K-ASSAY IgA Immunoturbidimetric Assay, REF: KAI-013 recallZ-0271-2023 · Kamiya Biomedical Company, LLC · initiated Sep 28, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Kamiya Biomedical Company, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

IgA Reagent may start showing cloudiness over time, which can affect assay performance.

FDA root cause category: Under Investigation by firm

What the company is doing

On 9/28/22, recall notices were sent to customers informing them to stop using affected kit lots and to return complete and partially used kits to the recalling firm. Further, if opaque turbidity is observed, or the controls of the affected assay are not within range, do not report the test results. Distributors are asked to contact their customers and to inform them of the recall. Alternately, distributors can send the recalling firm their customer list so the recalling firm can notify the customers directly. Complete and return the customer acknowledgement. The following firm contacts were provided: (206) 575-8068 and [email protected]

Recall details

Recall number
Z-0271-2023
Status
Open, Classified
Initiated by the company
Sep 28, 2022
Posted by the FDA
Nov 21, 2022
Recalling firm
Kamiya Biomedical Company, LLC
Quantity
165
Product code
CFF
FDA submissions
K993213
Distribution
Worldwide - US Nationwide distribution in the states of FL, NC, TX, TN, PA, CA, NY, IL, MA and the country of Czech Republic.

Product description

K-ASSAY IgA Immunoturbidimetric Assay, REF: KAI-013

Code information (models, lots, serial numbers)

UDI-DI: 00816426020092, Lot/ Expiration Date: H180/ 2023-01-31, K181/ 2023-03-31, N182/ 2023-06-30, D283/ 2023-09-30