MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32… recallZ-0344-2023 · Mobius Imaging, LLC · initiated Nov 18, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mobius Imaging, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing.

FDA root cause category: Other

What the company is doing

On November 18, 2022, the firm notified affected customers via Urgent Medical Device Correction letters. Customers were informed that a service representative will contact their facility to schedule a time to replace the batteries in the affected devices. The customer does not need to take additional action while waiting for the battery preplacement. If you have any questions or concerns regarding this update, please contact [email protected] or call Stryker Technical Support at 978 386 9619.

Recall details

Recall number
Z-0344-2023
Status
Terminated
Initiated by the company
Nov 18, 2022
Posted by the FDA
Dec 6, 2022
Recalling firm
Mobius Imaging, LLC
Quantity
3
Product code
JAK
FDA submissions
K180393
Distribution
Domestic distribution to consignees in Colorado, Tennessee, and Utah.

Product description

MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography (CT) system, software version 2.1.1

Code information (models, lots, serial numbers)

UDI-DI 00869346000200 Serial Numbers AIRO-0124, AIRO-0145, AIRO-0288