Omnicell IVX Station-For use in the automated preparation… recallZ-1617-2023 · OmniCell, Inc. · initiated Mar 27, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from OmniCell, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling
FDA root cause category: Under Investigation by firm
What the company is doing
Omincell notified Customers via phone on 3/27/2023 and by Urgent Field Action: Mandatory Correction Of Product Labeling Ivx Station (Limited Release) Letter on 3/31/2023 via email. Letter states reason for recall, health risk and action to take: We will analyze specific devices at your Limited Release sites. Upon completion of our analysis updated labeling will be provided and will include the revised dose accuracy range.
Recall details
- Recall number
- Z-1617-2023
- Status
- Open, Classified
- Initiated by the company
- Mar 27, 2023
- Posted by the FDA
- May 22, 2023
- Recalling firm
- OmniCell, Inc.
- Quantity
- 6 units
- Product code
- NEP
- Distribution
- AL, NY
Product description
Omnicell IVX Station-For use in the automated preparation of pharmaceutical admixtures and syringes Model Number: 298921373
Code information (models, lots, serial numbers)
UDI-DI: 00860006285005 Serial Numbers: IVXS000014, IVXS000015, IVXS000016, IVXS000017, IVXS000018. IVXS000019
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- IVX StationOmnicell, Inc. · System/Device, Pharmacy Compounding1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.