Dover recallZ-0361-2023 · Cardinal Health · initiated Nov 1, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cardinal Health or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of urinary retention
FDA root cause category: Equipment maintenance
What the company is doing
Cardinal Health issued Urgent Medical Device Recall Letter (Event-2022-04944) and Acknowledgement Form on November 1, 2022 to the direct consignees. Letter states reason for recall, health risk and action to take: Actions Required: 1. REVIEW your inventory for the affected product codes and lots. Location of product codes and lots are shown in Attachment 1. 2. SEGREGATE and quarantine all affected product upon review of your inventory. 3. NOTIFY any customers to whom you may have distributed/forwarded affected product to or will send the product on to about this product recall notice and share a copy of this notice. 4. RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to [email protected], whether you have affected product. CONTACT the appropriate Customer Service group to arrange for return and credit/replacement of any affected product Monday Friday between 8:00am - 5pm EST: " Hospital 800-965-5227 " Federal Government 800-444-1166 " Distributor 800-635-6021 " All Other Customers 888-444-5440 For questions related to this notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: [email protected] or call 800-292-9332.
Recall details
- Recall number
- Z-0361-2023
- Status
- Open, Classified
- Initiated by the company
- Nov 1, 2022
- Posted by the FDA
- Dec 2, 2022
- Recalling firm
- Cardinal Health
- Product code
- FCM
- FDA submissions
- K903340
- Distribution
- Nationwide
Product description
Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 1000 mL Drainage Bag Product Code: 3450LF Intended for urinary drainage from the bladder of a patient.
Code information (models, lots, serial numbers)
UDI-DI: Each 10884521009585 Case 20884521009582 Lot Numbers: 2216806064, 2214452064, 2212401164 2210219764, 2209525364, 2208816464 2206737964, 2206727564
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Dover 3450LFCardinal Health, Inc. · Tray, catheterization, sterile urethral, with or without catheter (kit)1 recall3 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 91087).
- Z-0356-2023 · Open, ClassifiedDover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 ml) 5ml Catheter Product Code: 2480 Intended for urinary drainage from the bladder of a patient
- Z-0357-2023 · Open, ClassifiedDover Closed Urethral Tray with Hydrogel Coated Red Rubber Catheter 16 Fr/Ch (5.3 mm) Product Code: 3408 Intended for urinary drainage from the bladder of a patient
- Z-0358-2023 · Open, ClassifiedDover Closed Urethral Tray with Hydrogel Coated Red Rubber Catheter 14 Fr/Ch (4.7 mm) Product Code: 3410 Intended for urinary drainage from the bladder of a patient
- Z-0359-2023 · Open, ClassifiedDover Closed Urethral Tray with Vinyl Catheter Product Code: 3420 Intended for urinary drainage from the bladder of a patient
- Z-0360-2023 · Open, ClassifiedDover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 1000 mL Drainage Bag Product Code: 3450 Intended for urinary drainage from the bla…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page