regard SHOULDER W BCH CHAIR, OR00356F, Item Number 800203006 recallZ-0294-2023 · ROi CPS LLC · initiated Oct 20, 2022
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ROi CPS LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Surgical convenience kits were distributed containing 3M surgical drapes which were subsequently recalled.
FDA root cause category: Component change control
What the company is doing
ROi CPS, LLC issued a MEDICAL DEVICE CORRECTION NOTICE on 10/20/2022 by email to its consignees. The notice explained the issue, the distribution dates, and kits specific to each consignee, and requested the following: "Actions to be taken by customers: 1. CHECK: Please review your inventory to determine if the manufacturer item numbers and lot numbers listed in the chart above are in your inventory. 2. QUARANTINE: Segregate and quarantine all affected kits in your inventory. 3. LABEL: Use the attached label template (Ro; CPS Correction INC13225-3M Recall Label Template-Avery 5960) to print warning labels to be affixed to all affected kits in your inventory. The label should be applied to each affected Kit and outer product case in a prominently visible location to end-users of the kits. The location chosen should not cover any other critical product information found on the existing product labeling. The label instructs users to remove and discard the 3M Steri-Drapes when the kit is opened for use. 4. REPLY: Complete the reply form included with this notice to indicate your actions in response to this notice. a. If you do not have inventory, please return the form indicating "0" affected inventory at your facility. b. Return the reply form to [email protected] 5. NOTIFY: Please immediately notify any customers to whom you have distributed or forwarded product affected by this medical device correction notice. You may include a copy of this notice, the notice from 3M, and the attached label template with your customer notification. 6. REPORT: Report any adverse health consequences experienced with the 3M products to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher's Lane, Rockville, MD 20852-9787"
Recall details
- Recall number
- Z-0294-2023
- Status
- Terminated
- Initiated by the company
- Oct 20, 2022
- Posted by the FDA
- Nov 23, 2022
- Recalling firm
- ROi CPS LLC
- Quantity
- 170 kits
- Product code
- OJH
- Distribution
- LA, MO, NC, FL
Product description
regard SHOULDER W BCH CHAIR, OR00356F, Item Number 800203006; ortho surgery convenience kit
Code information (models, lots, serial numbers)
UDI/DI 10194717107392, Lot Numbers: 91107, exp 1/8/2024; 91879, exp 3/18/2024
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Regard™ OR00356FRoi Cps, LLC · Orthopedic Tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91069).
- Z-0283-2023 · Terminatedregard LAMINECTOMY, NU01140, Item Number 900940; neuro surgery convenience kit
- Z-0284-2023 · Terminatedregard FOOT PACK, OR00123Y, Item Number 880118; orthopedic surgery convenience kit
- Z-0285-2023 · Terminatedregard TOTAL JOINT, OR00446F, Item Number 880291; orthopedic surgery convenience kit
- Z-0286-2023 · Terminatedregard KYPHOPLASTY, OR01101, Item Number 880469; orthopedic surgery convenience kit
- Z-0287-2023 · Terminatedregard BASIC NEURO, NU00259R, Item Number 800021018; neuro surgery convenience kit
- Z-0288-2023 · Terminatedregard BASIC NEURO, NU00259S, Item Number 800021019; neuro surgery convenience kit
- Z-0289-2023 · Terminatedregard SPINE PACK, NU00260H, Item Number 800022008; ortho surgery convenience kit
- Z-0290-2023 · Terminatedregard VITRECTOMY PACK, EY00266H, Item Number 800028008; eye surgery convenience kit
- Z-0291-2023 · Terminatedregard NEURO PACK, NU00278L, Item Number 800040012; nuero surgery convenience kit
- Z-0295-2023 · Terminatedregard SPINE PACK,NU00395J, Item Number 800241010; ortho surgery convenience kit
- Z-0296-2023 · Terminatedregard HAND PACK, OR00594K, Item Number 800243011; ortho surgery convenience kit
- Z-0297-2023 · Terminatedregard ACDF, NU00698H, Item Number 800246008; neuro surgery convenience kit
- Z-0298-2023 · Terminatedregard ACDF, NU00698I, Item Number 800246009; neuro surgery convenience kit
- Z-0300-2023 · Terminatedregard RETINAL PACK, EY00935C, Item Number 800730003; EYE surgery convenience kit
- Z-0301-2023 · Terminatedregard VITRECTOMY PK, EY00938C, Item Number 800733003; EYE surgery convenience kit
- Z-0302-2023 · Terminatedregard SPINE , NU00974F, Item Number 800769006; neuro surgery convenience kit
- Z-0303-2023 · Terminatedregard CVPART1BUNDLE- MERCYSTANTHONY , CV01005H, Item Number 830099008; cardiac surgery convenience kit
- Z-0304-2023 · Terminatedregard TOTAL HIP ARTHROPLASTY A&B, OR00032W, Item Number 880028023; ortho surgery convenience kit
- Z-0305-2023 · Terminatedregard SHOULDER ARTHROSCOPY, OR00049Q, Item Number 880044017; ortho surgery convenience kit
- Z-0306-2023 · Terminatedregard EYE TRAY, EY00099AE, Item Number 880099031; eye surgery convenience kit
- Z-0307-2023 · Terminatedregard FOOT PACK - MERCY SPRINGFIELD , OR00123Z, Item Number 880118026; ortho surgery convenience kit
- Z-0308-2023 · Terminatedregard SHOULDER PK, GS00125U, Item Number 880120021; ortho surgery convenience kit
- Z-0309-2023 · Terminatedregard SHOULDER PACK, GS00125V, Item Number 880120022; ortho surgery convenience kit
- Z-0310-2023 · Terminatedregard DAVINCI PACK, GS00137W, Item Number 880132023; gastric surgery convenience kit
- Z-0311-2023 · Terminatedregard MINOR NEURO, NU00140AD, Item Number 880135030; neuro surgery convenience kit
- Z-0312-2023 · Terminatedregard NEURO TRAY, NU00160R, Item Number 880158018; neuro surgery convenience kit
- Z-0313-2023 · Terminatedregard SPINE PACK , NU00193N, Item Number 880179014; neuro surgery convenience kit
- Z-0314-2023 · Terminatedregard MINOR DAVINCI PACK, GS00349M, Item Number 880251013; general surgery convenience kit
- Z-0315-2023 · Terminatedregard CATARACT, EY00450K, Item Number 880295011; eye surgery convenience kit
- Z-0316-2023 · Terminatedregard CATARACT, EY00450L, Item Number 880295012; eye surgery convenience kit
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page