402190522260521 495516 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-495516200318270318 514062 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-514062201119271119 533287 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-533287210630280629 554974 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-554974220207290206 561004 AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-561004220413290412 554974A AI-0024-S 24mm AcuTwist¿ Acutrak Compression Screw 10806378054648-554974A220207290206 347470 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-347470000000 352882 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-352882000000 360897 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-360897000000 360898 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-360898000000 360984 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-360984000000 380560 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-380560000000 383862 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-383862000000 385663 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-385663000000 360898A AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-360898A170717 426682 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-426682180511 430694 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-430694171219 453847 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-453847181025 527545 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-527545210503 533900 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-533900210707 537102 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-537102210803 553030 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-553030220119 553029 AI-0026 26mm AcuTwist¿ Acutrak Compression Screw 10806378054655-553029220119 203701 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-203701000000160531 206877 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-206877000000160630 209573 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-209573000000160730 212145 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-212145000000160930 219178 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-219178000000170131 220156 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-220156000000170131 221456 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-221456000000170330 226662 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-226662000000170531 233544 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-233544000000171030 236004 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-236004000000171231 239088 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-239088000000180130 250211 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-250211000000180630 253341 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-253341000000180831 263578 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-263578000000190131 265387 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-265387000000190331 270958 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-270958000000190331 273608 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-273608000000190731 276604 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-276604000000190831 293300 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-293300000000200430 305628 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-305628000000200930 310760 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-310760000000201231 326032 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-326032140831210831 336740 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-336740000000220131 347467 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-347467000000220430 359121 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-359121150928220927 371677 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-371677160225230224 376426 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-376426160303230303 384039 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-384039160816230816 395811 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-395811161014231014 398313 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-398313170113240113 407672 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-407672170428240427 419868 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-419868170815240814 422412 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-422412171201241130 433239 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-433239180125250124 427822 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-427822180503250502 440508 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-440508180430250429 450230 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-450230180911250910 453551 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-453551190122260121 430026 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-430026171213241212 456532 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-456532181204251203 453550 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-453550190117260116 459378 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-459378190111260110 459379 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-459379190111260110 430026A AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-430026A171213241212 537148 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-537148210803280802 521461 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-521461210303280302 554975 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-554975220207290206 560527 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-560527220407290406 561465 AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-561465220418290417 554975A AI-0026-S 26mm AcuTwist¿ Acutrak Compression Screw 10806378054662-554975A220207290206 207546 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-207546000000160731 216726 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-216726000000161231 219350 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-219350000000170131 221457 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-221457000000170330 225032 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-225032000000170530 233546 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-233546000000171031 237466 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-237466000000171230 248000 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-248000000000180531 253342 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-253342000000180831 258537 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-258537000000181130 261807 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-261807000000190131 264922 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-264922000000190228 266047 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-266047000000190331 270959 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-270959000000190331 276605 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-276605000000190831 293301 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-293301000000200430 310761 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-310761000000201231 323680 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-323680000000210430 336732 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-336732211231211231 342858 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-342858000000220331 358998 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-358998000000220731 358080 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-358080150831220830 369260 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-369260160218230217 384040 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-384040160816230816 404025 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-404025170130240130 407503 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-407503170427240426 417674 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-417674170721240720 419311 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-419311171010241009 433240 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-433240180125250124 427817 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-427817180430250429 440509 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-440509180430250429 450234 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-450234180911250910 453552 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-453552181206251205 453553 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-453553190227260226 455497 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-455497181116251115 453554 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-453554190110260109 459462 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-459462190114260113 459380 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-459380190111260110 492289 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-492289200217270216 504701 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-504701200721270721 433240A AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-433240A180125250124 504701B AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-504701B200721270721 504701C AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-504701C200721270721 504701A AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-504701A200721270721 541177 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-541177210907280906 521462 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-521462210303280302 550865 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-550865211215281214 560528 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-560528220407290406 561005 AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-561005000000 550865A AI-0028-S 28mm AcuTwist¿ Acutrak Compression Screw 10806378054686-550865A211215281214 347471 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-347471000000 352883 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-352883000000 360899 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-360899000000 360900 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-360900000000 360985 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-360985000000 380561 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-380561000000 383863 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-383863000000 385664 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-385664000000 380561A AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-380561A170717 383863A AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-383863A170717 424032 