Source Administration Set (SAS) used with the Medrad Intego… recallZ-0910-2023 · Bayer Medical Care, Inc. · initiated Nov 30, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bayer Medical Care, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Black specks may be present on the packaged needles
FDA root cause category: Under Investigation by firm
What the company is doing
Bayer US issued Urgent Medical Device Recall (RECALL ID: SA-22-RAD-05 ) Letter on 11/30/22 via email. Letter states reason for recall, health risk and action to take: 1. Review your current inventory for affected Lot 60272154 and quarantine, as appropriate. 2. Complete the included response form and submit through email to [email protected] or contact our Bayer Customer Service team at 1-800-633-7231, Press 0, then option 3. Please complete the Customer Response Form regardless of whether you have affected product or not. This action will assist us in tracking all impacted product. 3. You will receive a Return Goods Authorization (RGA) number to return any affected kits remaining in your inventory for a credit. 4. Place the affected product in a box and label the outside of the container with the RGA number in large, bold writing. Use the shipping label provided via email when you receive your RGA number. Please return affected product as soon as possible 5. Upon Bayer receipt of returned product, you will receive a credit on your account. Please distribute this letter to other appropriate departments and personnel within your facility who may need awareness of this recall.
Recall details
- Recall number
- Z-0910-2023
- Status
- Open, Classified
- Initiated by the company
- Nov 30, 2022
- Posted by the FDA
- Jan 6, 2023
- Recalling firm
- Bayer Medical Care, Inc.
- Quantity
- 2840 units
- Product code
- FRN
- FDA submissions
- K100798
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AZ, IN, MN, NY, OH, TX, VA and the country of China.
Product description
Source Administration Set (SAS) used with the Medrad Intego PET Infusion System, Sterile Catalog number: INT CSS Affected material number: 86566621 Intended to deliver accurate doses of 18F Fluorodeoxyglucose (FDG) or 18F Sodium Fluoride (NaF) radiopharmaceuticals and commonly used flushing solutions to patients during molecular imaging (PET Positron Emission Tomography - nuclear medicine) diagnostic procedures.
Code information (models, lots, serial numbers)
UDI: (01)00616258021478(11)201209(17)231209(10)60272154 Batch number: 60272154 Expiration date: December 9, 2023
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- MEDRAD® Intego Sterile Disposables INT CSSBayer Medical Care Inc. · Pump, Infusion1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page