BECKMAN COULTER IRISpec CA/CB/CC control, REF 900-7211, For… recallZ-0777-2023 · Beckman Coulter, Inc. · initiated Nov 18, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Beckman Coulter has become aware of an increase in customer complaints generated for intermittent IRISpec CB glucose control false positive results when used in conjunction with iChemVelocity Urine Chemistry Strips PN 800-7212

FDA root cause category: Under Investigation by firm

What the company is doing

The firm issued a MEDICAL DEVICE RECALL notice to its consignees on 11/18/2022 by mail and email. The notice explained the problem and requested the following actions be taken: "If IRISpec CB controls pass (QC results are within QC acceptance criteria): Continue to process patient samples per IFU (P/N 301-7146BE) If IRISpec CB controls fail (QC results are outside QC acceptance criteria): Refer to Addendum A below for troubleshooting No retrospective review of QC or patient sample results is necessary." "Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter."

Recall details

Recall number
Z-0777-2023
Status
Open, Classified
Initiated by the company
Nov 18, 2022
Posted by the FDA
Jan 3, 2023
Recalling firm
Beckman Coulter, Inc.
Quantity
3657 units
Product code
JJW
FDA submissions
K072640
Distribution
US and Canada, Brazil, Korea, Mexico, Puerto Rico, Taiwan

Product description

BECKMAN COULTER IRISpec CA/CB/CC control, REF 900-7211, For in vitro diagnostic use only.

Code information (models, lots, serial numbers)

UDI/DI 10837461002611, Lot 565-21: 11-30-2022