Skip to content

CRP Immunoturbidimetric reagent. Intended for the… recallZ-1286-2023 · Randox Laboratories · initiated Feb 1, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Randox Laboratories or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range

FDA root cause category: Under Investigation by firm

What the company is doing

Randox UK (manufacturer) extended Urgent Medical Device Removal letter to the US distributor via email on 2/1/23. The distributor contacted the customer via email on 2/6/23. Letter states reason for recall, health risk and action to take: " Discontinue use of and discard any of the above immediately. Provide Randox with photographic evidence of the destruction of the kits. " Review your reagent inventory of these products and assess your laboratories needs for reimbursement for discarded inventory. " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to [email protected] within five working days. If you have any questions or concerns, please contact Randox Technical Services.

Recall details

Recall number
Z-1286-2023
Status
Open, Classified
Initiated by the company
Feb 1, 2023
Posted by the FDA
Sep 30, 2024
Recalling firm
Randox Laboratories
Quantity
2 kits
Product code
DCK
FDA submissions
K953212
Distribution
PA, OH, WV

Product description

CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826

Code information (models, lots, serial numbers)

GTIN: 05055273201765 Batch: 588434

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page