BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is used… recallZ-0232-2023 · Becton Dickinson & Co. · initiated Oct 13, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton Dickinson & Co. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false resistance for clinical samples if the initial Quality Control (QC) testing of the panel batch passes, delaying treatment of the patient with the most appropriate antibiotic

FDA root cause category: Under Investigation by firm

What the company is doing

BD initiated the recall on October 13, 2022 via Urgent Medical Device Product Advisory states reason for recall, health risk and action to take: 1. Immediately inspect your inventory for the specific catalog and lot number listed above and ensure the contents of this Product Advisory are read and understood. 2. It is recommended to suspend reporting of Minocyline AST results to clinicians for the affected lot. 3. Share and post this Product Advisory letter within your facility network and forward to any customers to whom you may have distributed the product to ensure awareness. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements. BD Contact US Contact Information Areas of Support North American Regional Complaint Center Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Recall when prompted Mon Fri 8:00am and 5:00pm CT or Email: [email protected]

Recall details

Recall number
Z-0232-2023
Status
Open, Classified
Initiated by the company
Oct 13, 2022
Posted by the FDA
Nov 14, 2022
Recalling firm
Becton Dickinson & Co.
Quantity
13,900 EA
Product code
LON
FDA submissions
K070809
Distribution
US Nationwide distribution.

Product description

BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is used for the rapid identification and susceptibility of most aerobic and facultative anaerobic gram-positive bacteria of human origin with selected antimicrobial agents when used with the BD Phoenix Automated Microbiology System. Catalog (Ref) Number: 448607

Code information (models, lots, serial numbers)

UDI: (10) 1355046 (17) 230131 (30) 0025 (01) 30382904486076 Lot Number: 1355046 Exp. Date: 31JAN2023

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates