Turbett Surgical Container, TS1000: Indicated for enclosing… recallZ-1038-2023 · Turbett Surgical, Inc. · initiated Dec 5, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Turbett Surgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Pod may have weld failures breaching the sterile barrier and result in the use of unsterile instruments leading to infection or other major injuries.
FDA root cause category: Device Design
What the company is doing
***UPDATE*** Turbett Surgical issued via E-mails Product Alert directly to the Hospital departments utilizing the device on 12-27-2022. Letter states reason for recall, health risk and action to take: Actions to be taken by the Customer/User: 1. Immediately discontinue the use of the specific devices exhibiting one or more failure(s) listed below. " Missing side supports " Broken or cracked side tab support " Backwall weld separation " Side weld separation " Crack in the backwall seam 2. Email ([email protected]) with the following information. " Hospital name " Serial number(s) " Type of failure that your container exhibited 3. Turbett Surgical will contact you to organize pick up of the damaged unit and shipment of a replacement unit. If your container does not exhibit any of these failures, please continue to monitor and inspect the condition of the container and follow the Instructions For Use (IFU) as you continue utilizing your container. 4. If your container does not exhibit any of these failures, please email [email protected] stating that your container(s) do not exhibit any of these failures. Please continue to monitor and inspect your devices prior to assembling for sterilization AS WELL AS after sterilization and cooling to ensure expected outcomes. Reminder to follow the Instructions For Use (IFU) as you continue utilizing your TS-1200 container. If you have any questions about the content of this letter, please contact Mark Petras at [email protected]. ________________________________________________________________________ Turbett Surgical issued Emails to the Hospital department on 12/7/2022. Sales representatives were notified on 12/5/22 to switch out the other container. Letter states reason for recall, health risk and action to take: 1. Immediately verify that the specific serial numbered devices listed below, continues to operate without creating wet loads or the sterilization process is not compromised.
Recall details
- Recall number
- Z-1038-2023
- Status
- Open, Classified
- Initiated by the company
- Dec 5, 2022
- Posted by the FDA
- Jan 30, 2023
- Recalling firm
- Turbett Surgical, Inc.
- Quantity
- 6 units
- Product code
- FRG
- FDA submissions
- K153025
- Distribution
- US Nationwide distribution in the states of CA, IA, IN, NC, NH, NY, OH, LA, SD, TX, VA, WI.
Product description
Turbett Surgical Container, TS1000: Indicated for enclosing other medical devices that are to be sterilized by a healthcare provider
Code information (models, lots, serial numbers)
GTIN: 00859464007212 Lot Number: All Lot Numbers Serial Number: All Serial Numbers
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Turbett Surgical TS1000Turbett Surgical Inc · Wrap, Sterilization1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91314).
- Z-1037-2023 · Open, ClassifiedTurbett Surgical Container, TS1200: Indicated for enclosing other medical devices that are to be sterilized by a healthcare provider
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page