VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm… recallZ-0728-2023 · Icotec Ag · initiated Nov 3, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Icotec Ag or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Icotec issued a Field Action Notice to its consignee on 11/03/2022 by email. The notice explained the issue and that XXXX. The firm followed with an email to inform the consignee of their intent to recall the device. The notice requested the return of the device from the field. The firm will be replacing the devices.

Recall details

Recall number
Z-0728-2023
Status
Terminated
Initiated by the company
Nov 3, 2022
Posted by the FDA
Dec 19, 2022
Recalling firm
Icotec Ag
Quantity
6 devices
Product code
NKB
FDA submissions
K190545, K193423, K200596
Distribution
US Nationwide distribution in the state of GA.

Product description

VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702

Code information (models, lots, serial numbers)

UDI/DI: (01) 07640164844589, Serial Numbers: 200221-040, 200221-041, 200221-042, 200221-043, 200221-044, 200221-045