Centricity Universal Viewer 6.0 recallZ-0722-2023 · GE Medical Systems, LLC · initiated Oct 28, 2022
Class II Open, Classified All recalls posted in December 2022
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is the potential that after installation of the IAS tool, some studies can remain hidden without user notification. Access to those hidden studies requires GE Healthcare assistance, which can result in a delay in reading the study.
FDA root cause category: Software Manufacturing/Software Deployment
What the company is doing
GE Healthcare notified consignees on about 10/28/2022 via letter titled, "URGENT MEDICAL DEVICE CORRECTION - UPDATE." Consignees were instructed that the device can continued being used by following these instructions: 1) When a study is not present on the Study List, contact a GE Healthcare Service Representative for assistance in recovering the study. 2) In situations where study interpretation is time sensitive, interpret the study at the scanning unit or plan for an alternate means of reading studies for emergent cases. GE Healthcare will contact consignees to arrange for the correction. Consignees were also instructed to complete and return the acknowledgement form.
Recall details
- Recall number
- Z-0722-2023
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Oct 28, 2022
- Posted by the FDA
- Dec 16, 2022
- Recalling firm
- GE Medical Systems, LLC
- Quantity
- 236 units
- Product code
- LLZ
- FDA submissions
- K150420
- Distribution
- Nationwide distribution to CA, CO, DE, FL, IA, IL, MA, MI, MN, MO, MS, MT, ND, NJ, NY, OK, OR, SD, TX, WI, WV, WY. International distribution to Austria, Brazil, France, Germany, Ghana, Ireland, Italy, Kuwait, Poland, Saudi Arabia, South Africa, Spain, Switzerland, United Kingdom.
Product description
Centricity Universal Viewer 6.0. Used to display medical images (Including mammograms) and data from various imaging sources.
Code information (models, lots, serial numbers)
Show the full code information (6k characters)
Installed Product ID: 100666-1-Centricity Universal Viewer-14194768, 100666-1-Centricity Universal Viewer-20084199, 4405-1-Centricity Universal Viewer-00558662, 4405-1-Centricity Universal Viewer-19508102, 405-1-Centricity Universal Viewer-19050922, 4405-1-Centricity Universal Viewer-19050921, 104535-4-Centricity Universal Viewer-02396030, 104535-3-Centricity Universal Viewer-02396026, 104535-3-Centricity Universal Viewer-02396027, 104535-3-Centricity Universal Viewer-02396028, 104535-3-Centricity Universal Viewer-02396029, 104535-4-Centricity Universal Viewer-01806096, 4406-2-Centricity Universal Viewer-00558683, 4446-2-Centricity Universal Viewer-00558643, 4407-1-Centricity Universal Viewer-00558657, 4407-2-Centricity Universal Viewer-02315681, 4406-2-Centricity Universal Viewer-20501355, 4318-1-Centricity Universal Viewer-00558703, 4318-1-Centricity Universal Viewer-20080694, 10227-2-Centricity Universal Viewer-02349137, 4048-1-Centricity Universal Viewer-00558695, 4048-1-Centricity Universal Viewer-14767809, 4048-1-Centricity Universal Viewer-14767808, 4048-1-Centricity Universal Viewer-14767807, 4048-1-Centricity Universal Viewer-14767806, 4048-1-Centricity Universal Viewer-02256595, 9768-1-Centricity Universal Viewer-00558661, 9768-1-Centricity Universal Viewer-01972254, 4353-2-Centricity Universal Viewer-01589220, 13162-8-Centricity Universal Viewer-01452133, 13162-8-Centricity Universal Viewer-14776447, 13162-8-Centricity Universal Viewer-14776449, 4687-1-Centricity Universal Viewer-00558334, 100489-1-Centricity Universal Viewer-01819402, 100489-1-Centricity Universal Viewer-20791988, 104287-2-Centricity Universal Viewer-00555505, 4451-1-Centricity Universal Viewer-00558107, 100366-1-Centricity Universal Viewer-14777071, 100366-1-Centricity Universal Viewer-00558522, 4583-1-Centricity Universal Viewer-01367198, 2337-5-Centricity Universal Viewer-01810490, 2337-5-Centricity Universal Viewer-19670358, 4723-1-Centricity Universal Viewer-02303576, 3245-1-Centricity Universal Viewer-01451973, 1530-1-Centricity Universal Viewer-00558491, 1530-12-Centricity Universal Viewer-01301357, 4698-11-Centricity Universal Viewer-19667347, 4698-11-Centricity Universal Viewer-15961473, 4698-1-Centricity Universal Viewer-00558313, 