Atellica IM Cor recallZ-1504-2023 · Siemens Healthcare Diagnostics, Inc. · initiated Mar 16, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a negative bias with urine patient samples and urine cortisol Quality Control (QC) results intermittently out of range low.
FDA root cause category: Under Investigation by firm
What the company is doing
An Urgent Medical Device Correction notification letter dated 3/17/23 was sent to customers. Actions to be Taken by the Customer Discontinue use of urine patient samples with the Atellica IM Cortisol and ADVIA Centaur Cortisol assays until further notice. Customers may continue to use serum and plasma sample types with the Atellica IM Cortisol and ADVIA Centaur Cortisol assays. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. An additional Resolution Customer Notification letter dated 2/29/24 was sent to customers explaining an improvement to raw material specification that restores assay performance with urine samples. Customers are asked to follow the instructions for urine patient samples and urine cortisol QC as stated in the notification and if using products listed in the original field action notice continue following the guidance as stated in the Field Action Letter. Customers with any questions are to contact the Customer Care Center or local Siemens Healthineers technical support representative.
Recall details
- Recall number
- Z-1504-2023
- Status
- Open, Classified
- Initiated by the company
- Mar 16, 2023
- Posted by the FDA
- May 3, 2023
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 34,746 kits
- Product code
- JFT
- FDA submissions
- K142723
- Distribution
- Worldwide distribution - US Nationwide.
Product description
Atellica IM Cortisol 50T (Material Number 10995538), 250T (Material Number 10995537), and REF 250T (Material Number 11206248). In vitro diagnostic use in the quantitative determination of cortisol in human serum, plasma (EDTA and lithium heparin), and urine.
Code information (models, lots, serial numbers)
Atellica IM Cortisol (50T) UDI-DI: 00630414598659; Kit Lot Numbers: 50732343 50733343 88975345 29106347 29107347 and higher; Atellica IM Cortisol 250T UDI-DI: 00630414598642; Kit Lot Numbers: 50734343 88974345 29105347 and higher; Atellica IM Cortisol (REF) 250T UDI-DI: 00630414293608; Kit Lot Numbers: 53842343 58039343 94430345 and higher
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Atellica IM Cor 10995538Siemens Healthcare Diagnostics Inc. · FLUOROMETRIC, CORTISOL2 recalls0 reports
- Atellica IM Cor 11206248Siemens Healthcare Diagnostics Inc. · FLUOROMETRIC, CORTISOL1 recall0 reports
- Atellica IM Cor 10995537Siemens Healthcare Diagnostics Inc. · FLUOROMETRIC, CORTISOL1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92059).
- Z-1505-2023 · Open, ClassifiedADVIA Centaur Cortisol 50T (Material Number 10994924), ADVIA Centaur Cortisol 250T (Material Number 10994926), and ADVIA Centaur Cortisol REF 250T (Material Number 10994927). In vi…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page