R&D CBC-3D Hematology Control recallZ-0999-2023 · R & D Systems, Inc. · initiated Nov 7, 2022
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from R & D Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
For one lot of CBC-3D Hematology Control, the Normal Level may exhibit hemolysis or deterioration due to microbial contamination.
FDA root cause category: Under Investigation by firm
What the company is doing
The recalling firm issued letters to their OEM/own-label distributor (OLD) customers on 11/7/2022 via email. The letter informed the OEM/OLD of the issue, requested they inspect the Normal Level control for signs of hemolysis or deterioration, and to discard the tubes if hemolysis or deterioration are noted. They are to sign and return the Acknowledgement and Receipt form via email by 12/2/2022. The letter informs the OEM/OLD that a new lot of CBC-3D Normal Level Hematology Control will be available 1/9/2023. The letter did not request them to notify their customers. Updated OEM/OLD letters were issued 11/14/2022 which changed the acknowledgement form to indicate the OEM/OLD has initiated a recall for the product and is notifying their customers. The consignee letters for the recalling firm's customers were issued 11/10/2022 via email. Distributor consignees who have the product drop-shipped to their end users are notifying their own end users. The letter explains the issue, and instructs them to inspect the Normal Level control for signs of hemolysis or deterioration. If signs of hemolysis or deterioration were noted, the tube is to be discarded. An Acknowledgement and Receipt form was enclosed for return to the firm by 11/22/2022 via email. The letter informs the consignee that a new lot of CBC-3D Normal Level Hematology Control will be available 1/9/2023. Also attached to the letter is a Generic Customer Letter to be sent to the distributor's end user customer along with a Customer Medical Device Important Product Notice Return Form. The Generic Customer Letter is to be sent with a cover letter from the distributor partner. On 11/21/2022, the recalling firm issued an email to one customer to notify them that catalog number 3D502 was not impacted by the recall, as it had been listed on the notification issued on 11/10/2022.
Recall details
- Recall number
- Z-0999-2023
- Status
- Terminated
- Initiated by the company
- Nov 7, 2022
- Posted by the FDA
- Jan 19, 2023
- Recalling firm
- R & D Systems, Inc.
- Quantity
- 8,259 kits, 28,136 labeled loose tubes, and 15,926 unlabeled loose tubes
- Product code
- JPK
- FDA submissions
- K843962
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, FL, GA, IA, KY, LA, MA, MO, NC, NJ, NY, OH, TX and Puerto Rico. The countries of Brazil, China, Costa Rica, Czech Republic, France, Germany, Great Britain, Honduras, Hungary, Indonesia, Israel, Japan, Kuwait, Mexico, Netherlands, Nicaragua, Panama, Peru, Poland, Romania, San Salvador, South Africa, South Korea, Switzerland, Thailand, Turkey, Vietnam,.
Product description
R&D CBC-3D Hematology Control: (1) CBC-3D Normal Pack Hematology Control, REF 3D501, containing 10 x 2.0mL Control N; (2) CBC-3D TRI PACK Hematology Control, REF 3D503, containing 4 x 2.0 mL Controls L, N, and H; (3) CBC-3D for Mythic 18 Hematology Control, REF 3D503RX, containing 2 x 2.0 mL each Controls L, N, and H; (4) CBC-3D For MINDRAY Hematology Control, REF 3D506, containing 2 x 3.0 mL each Controls L, N, and H; (5) CBC-3D For MINDRAY Hematology Control, REF 3D506US, containing 2 x 3.0 mL each Controls L, N, and H; (6) CBC-3D TRI PACK Hematology Control, REF 3D507, containing 4 x 2.0 mL each Controls L, N, and H; (7) CBC-3D TRI PACK Hematology Control, REF 3D508, containing 2 x 2.0 m each Controls L, N, and H; (8) CBC-3D For ABACUS, REF D3D02 Hematology Control, containing 2 x 3.0 mL each Controls L, N, and H; (9) CBC-3D For ABACUS Hematology Control, REF D3D04, containing 4 x 3.0 mL each Controls L, N, and H. Loose tubes labeled CBC-3D Hematology Control in 2.0mL and 3.0mL sizes. OEM/Own-label distributor labels: (1) Alere 380 Hemo Control (L-N-H), REF AL35004-KIT, containing 2 x 3.0 mL each Controls L, N, and H; Kit lot #B1122; control lot #B1122N, no UDI (2) diatron Diacon 3 Hematology Control, Normal Level, 3.0 mL tubes, lot #B1122N (3) diatron Diacon 3 Hematology Control, REF DN35002-SET, containing 6 x 3.0 mL Control N, kit lot #B1122, control lot #B1122N (4) diatron Diacon 3 Hematology Control, REF DN35004-KIT, containing 2 x 3.0 mL each Controls L, N, and H, kit lot #B1122, Normal control lot #B1122N. (5) Elite 3 HEM Control Normal, REF HEM00014, containing 6 x 3.0 mL Normal controls, kit lot #B1122, control