Ventana Pd-L1 (Sp142) recallZ-0907-2023 · Ventana Medical Systems, Inc. · initiated Nov 18, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ventana Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent positivity, which could potentially lead to a false-negative PD-L1 result.

FDA root cause category: Process design

What the company is doing

On 11/18/2022, the firm sent an "URGENT MEDICAL DEVICE CORRECTION", Letter via UPS ground to customer to inform them that internal comparison studies revealed unacceptable, light staining with some VENTANA PD-L1 (SP142) lots. This can cause a reduction in both staining intensity and percent positivity, which could potentially lead to a false-negative PD-L1 result. Actions required of Customers: 1) Immediately discontinue the use of the VENTANA PD-L1 (SP142) lot numbers listed in Table 2 and discard any remaining inventory according to your local guidelines. 2) Be aware of the new expiration dates for the VENTANA PD-L1 (SP142) lot numbers listed in Tables 3 and 4. 3) Review negative results for the lots listed in Tables 2 and 3 as directed. 4) Follow standard laboratory operating procedures to investigate any suspected false negative results generated with the affected lots. 5) Consult with the physician or pathologist at their facility to determine specific clinical implications for their patients. 6) Refer to the Product Credit section on the last page of this UMDC for the type of credit available to them and their actions needed. 7) Complete all sections of the enclosed Recall Response Form (TP-01699) and email it to: [email protected] even if they are not requesting product credit. 8) Put PD-L1 (SP142) in the subject line of the email. Or fax it to 1-888-345-0279. 9) Be aware product will not be sent nor will credit be processed until the Recall Response Form is received. 10) Post this UMDC on or near the affected products/lots as a reminder to continuously check the expiration date and the instruction. 11) File this Urgent Medical Device Correction (UMDC) for future reference. For questions, contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-227-2155.

Recall details

Recall number
Z-0907-2023
Status
Open, Classified
Initiated by the company
Nov 18, 2022
Posted by the FDA
Jan 6, 2023
Recalling firm
Ventana Medical Systems, Inc.
Quantity
8,927 kits
Product code
PLS
FDA submissions
P160002S017
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO,CT, DC, DE, FL, GA, IL, IN, KS, KY, MA, MD, MI, MN, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, and WI. The countries of Canada, Austria, Belgium, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Luxembourg, Netherlands, Norway, Poland, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom, Albania, Armenia, Bangladesh, Belarus, Costa Rica, Cote d'Ivoire, Cuba, Dominican Republic, Egypt, El Salvador, Ghana, Guatemala, Iran, Israel, Jordan, Kenya, Kosovo, Kuwait, Lebanon, Mauritius, Morocco, Myanmar, Nigeria, North Macedonia, Oman, Panama, Philippines, Russian Fed., Saudi Arabia, Serbia, South Africa, Tanzania, Turkey, Turkmenistan and United Arab Emirates. Update (04-Jan-2023): Argentina, Australia, Brazil, Chile, China, Colombia, Ecuador, Hong Kong, Honduras, Indonesia, India, Japan, Korea, Mexico, Malaysia, New Zealand, Peru, Pakistan, Paraguay, Singapore, Thailand, Taiwan, Uruguay and Vietnam.

Product description

VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 08008540001 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 07011571001

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

UDI-DI Code: 04015630967155 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 F30596 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 G27114 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 G31438 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H03958 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H13049 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H35110 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J12769 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J18775 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J25361 UDI-DI Code: 07613336106709 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 F31427 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G06715 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G16831 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G24564 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H00001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H07164 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H15132 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H18578 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H23050 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J00572 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J01568 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J14244 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J20989 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 H08373 UDI-DI Code: 04015630976591 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 H17865 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 H30521 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 H35111 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 J11932 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 J20132