SINGLE USE DISTAL COVER recallZ-1613-2023 · Olympus Corporation of the Americas · initiated Apr 18, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Single Use Distal Cover MAJ-2315 may cause mucosal injury and/or become detached from the endoscope during use. Consequences of a detached cover could include the risk of aspiration, inhalation, or obstruction that will require urgent removal, and/or burns from an uncovered distal end.
FDA root cause category: Device Design
What the company is doing
An URGENT - Medical Device Recall notification letter dated 4/18/23 was sent to customers. Instructions for Use: Correct application and inspection of the distal cover before the procedure is critical. In addition, inspection of the distal end after the procedure is also important. Olympus representatives are available to provide additional support and training. Please contact your local Olympus facility to further discuss. Action Required: Our records indicate that you have purchased or may use the MAJ-2315 Distal Covers with the original design. Olympus requires you to take the following action: 1. Immediately locate and no longer use the MAJ-2315 with the original design. The original design can be identified by the item code (below) or by lot numbers less than H2530. The image below depicts the area where the lot number is identified. The lot number is on the Carton box and Pack. Model: MAJ-2315 Item code: N5786100 Lot number: Lower than H2530 2. Contact our Olympus Customer Service Team at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility upon return of affected product. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a) Go to https://olympusamerica.com/recall b) Enter the recall number: 0423 c) Complete the form as instructed. 4. If you have distributed these devices outside your facility, please notify your customers of this matter immediately. Your notification to your customers may be enhanced by including a copy of this letter. Olympus requests that you report complaints, including mucosal injuries, distal cover detachment and adverse events, to Olympus. Please report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, to report complaints. Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration
Recall details
- Recall number
- Z-1613-2023
- Status
- Open, Classified
- Initiated by the company
- Apr 18, 2023
- Posted by the FDA
- May 19, 2023
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 309278 pieces
- Product code
- FDT
- FDA submissions
- K193182
- Distribution
- US Nationwide.
Product description
Olympus Single Use Distal Cover Model Number MAJ-2315 used with EVIS EXERA III Duodenovideoscope TJF-Q190V; Item code: Catalogue #N5786100
Code information (models, lots, serial numbers)
UDI-DI: 04953170403019; Item code: Catalog N5786100 Lots Lower than H2530
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- SINGLE USE DISTAL COVER MAJ-2315Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall520 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page