REVACLEAR recallZ-0760-2023 · Baxter Healthcare Corporation · initiated Nov 23, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for Internal Blood Leaks due to a twisted gasket in the dialyzer

FDA root cause category: Process change control

What the company is doing

Baxter Healthcare sent a notification via USPS first class mail on about 11/23/2022 and to home patients on about 11/29/2022. Consignees were instructed to locate and remove all affected product from your inventory, contact Baxter Healthcare Center for Service to arrange for return and credit, and acknowledge the receipt of the notification by completing a reply form on the customer portal. Note: this letter describes two issues. This recall addresses Internal Blood Leaks being caused by a twisted gasket in the dialyzer. The second issue is linked to reports of Polyurethane Displacement in the dialyzer, which is addressed in a separate recall.

Recall details

Recall number
Z-0760-2023
Status
Open, Classified
Initiated by the company
Nov 23, 2022
Posted by the FDA
Dec 22, 2022
Recalling firm
Baxter Healthcare Corporation
Quantity
3,610,944 units
Product code
KDI
FDA submissions
K130039
Distribution
Worldwide distribution. US Nationwide including Guam and St. Thomas USVI; Canada, China, Australia, New Zealand, Austria, Egypt, France, Germany, Ireland, Italy, Morocco, Qatar, Saudi Arabia, South Africa, Spain, United Kingdom, Bahamas, Barbados, Belize, Bermuda, Colombia, Dominican Republic, Ecuador, Guatemala, Honduras, Mexico, Panama, and Trinidad and Tobago.

Product description

Revaclear 400 Dialyzer. Product Code: 114746L. Used to treat chronic and acute renal failure by hemodialysis.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

UDI-DI: 07332414124076; Product Code: 114746L; Lot Numbers: C421222601, C421222701, C421222801, C421222901, C421223001, C421223101, C421223201, C421223301, C421223401, C421223501, C421223601, C421223701, C421223801, C421223901, C421224001, C421224101, C421224201, C421224301, C421224401, C421224601, C421224701, C421224801, C421224901, C421225001, C421225101, C421225201, C421225301, C421225401, C421225501, C421225601, C421225701, C421225801, C421225901, C421226001, C421226101, C421226201, C421226301, C421226401, C421226501, C421226601, C421226701, C421226801, C421226901, C421227001, C421227101, C421227201, C421227301, C421227401, C421227501, C421227601, C421227701, C421227801, C421227901, C421228001, C421228101, C421228201, C421228301, C421228401, C421228501, C421228601, C421228701, C421228801, C421228901, C421229001, C421229101, C421229201, C421229301, C421229401, C421231001, C421240001, C421240101, C421240201, C421240301, C421240401, C421240501, C421240601, C421240701, C421240801, C421240901, C421241001, C421241101, C421241201, C421241301, C421241401, C421241501, C421241601, C421241701, C421241801, C421241901, C421242001, C421242101, C421242201, C421242301, C421242401, C421242501, C421242601, C421242801, C422200101, C422200201, C422200301, C422200501, C422200601, C422200701, C422200801, C422200901, C422201001, C422201101, C422201201, C422201701, C422201801, C422202101, C422202201, C422202301, C422205001, C422205101, C422205201, C422205301, C422205701, C422205801, C422205901, C422206001, C422206101, C422206201, C422206301, C422206401, C422206501, C422206601, C422206701, C422206801, C422206901, C422207001, C422207201, C422207301, C422207401, C422207501, C422207601, C422207701, C422207801, C422207901, C422208001, C422208101, C422208201, C422208301, C422208401, C422208501, C422208601, C422208701, C422208801, C422208901, C422209001, C422209101, C422209201, C422209301, C422209401, C422209501, C422209801, C422209901, C422210001, C422210101, C422210201, C422210301, C422210401, C422210501, C422210601, C422210701, C422211201, C422211301, C422211401, C422211501, C422211601, C422211701, C422211801, C422211901, C422212001, C422212101, C422212201, C422212301, C422212401, C422212501, C422212601