lot #B1122N; (6) Erba H3 CON N, REF HEM00034, 3 mL vials, lot #E1122N, exp. 2023-02-10 (7) Horiba Medical Minotrol 16 Whole Blood Hematology Control, Cat. #5300000277, containing 3 x 2.5mL tubes (1 each Control 1, Control 2 (Normal), and Control 3; Cat. #5300000278, containing 6 x 2.5mL tubes (2 each Control 1, Control 2 (Normal), and Control 3; and Cat. #5300000283, containing 12 x 2.5mL tubes (4 each Control 1, Control 2 (Normal), and Control 3; kit lot #MX438, exp. 2023-01-05, Control 2 lot #MX438N. (8) 2022 3rd Test Event, CBC PLT 3 PART DIFF 2022-912-3D; and 2022 3rd Test Event, CBC PLT 3 PART DIFF 2022-914-3D. (9) mindray BC-3D, P/N 040-001423-00. (10) MINDRAY BC-3D CONTROL, REF BC3D01, containing 1 x 3.0 mL Controls L, N, and H; REF BC3D02, containing 2 x 3.0 mL Controls L, N, and H; and REF BC3D6N, containing 6 x 3.0 mL Control N; kit lot #B1122, control lot #B1122N. (11) MINDRAY BC-3D CONTROL, REF BC3D02US, containing 2 x 3.0 mL Level L, N, and H, kit lot number B1122, control lot #B1122N. (12) Nihon Kohden MEK-3D Hematology Control, REF 3DN03, containing MEK-3DN x 3 tubes; REF 3DN06, containing MEK-3DN x 6 tubes; and REF 3DN12, containing MEK-3DN x 12 tubes, kit lot #B22Y, control lot #B22YN. (13) Nihon Kohden MEK-3DN Normal Hematology Control tubes, 2.0 mL, lot #B22YN. (14) Orphee MYT-3D, REF MYT301, containing 2 x 2.5 mL Control N; REF MYT302, containing 2 x 2.5 mL each Controls L, N, and H; REF MYT304N, containing 4 x 2.5 mL Control N; REF 306N, containing 6 x 2.5 mL Control N; kit lot #B1122, control lot #B1122N; Myt-3D VET, REF MYT303-VET, containing 1 x 2.5 mL each Control L, N, and H. (15) Siemens ADVIA 360 Control, REF 11170853, containing 2 x 3.0 mL each Controls Low, Normal & High, kit lot #B0822, GTIN 00630414600833, control lot #B1122N. (16) URIT HQ-3DIFF, REF 5411000306, containing 2 x 3.0 mL each Controls L, N, and H, kit lot #B1122, control lot #B1122N; URIT HQ-3DIFF Hematology Control N tubes, 3.0 mL, lot #B1122N. (17) Woodley Haematology Control, REF WD1154, containing 6 x 2.0 mL Control N; and REF WD1154A, containing 2 x 2.0 mL each Controls L, N, and H, kit lot number B1122, control lot #B1122N;
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
R&D labeled product lot numbers: (Products 1-9): Lot B1122, exp. date 2023-02-05; UDI numbers for the following catalog numbers: 3D501 - UDI 10815762020091; 3D503 - UDI 00815762020117; 3D503RX - UDI 00815762020124; 3D506 - UDI N/A OUS only; 3D506US - UDI 00815762020162; 3D507 - UDI 00815762020179; 3D508 - UDI 00815762020186; D3D02 - UDI 00815762020230; and D3D04 - UDI 00815762020247. OEM/Own-label distributor labeled product lot numbers (Products 1-17) (1) Kit lot #B1122; control lot #B1122N. (2) Lot #B1122N. (3) Kit lot #B1122, control lot #B1122N. (4) Kit lot #B1122, control lot #B1122N. (5) Kit lot #B1122, control lot #B1122N. (6) Lot #E1122N. (7) Kit lot #MX438, Control 2 lot #MX438N. (8) Lot #2022-912-3D and Lot #2022-914-3D. (9) P/N 040-001423-00. (10) Kit lot #B1122, control lot #B1122N. (11) Kit lot number B1122, control lot #B1122N. (12) Kit lot #B22Y, control lot #B22YN. (13) Lot #B22YN. (14) Kit lot #B1122, control lot #B1122N. (15) Kit lot #B0822, GTIN 00630414600833, control lot #B1122N. (16) Kit lot #B1122, control lot #B1122N. (17) Kit lot number B1122, control lot #B1122N;
Devices named in this recall
9 devices in the FDA device registry whose identifier appears in the recall's code information.
- ADVIA® 360 Control 11170853Siemens Healthcare Diagnostics Inc. · MIXTURE, HEMATOLOGY QUALITY CONTROL1 recall0 reports
- Research and Diagnostic Systems 3D503Research And Diagnostic Systems, Inc. · Mixture, Hematology Quality Control1 recall0 reports
- Research and Diagnostic Systems 3D503RXResearch And Diagnostic Systems, Inc. · Mixture, Hematology Quality Control1 recall0 reports
- Research and Diagnostic Systems 3D506USResearch And Diagnostic Systems, Inc. · Mixture, Hematology Quality Control1 recall0 reports
- Research and Diagnostic Systems 3D507Research And Diagnostic Systems, Inc. · Mixture, Hematology Quality Control1 recall0 reports
- Research and Diagnostic Systems 3D508Research And Diagnostic Systems, Inc. · Mixture, Hematology Quality Control1 recall0 reports
- Research and Diagnostic Systems D3D02Research And Diagnostic Systems, Inc. · Mixture, Hematology Quality Control1 recall0 reports
- Research and Diagnostic Systems D3D04Research And Diagnostic Systems, Inc. · Mixture, Hematology Quality Control1 recall0 reports
- Research and Diagnostic Systems 3D501Research And Diagnostic Systems, Inc. · Mixture, Hematology Quality Control1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page