Ultraview SL (UVSL) Command Module, Model 91496 recallZ-1503-2023 · Spacelabs Healthcare, Inc. · initiated Apr 12, 2023
Class II Open, Classified All recalls posted in May 2023
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Spacelabs Healthcare, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent with no alarms. This may cause clinicians to miss a potentially catastrophic event.
FDA root cause category: Component change control
What the company is doing
On 4/12/23 correction notices were sent and customers were provided the following information: In software versions up to and including version 2.03.06: "Processing Suspended" messaging overlays the waveforms, question marks (?) replace the typically found numerical data, and there is an "alarm off" symbol on the far right of the parameter zone while the product in in Suspend status. The patient s ECG and Respiratory waveforms remain displayed in the parameter zone of the user interface while the product is in Suspend status. In software version 2.04 and later, "Processing Suspended" messaging overlays the waveform zone, question marks (?) replace the typically found numerical data, and there is an "alarm off" symbol on the far right of the parameter zone while the product in in suspend status; however, the active patient waveforms are no longer displayed. We believe that the halted display of parameter waveforms more clearly communicate that the product is in Suspend status. The recalling firm recommends: 1) All users must be adequately trained on all versions of the product that will be used. 2) Ensure that all users are aware that Suspend Processing status means that all parameter processing and alarms are stopped and will not function. 2) The Suspended state must be manually re-engaged by pressing the "Resume Processing" button to return to active processing and return the alarms to the intended operation. 3) Please pass this notice on to those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. 4) All customers should acknowledge receipt and understanding of the notice at the following URL: https://spacelabshealthcare.com/support/recall-reply-form/2023-00001-c/ Questions regarding this correction can be directed to Technical Support at 1-800-522-7025, select option 2.
Recall details
- Recall number
- Z-1503-2023
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Apr 12, 2023
- Posted by the FDA
- May 3, 2023
- Recalling firm
- Spacelabs Healthcare, Inc.
- Quantity
- 1,088
- Product code
- DSI
- FDA submissions
- K103142
- Distribution
- US Nationwide distribution in the states of SD, KY, MO, WA, OH, NC, CA, OR, TX, GA, IL, OK, SC, PA, AL, KS, FL, ID, MI, MN, IA, NJ, MD, VA, AR, CT, NE, LA, AK, MS, ME, WV, NY, AZ, TN, WY, IN, MA, MT, CO, DE, ND, WI, NV, RI, NH, UT.
Product description
Ultraview SL (UVSL) Command Module, Model 91496
Code information (models, lots, serial numbers)
Software version 2.03.06 and earlier and software version 2.04 and later.
Other devices under the same FDA submission
2 devices cleared under K103142 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Ultraview SL Ultraview SL®Spacelabs Healthcare (Washington), Inc · SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE2 recalls2 reports
- Ultraview SL UV SL Command ModuleSpacelabs Healthcare (Washington), Inc · DETECTOR AND ALARM, ARRHYTHMIA1 recall3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page