Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model… recallZ-0886-2023 · Datascope Corp. · initiated Dec 19, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Cardiosave IABP may shut down unexpectedly due to blood entering in the Cardiosave IABP when therapy is provided with a compromised intra-aortic balloon catheter.
FDA root cause category: Device Design
What the company is doing
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 12/19/2022 was sent to customers. It contains clinical guidance to augment current clinical recommendations of patient management strategies should an IAB perforation occur. Maintain established patient and device management strategies regarding timing to IAB catheter removal. User Actions to be taken include to examine inventory immediately to determine if affected units are on hand, if a blood back event is suspected, remove the Cardiosave from patient use, and report the event to appropriate biomedical engineering staff for inspection prior to next patient use, and ensure that all Cardiosave Intra-Aortic Balloon Pump users at your facility are aware of this notice. Customers were also instructed to sign and return the MEDICAL DEVICE CORRECTION - RESPONSE FORM and notify customers if affected units were further distributed. The Cardiosave IABP units distributed going forward will be added to the scope and newly affected consignee(s) will be notified of the field correction. Letter dated 11/21/23 titled THIS IS AN UPDATE TO THE PREVIOUSLY INITIATED ACTION URGENT MEDICAL DEVICE CORRECTION was sent to customers. This notification is an update to the previously initiated action, dated December 19, 2022, regarding potential for blood to enter into the Cardiosave Hybrid and Rescue IABPs. The update is primarily to share new recommendations, such as a new labeling addendum and to consider the benefits and risks of using Cardiosave IABP on a patient during transport. Instructions were the same as the first letter. Updated letters will be sent to OUS consignees, beginning on 07/31/2025 alerting consignees of an upcoming component correction and software update to resolve the previous issues. Letters will be distributed as approval is received for each country. Once approval is received in the US, US consignees will be notified.
Recall details
- Recall number
- Z-0886-2023
- Status
- Open, Classified
- Initiated by the company
- Dec 19, 2022
- Posted by the FDA
- Jan 19, 2023
- Recalling firm
- Datascope Corp.
- Quantity
- 46 Total
- Product code
- DSP
- FDA submissions
- K181122
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Albania, Angola, Argentina, Australia, Austria, AZERBAIJAN, BAHRAIN, BANGLADESH, Barbados, Belarus, Belgium, BOSNIA and HERZEGOVINA, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, C¿te d'Ivoire, Croatia, CZECH REPUBLIC, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Ghana, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, TRINIDAD AND TOBAGO, Tunisia, Turkey, Turkmenistan, Ukraine, UNITED ARAB EMIRATES, United Kingdom, Vietnam, Yemen, and Zambia.
Product description
Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85, 0998-UC-0800-75, 0998-UC-0800-83, 0998-UC-0800-85.
Code information (models, lots, serial numbers)
All Serial Numbers. May also be designated as part number D998 etc. Model Number, UDI-DI: 0998-00-0800-75, 10607567112312; 0998-00-0800-83, 10607567108407; 0998-00-0800-85, 10607567113449; 0998-UC-0800-75, 10607567112312; 0998-UC-0800-83, 10607567108407; 0998-UC-0800-85, 10607567113449.
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE RESCUE, CHINESE 0998-00-0800-75Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls12 reports
- CARDIOSAVE RESCUE, 3.1 EDITION 0998-00-0800-85Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 91179).
- Z-0885-2023 · Open, ClassifiedCardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-36, 0998-00-0800…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page