iQ200 recallZ-1596-2023 · Beckman Coulter, Inc. · initiated Apr 5, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have been investigated.
FDA root cause category: Under Investigation by firm
What the company is doing
Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 04/05/2023 by mail. The notice explained the issue and the hazard and requested the following actions be taken: "Examine your specimen settings to determine if the optional 'Sperm Present' and 'Previous Sample Had Sperm' flags are enabled. If disabled, no further action is needed by your laboratory. If enabled, follow the actions below If the presence of sperm is identified: Review the previous specimen for the presence of sperm. Follow recommendations under Previous Sample Had Sperm in the iQ200 Series Instructions For Use (IFU) 300-4320CE and 300-4321EE and DxU 850m and DxU 840m Iris IFU (C49320AB) in Chapter 6, Data Review, Flags, Previous Sample Had Sperm, Recommendations. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter."
Recall details
- Recall number
- Z-1596-2023
- Status
- Open, Classified
- Initiated by the company
- Apr 5, 2023
- Posted by the FDA
- May 17, 2023
- Recalling firm
- Beckman Coulter, Inc.
- Quantity
- 1068 units
- Product code
- KQO
- FDA submissions
- K093861, K210127, K022774
- Distribution
- Worldwide Distribution
Product description
iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Numbers C10683 and 700-3375; in-vitro diagnostic device, urine analysis, c) iQ200 Select 1500, Part Number 700-3347, d) iQ200 Sprint, Part Number 700-3325
Code information (models, lots, serial numbers)
a) iQ 200 Select, Part Numbers C10684, UDI/DI 15099590695484 and 700-3345, UD/DI 10837461001300; b) iQ 200 Elite, Part Numbers C10683, UDI/DI 15099590697129 and 700-3375, UDI/DI 10837461001355; c) iQ200 Select 1500, Part Number 700-3347, UD/DI 10837461001317; d) iQ200 Sprint, Part Number 700-3325, UDI/DI 10837461001256; All analyzers with Windows 10, Software Version 8.1;
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- iQ200 Elite Plus (US) (2008 Version) (NA)Beckman Coulter Ireland Inc. · Automated Urinalysis System3 recalls2 reports
- iQ200 Sprint Pro (NA)/SPRINT w load/unload (NA)/SPRINT Pro w load/unload (NA)/Sprint PBeckman Coulter Ireland Inc. · Automated Urinalysis System3 recalls1 report
- iQ200 Select/Elite Pro (NA)Beckman Coulter Ireland Inc. · Automated Urinalysis System3 recalls0 reports
- iQ200 Select Plus Kit 2008 Analyzer (NA)/ Select Pro (NA)Beckman Coulter Ireland Inc. · Automated Urinalysis System2 recalls4 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92056).
- Z-1597-2023 · Open, ClassifiedDxU 850m Iris Urine Microscopy Analyzer Catalog Number C49513, in-vitro diagnostic device, urine analysis
- Z-1598-2023 · Open, ClassifiedDxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic device, urine analysis
- Z-1599-2023 · Open, ClassifiedUA WIN 10 Kit, Catalog Number C52900, in-vitro diagnostic device, urine analysis
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page