Skip to content

HF Resection Electrodes, Model Numbers WA22602S, WA22603S… recallZ-0424-2023 · Olympus Corporation of the Americas · initiated Nov 4, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An incompatible HF cable may be packaged with the HF Resection Electrodes. This cable cannot be connected to the electrosurgical generator and may lead to delay or cancellation of the surgical procedure.

FDA root cause category: Packaging process control

What the company is doing

On November 4, 2022, the firm notified affected customers. Customers were instructed to identify any affected product in their inventory. Affected product should be returned to Olympus, who will issue a credit to the customer upon return of affected product. Those who have distributed the devices outside their facility should notify their customers of this removal action immediately. To return product, contact your Olympus customer service representative at 1-800-848-9024.

Recall details

Recall number
Z-0424-2023
Status
Open, Classified
Initiated by the company
Nov 4, 2022
Posted by the FDA
Dec 7, 2022
Recalling firm
Olympus Corporation of the Americas
Quantity
186 boxes (930 pieces)
Product code
FJL
FDA submissions
K171965
Distribution
US Nationwide domestic distribution.

Product description

HF Resection Electrodes, Model Numbers WA22602S, WA22603S, WA22621S and WA22657S

Code information (models, lots, serial numbers)

Model Number/ UDI-DI/ Lot Number WA22602S 14042761085127 1000090824 WA22603S 14042761085134 1000092201 WA22621S 14042761085172 1000080316 WA22657S 14042761085301 1000092202

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page