100159-1-Centricity Universal Viewer-00558324, 13295-1-Centricity Universal Viewer-01292684, 13295-1-Centricity Universal Viewer-19609993, 13295-1-Centricity Universal Viewer-19611189, 101598-1-Centricity Universal Viewer-00667138, 4382-1-Centricity Universal Viewer-02268547, 102768-1-Centricity Universal Viewer-02179353, 102768-1-Centricity Universal Viewer-02255613, 102768-1-Centricity Universal Viewer-16174584, 13228-1-Centricity Universal Viewer-00558678, 4637-1-Centricity Universal Viewer-00558372, 4637-1-Centricity Universal Viewer-02176556, 9094-1-Centricity Universal Viewer-01766830, 9094-1-Centricity Universal Viewer-01977629, 4433-1-Centricity Universal Viewer-00558704, 4433-1-Centricity Universal Viewer-17737529, 100864-1-Centricity Universal Viewer-00558400, 100104-1-Centricity Universal Viewer-16204735, 100104-1-Centricity Universal Viewer-01299102, 4449-1-Centricity Universal Viewer-00558276, 4449-1-Centricity Universal Viewer-20689478, 100581-1-Centricity Universal Viewer-01814001, 1027-1-Centricity Universal Viewer-00558706, 1027-1-Centricity Universal Viewer-20759776, 101690-1-Centricity Universal Viewer-00558680, 101690-1-Centricity Universal Viewer-20091140, 4441-1-Centricity Universal Viewer-00558677, 4483-1-Centricity Universal Viewer-00558457, 4483-1-Centricity Universal Viewer-20583582, 9796-1-Centricity Universal Viewer-00558571, 13225-1-Centricity Universal Viewer-02032334, 2403-7-Centricity Universal Viewer-00558347, 2403-7-Centricity Universal Viewer-01973509, 11042-1-Centricity Universal Viewer-02285888, 11042-1-Centricity Universal Viewer-19603374, 12966-1-Centricity Universal Viewer-01283020, 12966-1-Centricity Universal Viewer-20025013, 4615-1-Centricity Universal Viewer-01452088.4615-1-Centricity Universal Viewer-20080545, 4347-1-Centricity Universal Viewer-00558520, 100773-1-Centricity Universal Viewer-00558484, 100773-1-Centricity Universal Viewer-20101378, 4487-1-Centricity Universal Viewer-01291589, 4487-1-Centricity Universal Viewer-02097167, AP6264-Centricity Universal Viewer-00000180, 313741-Centricity Universal Viewer-01422997, 313741-Centricity Universal Viewer-02097256, 313741-Centricity Universal Viewer-15591262, A91901-Centricity Universal Viewer-00044549, M41980-Centricity Universal Viewer-00044515, HC0399-Centricity Universal Viewer-01282972, CC4105-Centricity Universal Viewer-01282278, PC1695-Centricity Universal Viewer-01447701, 786787-Centricity Universal Viewer-01283171, HBON01-Centricity Universal Viewer-00260801, 691749-Centricity Universal Viewer-01539092, HC4105-Centricity Universal Viewer-00001393, 509940-Centricity Universal Viewer-01813036, 33637-Centricity Universal Viewer-15773152. HC0245-Centricity Universal Viewer-00059319 600791-Centricity Universal Viewer-02099049 485135-Centricity Universal Viewer-01447706 PC0023-Centricity Universal Viewer-01282279 509585-Centricity Universal Viewer-01826272 601860-Centricity Universal Viewer-01373156 TP0575-Centricity Universal Viewer-01825777 GH1077-Centricity Universal Viewer-01724675 81111-Centricity Universal Viewer-01286673 A59927-Centricity Universal Viewer-00040382 A61971-Centricity Universal Viewer-01891850 KW0000-Centricity Universal Viewer-01416913 PL3673-Centricity Universal Viewer-00038810 PL1494-Centricity Universal Viewer-00038799 PL2970-Centricity Universal Viewer-00038797 PL1124-Centricity Universal Viewer-00038795 PL1499-Centricity Universal Viewer-00038800 PL1501-Centricity Universal Viewer-00038802 PL2624-Centricity Universal Viewer-00038804 PL3277-Centricity Universal Viewer-00038831 PL3362-Centricity Universal Viewer-20631670 PL0326-Centricity Universal Viewer-02040892 PL2975-Centricity Universal Viewer-00162605 PL2971-Centricity Universal Viewer-00038808 SA1339-Centricity Universal Viewer-00116444 16218680-Centricity Universal Viewer-02266272 ZA2551-Centricity Universal Viewer-00116834 MPX352-Centricity Universal Viewer-19629107 DP7603-Centricity Universal Viewer-20093544 FP5028-Centricity Universal Viewer-00110911 124774-Centricity Universal Viewer-00000208 482896-Centricity Universal Viewer-00545574 85021-Centricity Universal Viewer-01426096
Other devices under the same FDA submission
1 device cleared under K150420 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page