Philips Fetal Spiral Electrode-intended for patients… recallZ-0908-2023 · Philips North America Llc · initiated Nov 21, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Fetal Spiral Electrode (FSE) spiral tip of the FSE breaking off during use and requiring surgical intervention to remove the broken tip from the neonate patient. Philips has decided to discontinue the distribution of this product.
FDA root cause category: Under Investigation by firm
What the company is doing
Philips Medical issued to U.S. customers, the Urgent Medical Device Correction (UMDC) letter via certified mail 11/21/22. Philips Market Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take. Philips has decided to discontinue the distribution of this product. To limit the potential for the spiral tip of the FSE to break off in the patient s scalp, it is important to avoid over rotation during attachment or pulling the tip from the fetal scalp Customer should complete the Urgent Field Safety Notice Response Form online to submit both their acknowledgement of this recall (mandatory) and actions taken, as well as request for credit (optional). To request credit go to URL: https://forms.office.com/r/GsLVh2gYZi Pass this notice to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. (If appropriate). If you choose, discard all FSEs sold by Philips and source from an alternative supplier available in your country. If you choose to discard the FSEs, we will issue a credit after you complete the online response form. Actions planned by Philips Hospital Patient Monitoring to correct the problem " Philips is continuing to investigate additional root causes with FSE supplier " Philips will immediately discontinue selling the FSEs " Philips will reimburse for any discarded FSEs
Recall details
- Recall number
- Z-0908-2023
- Status
- Open, Classified
- Initiated by the company
- Nov 21, 2022
- Posted by the FDA
- Jan 6, 2023
- Recalling firm
- Philips North America Llc
- Quantity
- 69,679 cases (1,741,975 units)
- Product code
- HGP
- FDA submissions
- K030691
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, China, Croatia, Cura¿ao, Czechia, Denmark, Faroe Islands, Finland, France, Germany, Hong Kong, Iceland, India, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Oman, Portugal, Puerto Rico, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Tanzania, United Republic Trinidad and Tabago, United Arab Emerates, United Kingdom.
Product description
Philips Fetal Spiral Electrode-intended for patients requiring fetal heart rate monitoring during labor. PN: 989803137631
Code information (models, lots, serial numbers)
UDI: (01)20884838007431. All devices shipped to and received by the customer from Philips between May 2020 until end Nov 2022
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Fetal Measurements FETAL SPIRAL ELECTRODEPhilips Medical Systems Hsg · ELECTRODE, CIRCULAR (SPIRAL), SCALP AND APPLICATOR1 recall1 death92 reports
Other devices under the same FDA submission
1 device cleared under K030691 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Spiral Fetal Scalp Electrode CNS000004Clinical Innovations, LLC · Electrode, Circular (Spiral), Scalp And Applicator10 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.