Skip to content

Revaclear 400 dialyzers recallZ-0763-2023 · Baxter Healthcare Corporation · initiated Nov 23, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Reports of Polyurethane Displacement in the dialyzer, and might be identified as a blood leak

FDA root cause category: Process control

What the company is doing

Baxter Healthcare notified healthcare facilities on 11/23/2022 and home patients on 11/29/2022, via first class mail. Consignees were instructed to be aware of the difference in appearances between Polyurethane Displacement and an Internal Blood Leak and ensure staff is also aware. Additionally, consignees were instructed that if there is a blood leak alarm, to follow your clinical practice to stop treatment, and if only the Polyurethane Displacement is observed, to continue use of the dialyzer. Consignees were asked to acknowledge receipt of the notification by completing and returning the response form.

Recall details

Recall number
Z-0763-2023
Status
Open, Classified
Initiated by the company
Nov 23, 2022
Posted by the FDA
Dec 22, 2022
Recalling firm
Baxter Healthcare Corporation
Quantity
6,338,280 units
Product code
KDI
FDA submissions
K130039
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Australia, New Zealand, Austria, Egypt, France, Germany, Ireland, Italy, Morocco, Qatar, Saudi Arabia, South Africa, Spain, United Kingdom, Bahamas, Barbados, Belize, Bermuda, Colombia, Dominican Republic, Ecuador, Guatemala, Honduras, Mexico, Panama, St. Thomas, Trinidad and Tobago.

Product description

Revaclear 400 dialyzers. Indicated for treatment of chronic and acute renal failure by hemodialysis.

Code information (models, lots, serial numbers)

a. UDI-DI: 07332414124076; Product Code: 114746L; All lot numbers from C421221401-C422228501. b. UDI-DI: 07332414126209; Product Code: 114746M; All lot numbers from C422218601-C422219901

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 91223).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page