Ingenia recallZ-0459-2023 · Philips North America Llc · initiated Nov 22, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT Medical Device Correction Notice dated 11/22/22 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users 1. Systems that do not yet have the SmokeDetector Interlock feature installed are identified in Appendix A. If your system is listed in Appendix A immediately discontinue use of your impacted MR system(s) until Philips has installed the SmokeDetector Interlock. " Post a Do not use notice on or near the impacted MR system(s) " Post this notice near the affected MR system(s) for ease of reference. " Circulate this notice to all users of this device so that they are aware of the product issue and associated hazard/harm until this issue has been resolved. " Please complete and return the attached customer response form to Philips promptly and no later than 30 days from receipt of this letter. " Note: Once your system has the SmokeDetector Interlock installed, please follow the advice in steps 2 through 5. 2. Philips has installed a SmokeDetector Interlock feature on the systems listed in Appendix B. Immediately confirm if your system has the SmokeDetector Interlock installed by confirming your serial number is listed in Appendix B. " Post this notice near the affected MR system(s) for ease of reference. " Circulate this notice to all users of this device so that they are aware of the product issue and associated hazard/harm until this issue has been resolved. " Please complete and return the attached customer response form to Philips promptly and no later than 30 days from receipt of this letter. 3. The SmokeDetector Interlock feature is integrated with the MR System and powers off the relevant part of the system (Gradient Amplifier) when smoke is detected in order to prevent further development of smoke and/or fire. If the SmokeDetector Interlock has detected smoke and has powered off the Gradient Amplifier: " Remove patient from the system according to the instructions for use because
Recall details
- Recall number
- Z-0459-2023
- Status
- Open, Classified
- Initiated by the company
- Nov 22, 2022
- Posted by the FDA
- Dec 8, 2022
- Recalling firm
- Philips North America Llc
- Quantity
- 566 total
- Product code
- LNH
- FDA submissions
- K183063, K193215, K213516, K173356, K173451
- Distribution
- US Nationwide.
Product description
Philips Ingenia Elition S MR System, Model Numbers 781357, 782106, 782137
Code information (models, lots, serial numbers)
Model 781357: UDI-DI: 00884838088108; Serial Numbers: 46002 46014 46025 46065 46077 46088 46097 46109 46071; Model 782106: UDI-DI: 00884838098329; Serial Numbers: 46224 46225 46265 46269 46311; Model Number 782137: UDI-DI: 00884838108615; Serial Numbers: Included in Letter but No Devices Distributed in the US.
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Ingenia Elition SPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging13 recalls2 reports
- Ingenia Ingenia Elition SPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls3 reports
- Ingenia Ingenia Elition SPhilips Medical Systems Nederland B.V. · System, Nuclear Magnetic Resonance Imaging7 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 91195).
- Z-0460-2023 · Open, ClassifiedPhilips Ingenia Elition X MR System, Model Numbers 781358, 782107, 782119, 782136
- Z-0462-2023 · Open, ClassifiedPhilips MR 7700 System, Model Number 782120
- Z-0463-2023 · Open, ClassifiedPhilips Upgrade to MR 7700 System, Model Number 782130
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page