Future Diagnostics STAT-IntraOperative-Intact-PTH… recallZ-1016-2023 · Future Diagnostics Solutions B.V. · initiated Dec 6, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Future Diagnostics Solutions B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to an issue of high coefficient of variability (CV) problem caused by the accuspheres inside the Immunoassay kits.

FDA root cause category: Under Investigation by firm

What the company is doing

On 12/06/2022, the firm send a Recall Notice via email to customer informing them that The PTH kit 4K-IPT-00 with lot numbers: M78559, M78560 and M78636 can occasionally give a high duplicate %CV outside the recommended maximum of 15%CV. As a result of the failure in the duplicate CV requires a repeat run it may cause a delay in the patient result turnaround time or in system calibration depending on what samples are run and can lead to a wrong sample result. Actions to be taken by customers: -Immediately identify and quarantine the affected lot numbers. -Further instructions will be provided by the Future Diagnostics. For questions, send an email to [email protected].

Recall details

Recall number
Z-1016-2023
Status
Open, Classified
Initiated by the company
Dec 6, 2022
Posted by the FDA
Jan 25, 2023
Recalling firm
Future Diagnostics Solutions B.V.
Quantity
342 kits
Product code
CEW
FDA submissions
K000471
Distribution
Worldwide distribution. US states of AZ, CA, CO, LA, OH, PA, and TX; Austria, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, Spain, Sweden, and United Kingdom

Product description

Future Diagnostics STAT-IntraOperative-Intact-PTH (STAT-IO-I-PTH) Immunoassay kit REF 4K-IPT-00. in vitro diagnostic.

Code information (models, lots, serial numbers)

Catalog Number: 4K-IPT-00 UDI-DI: 08719326045214-230915 Lot Numbers: M78559 M78560 M78636