Why it was recalled

Due improperly formulated and released bulk lot of fluorescence in situ hybridization (FISH) probes resulting in false positive identification results for Staphylococcus aureus and Streptococcus spp.

FDA root cause category: Mixed-up of materials/components

What the company is doing

On 11/04/2022, the firm communicated an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that the firm has become aware of complaints concerning false positive identification results for Staphylococcus aureus and Streptococcus spp. when using the Accerlerate PhenoTest BC kits. Customer are instructed to: (1) Share this customer notification with all who need to be made aware within their organization. (2) Appropriately destroy all affected Accelerate PhenoTest BC kits by defacing all labeling and disposing in accordance with site biohazard or waste policies and procedure. (3) Complete the Recall Response Form and return it to Accelerate Diagnostics. (4) Retrospectively review runs from affected lots, specifically those runs with positive identification results for either Staphylococcus aureus or Streptococcus spp. Follow their institutional policies and processes to review prior patient results related to the described issue. (5) Continue to report all technical issues through Accelerate Technical Support. Contact for Questions: NA Customer Service & Technical Support Toll free: +1 (888) 586 2939 opt. 4 Direct: +1 (520) 365 3100 Email: [email protected]

Recall details

Recall number
Z-0345-2023
Status
Open, Classified
Initiated by the company
Oct 17, 2022
Posted by the FDA
Dec 6, 2022
Recalling firm
Accelerate Diagnostics Inc
Quantity
200 kits
Product code
PRH
FDA submissions
K192665
Distribution
U.S.: AR, CA, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, MO, MS, MT, NC, ND, NJ, NY, OH, PA, RI, SC, TN, TX, VA, WA, and WV. O.U.S.: Germany, Italy, Kuwait, Latvia, Poland, Portugal, Romania, Saudi Arabia, Spain, and United Arab Emirates

Product description

Accelerate PhenoTest BC kit REF 10101018

Code information (models, lots, serial numbers)

Part Number: 10101018 UDI-DI Code: 00862011000307 Lot Numbers: 5549A 5566A 5588A 5606A 5612A

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 91075).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page