Skip to content

OCEAN recallZ-0235-2023 · Atrium Medical Corporation · initiated Oct 18, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Atrium Medical Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Warehouse experienced temperature excursions in July and August 2021. It was determined that there were no associated harms to patients.

FDA root cause category: Storage

What the company is doing

Atrium/ Getinge issued Urgent Medical Device Removal on 10/18/21 via email with Delivery and Read receipt. Letter states reason for recall, health risk and action to take: Our records indicate that McKesson has received the Atrium Ocean Wet Suction Water Seal Chest Drains from the lot listed on Page 1. " Please examine your inventory and confirm that you have no remaining Atrium Ocean Wet Suction Water Seal Chest Drain 2002-000 with the REF/LOT number listed in this notice. Please return any product that may be affected under this Field Action. " Please respond to this email using the response form on page 3 to acknowledge that you have received this notification. If you have any questions, please contact your Atrium/Getinge representative or call the Atrium/Getinge Customer Support at (603) 880-1433 (press option 1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).

Recall details

Recall number
Z-0235-2023
Status
Open, Classified
Initiated by the company
Oct 18, 2022
Posted by the FDA
Nov 15, 2022
Recalling firm
Atrium Medical Corporation
Quantity
24 units (4cases/6)
Product code
KDQ
FDA submissions
K043582
Distribution
US Nationwide distribution in the state of VA.

Product description

Atrium Ocean Wet Suction Water Seal Chest Drain- Product Code: 2002-000

Code information (models, lots, serial numbers)

UDI-DI: 20650862100017 - case / 00650862100013- each Lot Number: 472326, 473226

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

8 devices cleared under K043582 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page