GI4000 Electrosurgical Unit - intended to deliver… recallZ-0241-2023 · Steris Corporation · initiated Oct 17, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Steris Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which could cause a full power loss. If the regulator failure were to occur during a patient procedure, a procedural delay may result

FDA root cause category: Component design/selection

What the company is doing

STERIS issued Urgent Medical Device Field Correction on 10/17/22. Letter states reason for recall, health risk and action to take: All affected Customers will be contacted by a STERIS Representative to arrange the return of their GI4000 Electrosurgical Units to complete the correction. User Action Please ensure the following steps are completed: 1. Work with STERIS Representative to arrange pick up of your GI4000 Electrosurgical Unit. 2. STERIS will repair or replace the GI4000 Electrosurgical Unit for your facility. If you have questions regarding this matter, please contact Jennifer Bailey, Product Manager at 440-358-6075 or your local STERIS Representative will provide Customer Notification Letters to all Customers who purchased affected GI4000 Electrosurgical Units. STERIS will request that all affected GI4000 Electrosurgical Units be returned to complete the correction. While the units are being reworked, Customers will be provided with a loaner unit. Cintron Medical will be changing the receiving inspection plan for the PCBA. The inspection plan will have a step to verify the orientation of the diodes.

Recall details

Recall number
Z-0241-2023
Status
Terminated
Initiated by the company
Oct 17, 2022
Posted by the FDA
Nov 16, 2022
Recalling firm
Steris Corporation
Quantity
5 units
Product code
GEI
FDA submissions
K192265
Distribution
US Nationwide distribution in the states of Arizona, Illinois, Tennessee, Virginia.

Product description

GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or without delivery of argon gas, in flexible endoscopic applications. MODEL #(s): G1110001

Code information (models, lots, serial numbers)

UDI: 00724995180256 Serial Numbers: 0032-0822-003, 0032-0622-009, 0032-0922-001, 0032-0822-001, and 0032-0082-002