Skip to content

Distractor Rack, REF 5598007, which may be packaged in… recallZ-1574-2023 · Medtronic Sofamor Danek USA Inc · initiated Nov 19, 2021

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Sofamor Danek USA Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor.

FDA root cause category: Under Investigation by firm

What the company is doing

Medtronic issued an URGENT; Medical Device Removal notice on 11/19/2021. The notice explained the issue and the risk and requested the removal of the affected product from use pending return. It also directed the consignee to transfer the notice to other organizations which could be affected.

Recall details

Recall number
Z-1574-2023
Status
Open, Classified
Initiated by the company
Nov 19, 2021
Posted by the FDA
May 10, 2023
Recalling firm
Medtronic Sofamor Danek USA Inc
Quantity
4 units
Product code
LXH
Distribution
US: IN, CT, NE

Product description

Distractor Rack, REF 5598007, which may be packaged in Modular Distractor Set, Loaner Kit Number SPS03139; for use in spinal surgery

Code information (models, lots, serial numbers)

UDI/DI (GTIN) 00763000401535, Lot Numbers EM21D006, EM21D007; Loaner Kit Lot numbers: 0072, 0073, 0077, 0098, 0103

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page