2.5 mm LifePort Endotracheal Tube Adapters 10-pack… recallZ-0371-2023 · Bunnell, Inc. · initiated Nov 2, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bunnell, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Some adaptors, for use with high frequency ventilators, were packaged improperly, and may deform at the tip, which may impair the ability to pass a suction catheter or may affect normal ventilator performance, which may lead to hypercarbia, hypoxia, and gas trapping.
FDA root cause category: Packaging process control
What the company is doing
On 11/02/22, recall notices were emailed to customers, who were asked to remove the affected adapters and affected adapters from patient circuit kits and discard. The remainder of the Patient Circuit Kit may be retained for future use. As a reminder, customers may use a 3.0 mm LifePort Adapter (included in each Patient Circuit Kit) to connect to a 2.5 mm endotracheal tube if this is the only available product for patient care, per the instructions for use in the User Manual (model 204) and Operator's Manual (model 203). Complete are return the acknowledgement and receipt form. Customers with additional questions are encouraged to call 800-800-4358 or email [email protected]
Recall details
- Recall number
- Z-0371-2023
- Status
- Open, Classified
- Initiated by the company
- Nov 2, 2022
- Posted by the FDA
- Dec 3, 2022
- Recalling firm
- Bunnell, Inc.
- Quantity
- 1606 Assemblies
- Product code
- LSZ
- FDA submissions
- P850064S035
- Distribution
- US Nationwide Distribution: PA, NY, AR, AZ, TN, SC, MA, FL, CA, TX, MD, VA, NV, IL, IN, IA, HI, KY, LA, WI, GA, OH, MI, MO, MN, NJ, MS, NC, ND, NM, UT, OK, WA, RI; OUS: Canada
Product description
2.5 mm LifePort Endotracheal Tube Adapters 10-pack, Catalog: 9025; and 2.5 mm LifePort Endotracheal Tube Adaptors inside Patient Circuit Kits, Catalog: 938
Code information (models, lots, serial numbers)
Catalog/UDI/Lot: 9025/10616120000034/22042320 ; and 2.5 mm LifePort Endotracheal Tube Adaptors, Lot: 22042320, inside 938/10616120000089, 00616120000082/22052381, 22052407, 22062431, 22062458, 22062478
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Bunnell 903Bunnell, Incorporated · Ventilator, High Frequency1 recall0 reports
- LifePort 2.5 mmBunnell, Incorporated · Ventilator, High Frequency1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.