MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32… recallZ-0957-2023 · Mobius Imaging, LLC · initiated Dec 7, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mobius Imaging, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Stryker received one report of an injury to an untrained user that occurred while the untrained user was moving an Airo TruCT unit in reverse. The firm will increase visibility of instructions for users to consider when transporting the unit in reverse by providing labels to be affixed onto the AIRO.

FDA root cause category: Use error

What the company is doing

On December 7, 2022, the firm notified customers via Urgent Medical Device Correction letters. Customers were informed that the firm would add additional warning labels to the Airo to increase visibility of instructions for users to consider when transporting an Airo TruCT unit in reverse. A service representative will contact your facility once labels are available (anticipated February 2023) to schedule a time to affix the labeling to your Airo TruCT unit. In the interim, please ensure all Airo TruCT users and other relevant personnel in your facilities are informed of this issue. If the devices have been distributed to other organizations, inform Stryker so that they may inform personnel at the new locations appropriately. If you have any questions or concerns regarding this notification, please contact [email protected] or call Stryker Technical Support at 978 386 9619.

Recall details

Recall number
Z-0957-2023
Status
Open, Classified
Initiated by the company
Dec 7, 2022
Posted by the FDA
Jan 17, 2023
Recalling firm
Mobius Imaging, LLC
Quantity
303
Product code
JAK
FDA submissions
K180393
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Chinese Taipei, Denmark, Egypt, France, Germany, India, Italy, Japan, Kazakhstan, Korea, Republic of Kuwait, Lithuania, Malaysia, Mexico, Morocco, Nepal, Netherlands, Norway, Pakistan, Poland, Russian Federation, Saudi Arabia, Spain, Switzerland, Turkey, United Arab Emirates, United Kingdom, Viet Nam.

Product description

MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography system.

Code information (models, lots, serial numbers)

UDI-DI: 00869346000200 ALL Serial numbers

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.