Arrow AutoCAT2 Intra-Aortic Balloon Pump, AEROAUTOCAT 2… recallZ-0431-2023 · ARROW INTERNATIONAL Inc. · initiated Oct 17, 2022
Class I Open, Classified All recalls posted in December 2022
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ARROW INTERNATIONAL Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.
FDA root cause category: Pending
What the company is doing
The firm issued an URGENT MEDICAL DEVICE NOTIFICATION to its consignees on 10/17/2022 by FedEx 2-day mail. To reduce the risk of short battery run time: .Plug IABP into an AC outlet during patient use, and when not in use. .Ensure the battery is fully charged prior to transporting patients, and a fully charged backup IABP device is available. .Replace the batteries when Battery Run Time is less than 90 minutes, there is visual damage to the battery, or Battery has been in service for 3 years. . Perform Battery Load Test at least every 12 months by qualified service personnel. If there is an issue with the battery load, immediately quarantine the device and contact Teleflex Customer Service to report the issue and receive service support for the affected IABP device. If a battery load test has not been performed in the past 12 months, perform battery load test prior to transporting patients with affected IABP devices. .If the IABP device battery fails while in use, immediately connect to an AC power source to continue therapy. .If a source of AC power is not available, transfer the patient to a fully charged back-up IABP device. .If pumping cannot be restored within 15 to 30 minutes, manually inflate and deflate the IAB several times per hour to reduce the risk of thrombus formation and consider removing the balloon. Facility Actions: Send a copy of this notification to all relevant personnel in your organization. Place a copy of this Notice with all affected products. Immediately check your inventory of Arrow AutoCAT2 and Arrow AC3 Optimus IABP, whether stored or in use. Return the Acknowledgement Form to Teleflex Customer Service. Hours of operation are 8am-7pm EST. Telephone: 1-866-396-2111, FAX: 1-855-419-8507 Email: [email protected].
Recall details
- Recall number
- Z-0431-2023
- Recall class
- Class I (most serious)
- Status
- Open, Classified
- Initiated by the company
- Oct 17, 2022
- Posted by the FDA
- Dec 15, 2022
- Recalling firm
- ARROW INTERNATIONAL Inc.
- Quantity
- 81 units
- Product code
- DSP
- FDA submissions
- K060309
- Distribution
- Worldwide
Product description
Arrow AutoCAT2 Intra-Aortic Balloon Pump, AEROAUTOCAT 2 WAVE, REF IAP-0535 (IPN000329), cardiac pump
Code information (models, lots, serial numbers)
UDI/DI (01) 0 0801902 05173 8, (01) 0 0801902 05173 9, All Lot/Serial Numbers
Other devices under the same FDA submission
11 devices cleared under K060309 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ARROW IPN000320Teleflex Incorporated · System, balloon, intra-aortic and control2 deaths383 reports
- ARROW IPN000302Teleflex Incorporated · System, balloon, intra-aortic and control54 reports
- ARROW IPN000325Teleflex Incorporated · System, balloon, intra-aortic and control44 reports
- ARROW IPN000324Teleflex Incorporated · System, balloon, intra-aortic and control14 reports
- ARROW IPN000329Teleflex Incorporated · System, balloon, intra-aortic and control8 reports
- ARROW IPN000308Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000312Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000333Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000334Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000323Teleflex Incorporated · System, balloon, intra-aortic and control1 report
- ARROW IPN000328Teleflex Incorporated · System, balloon, intra-aortic and control1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 91021).
- Z-0426-2023 · Open, ClassifiedArrow AutoCAT2 Intra-Aortic Balloon Pump, AUTOCAT 2 IAPB, REF IAP-0400 (IPN000302), cardiac pump
- Z-0427-2023 · Open, ClassifiedArrow AutoCAT2 Intra-Aortic Balloon Pump, AUTOCAT2 IAPB, REFURBISHED, REF IAP-0400X (IPN000311), cardiac pump
- Z-0428-2023 · Open, ClassifiedArrow AutoCAT2 Intra-Aortic Balloon Pump, AEROAUTOCAT2, REF IAP-0435 (IPN000312), cardiac pump
- Z-0429-2023 · Open, ClassifiedArrow AutoCAT2 Intra-Aortic Balloon Pump, AUTOCAT2 WAVE, REF IAP-0500 (IPN000320), cardiac pump
- Z-0430-2023 · Open, ClassifiedArrow AutoCAT2 Intra-Aortic Balloon Pump, AUTOCAT2 WAVE REFURBISHED, REF IAP-0500X (IPN000328), cardiac pump
- Z-0432-2023 · Open, ClassifiedArrow AutoCAT2 Intra-Aortic Balloon Pump, AEROAUTOCAT2 WAVE REFURBISHED, REF IAP-0535X (IPN000329), cardiac pump
- Z-0444-2023 · Open, ClassifiedArrow AutoCAT 2 Intra-Aortic Balloon Pump, Arrow AutoCAT 2 IABP, REF IAP-0400E (IPN000305), cardiac pump
- Z-0445-2023 · Open, ClassifiedArrow AutoCAT 2 Intra-Aortic Balloon Pump, Arrow AutoCAT 2 IABP, REF IAP-0400F (IPN000306), cardiac pump
- Z-0446-2023 · Open, ClassifiedArrow AutoCAT 2 Intra-Aortic Balloon Pump, Arrow AutoCAT 2 IABP, REF IAP-0400J (IPN000308), cardiac pump
- Z-0447-2023 · Open, ClassifiedArrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500D (IPN000321), cardiac pump
- Z-0448-2023 · Open, ClassifiedArrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500E (IPN000322), cardiac pump
- Z-0449-2023 · Open, ClassifiedArrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500EX (IPN000322), cardiac pump
- Z-0450-2023 · Open, ClassifiedArrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500F (IPN000323), cardiac pump
- Z-0451-2023 · Open, ClassifiedArrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500I (IPN000324), cardiac pump
- Z-0452-2023 · Open, ClassifiedArrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500J (IPN000325), cardiac pump
- Z-0453-2023 · Open, ClassifiedArrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500NL (IPN000326), cardiac pump
- Z-0454-2023 · Open, ClassifiedArrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500SV (IPN000327), cardiac pump
- Z-0455-2023 · Open, ClassifiedArrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-05001 (IPN000333), cardiac pump
- Z-0456-2023 · Open, ClassifiedArrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500J (IPN000334), cardiac pump
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page