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-424032180110 430695 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-430695171219 453420 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-453420181018 453560 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-453560181022 453561 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-453561181022 535721 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-535721210722 546945 AI-0030 30mm AcuTwist¿ Acutrak Compression Screw 10806378054693-546945211027 200696 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-200696000000160331 202571 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-202571000000160331 204014 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-204014000000160530 208830 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-208830000000160730 207547 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-207547000000160731 211279 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-211279000000160930 218280 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-218280000000170131 219179 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-219179000000170131 220157 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-220157000000170131 221458 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-221458000000170330 222722 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-222722000000170330 225033 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-225033000000170530 228561 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-228561000000170630 231970 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-231970000000170930 233547 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-233547000000171031 236974 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-236974000000171231 236005 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-236005000000171230 239092 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-239092000000180228 250212 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-250212000000180630 252668 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-252668000000180731 254693 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-254693000000180830 255720 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-255720000000180830 260407 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-260407000000181231 263579 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-263579000000190228 264923 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-264923000000190228 265389 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-265389000000190331 266048 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-266048000000190331 270960 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-270960000000190331 272210 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-272210000000190731 273609 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-273609000000190731 276606 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-276606000000190831 278310 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-278310000000190930 283679 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-283679000000191231 293302 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-293302000000200430 301353 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-301353000000200831 305629 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-305629000000201031 310762 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-310762000000201231 315881 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-315881000000210331 318864 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-318864000000210430 323681 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-323681000000210630 326033 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-326033000000210831 331751 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-331751000000210831 336738 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-336738000000220131 344083 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-344083000000220131 336744 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-336744000000220131 342859 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-342859000000220331 349766 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-349766000000220630 359123 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-359123150828220827 357434 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-357434150828220827 353312 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-353312151012221011 362027 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-362027151112221111 365607 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-365607151204221203 369259 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-369259160104230103 369258 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-369258160212230211 376427 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-376427160303230303 371679 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-371679160322230322 379201 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-379201160628230628 387264 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-387264161003231003 397029 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-397029170109240109 403330 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-403330170320240319 410451 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-410451170607240606 414901 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-414901170823240822 417005 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-417005171025241024 424157 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-424157180110250109 422413 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-422413180116250115 433241 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-433241180125250124 440510 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-440510180430250429 427959 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-427959180529250528 430028 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-430028171213250906 430029 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-430029171213250830 444131 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-444131181005251004 448366 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-448366180813250812 450235 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-450235180911250910 453418 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-453418181127251126 452526 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-452526181206251205 453419 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-453419181206251205 453556 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-453556190123260122 453557 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-453557190110260109 453558 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-453558190110260109 456533 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-456533181204251203 453555 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-453555190117260116 459381 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-459381190111260110 459383 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-459383190111260110 459382 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-459382190111260110 471403 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-471403190522260521 478950 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-478950190904260903 504702 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-504702200721270721 453419A AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-453419A181206251205 507089 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-507089200824270824 514063 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-514063201119271119 453419B AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-453419B181206251205 524654 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-524654210331280330 524654A AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-524654A210331280330 541178 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-541178210907280906 533288 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-533288210630280629 521463 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-521463210303280302 554978 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-554978220207290206 554977 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-554977220207290206 554979 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-554979220207290206 554980 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-554980220207290206 560529 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-560529220407290406 561466 AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-561466220418290417 554980A AI-0030-S 30mm AcuTwist¿ Acutrak Compression Screw 10806378054709-554980A220207290206
Devices named in this recall
14 devices in the FDA device registry whose identifier appears in the recall's code information.
- ACUMED AI-0010-SAcumed LLC · Screw, fixation, bone1 recall0 reports
- ACUMED AI-0012-SAcumed LLC · Screw, fixation, bone1 recall0 reports
- ACUMED AI-0014-SAcumed LLC · Screw, fixation, bone1 recall0 reports
- ACUMED AI-0016-SAcumed LLC · Screw, fixation, bone1 recall0 reports
- ACUMED AI-0018-SAcumed LLC · Screw, fixation, bone1 recall0 reports
- ACUMED AI-0020Acumed LLC · Screw, fixation, bone1 recall0 reports
- ACUMED AI-0020-SAcumed LLC · Screw, fixation, bone1 recall0 reports
- ACUMED AI-0022-SAcumed LLC · Screw, fixation, bone1 recall0 reports
- ACUMED AI-0024-SAcumed LLC · Screw, fixation, bone1 recall0 reports
- ACUMED AI-0026Acumed LLC · Screw, fixation, bone1 recall0 reports
- ACUMED AI-0026-SAcumed LLC · Screw, fixation, bone1 recall0 reports
- ACUMED AI-0028-SAcumed LLC · Screw, fixation, bone1 recall0 reports
- ACUMED AI-0030Acumed LLC · Screw, fixation, bone1 recall0 reports
- ACUMED AI-0030-SAcumed LLC · Screw, fixation, bone1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 90996).
- Z-0767-2023 · Open, ClassifiedPart Number/Part Description: 30-0033-S 30mm Acutrak Fusion Device 6mm Nose AM-0025-S 25.0mm Acutrak 4/5 Bone Screw AM-0030-S 30.0mm Acutrak 4/5 Bone Screw AM-0035-S 35.0…
- Z-0768-2023 · Open, ClassifiedAcutrak 2 Part Number/Part Description: 30-0021 25.0mm Acutrak 2 - 5.5 Screw 30-0021-S 25.0mm Acutrak 2 - 5.5 Screw 30-0023 30.0mm Acutrak 2 - 5.5 Screw 30-0023-S 30.0mm Acutrak 2 …
- Z-0770-2023 · Open, ClassifiedDual-Trak Clavicle Screws Part Number/Part Description: 40-0136 3.0mm x 80mm Dual-Trak Clavicle Screw 40-0136-S 3.0mm x 80mm Dual-Trak Clavicle Screw 40-0137 3.0mm x 90mm Dual-Trak…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page