Skip to content

Southmedic SAFETY CARTRIDGE WITH SOUTHMEDIC PLUS SURGICAL… recallZ-0279-2023 · Southmedic, Inc. · initiated Oct 27, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Southmedic, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Upon review of its batch record, Southmedic identified that lot # W92761 was not sent out for sterilization, and was instead transferred directly to distribution where it was sold and distributed to customers as non-sterile product labelled as sterile.

FDA root cause category: Under Investigation by firm

What the company is doing

Southmedic sent and URGENT: MEDICAL DEVICE RECALL notice its consignees on 10/28/2022 by email. The notice explained the issue and requested the product be destroyed. The response forms indicate that the direct consignee notify its customers

Recall details

Recall number
Z-0279-2023
Status
Open, Classified
Initiated by the company
Oct 27, 2022
Posted by the FDA
Nov 23, 2022
Recalling firm
Southmedic, Inc.
Quantity
450 units
Product code
GES
Distribution
US Nationwide distribution in the state of IL.

Product description

Southmedic SAFETY CARTRIDGE WITH SOUTHMEDIC PLUS SURGICAL BLADE, REF 73-8020, STERILE EO

Code information (models, lots, serial numbers)

GTIN 620974002628, Lot number: W92761

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.