Skip to content

CODMAN recallZ-0909-2023 · Integra LifeSciences Corp. · initiated Dec 2, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Integra LifeSciences Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Integra LifeSciences identified that one lot was released with out-of-specification endotoxin results. The worst-case risk for the use of affected Surgical Patties is potentially a minor transient fever.

FDA root cause category: Process control

What the company is doing

On December 2, 2022, the firm notified customers via "URGENT VOLUNTARY MEDICAL DEVICE RECALL" letters. Customers were instructed to immediately remove units of affected product from service. Customers should complete the Acknowledgement Form and return to the recalling company. Once received, Customer Service will contact the customer to aid them in returning the affected product. Should you have any questions regarding these instructions, please contact Customer Service: Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: [email protected]

Recall details

Recall number
Z-0909-2023
Status
Open, Classified
Initiated by the company
Dec 2, 2022
Posted by the FDA
Jan 6, 2023
Recalling firm
Integra LifeSciences Corp.
Quantity
168 US; 3 OUS
Product code
HBA
FDA submissions
K193346
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, CO, CT, FL, IN, KY, MA, MI, MO, NC, NJ, NY, OH, PA, TX, VA, WV and the countries of Chile, Australia, New Zealand, France, Romania, United Kingdom, and Ireland.

Product description

Codman Surgical Patties, Part number 801400; SURG PATXRAY 1/2X1/2-200; Lot 6396457

Code information (models, lots, serial numbers)

UDI-DI 10381780514961 Lot Number: 6396457 Expiration date 06/30/2027

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K193346 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page