VITROS recallZ-0220-2023 · Ortho-Clinical Diagnostics, Inc. · initiated Oct 13, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Calibration disk Data Release Version (DRV) 6194 to 6200 contain inappropriate calibration mathematics for VITROS Calibration Kit 9 for calibrating VITROS PHYT Slides, for all lots of slide generation 22 (GEN 22). The erroneous mathematics on the calibration disk prevents the VITROS 250/350 Chemistry System from successfully calibrating, leading to delayed results.
FDA root cause category: Under Investigation by firm
What the company is doing
Ortho Clinical Diagnostics issued Urgent Letter (CL2022-260) to VITROS 250/350 Chemistry System consignees. Letter states reason for recall, health risk and action to take: Upon receiving this communication, load Calibration Diskette DRV 6201 on your VITROS 250/350 Chemistry System and calibrate VITROS PHYT Slides, GEN 22 using VITROS Calibrator Kit 9. Complete the enclosed Confirmation of Receipt form no later than October 21, 2022. Please forward this notification if the affected product was distributed outside of your facility. Save this notification with your user documentation or post this notification by each VITROS 250/350 Chemistry System until Calibration Diskette DRV 6201 has been received and loaded by your facility. If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Ortho Care" Technical Solutions Center. If you have further questions, please contact the Ortho Care Technical Solutions Center at 1-800-421-3311
Recall details
- Recall number
- Z-0220-2023
- Status
- Open, Classified
- Initiated by the company
- Oct 13, 2022
- Posted by the FDA
- Nov 10, 2022
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Quantity
- 17,396 Units
- Product code
- DKB
- Distribution
- Worldwide Distribution: US (Nationwide) and Countries (Foreign) including: Australia Bermuda DV04 Bermuda, HM 13 Bermuda, HM08 Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Norway Portugal Russia Singapore 609917 Spain Sweden United Kingdom
Product description
VITROS Chemistry Products Calibrator Kit 9 -(GEN) 22 of VITROS PHYT Slides For in vitro diagnostic use only. Product Code: 8568040 . VITROS Chemistry Products Calibrator Kit 9 is used to calibrate VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of ACET, CRBM, DGXN, PHBR, and PHYT
Code information (models, lots, serial numbers)
UDI-DI: 10758750005031 Kit Lots/Exp. Date: 0911 24-MAY-2023; 0921 04-OCT-2023; 0942 06-JUN-2024; 0980 26-OCT-2022; 0991 01-MAR-2023 VITROS Chemistry Products PHYT Slides, GEN 22, All lots Calibration Diskette Data Release Versions (DRVs) 6194 through 6200 supporting VITROS Calibrator Kit 9.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- VITROS 8568040Ortho-Clinical Diagnostics, Inc. · CALIBRATOR, SECONDARY1 recall18 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page