CODMAN recallZ-1614-2023 · Integra LifeSciences Corp. · initiated Apr 11, 2023
Class II Open, Classified All recalls posted in May 2023
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Integra LifeSciences Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Released with discoloration (rust) on the Drill Chuck and inner chuck of Codman Cranial Hand Drill may result in inflammation, local toxicity, systemic toxicity, sensitivity, revision surgery, procedural delay, inconvenience to user, and/or infection.
FDA root cause category: Under Investigation by firm
What the company is doing
Codman issued Urgent Medical Device Recall Letter on 4/11/23 via Fed'X. Letter states reason for recall, health risk and action to take: 1. If you do have units of the affected product listed in Table 1, remove them immediately from service. 2. If you do have affected product, check the box on the enclosed form I do have affected product. Record the total quantity of the affected product that you have. 3. If you do not have affected product, check the box, I do not have affected product. 4. Complete the attached Acknowledgement Form and return to [email protected] i. or FAX to 1-609-750-4220. 5. Keep a copy of the form for your records. 6. When your form is received, and it is noted that you have affected product, Customer Service will contact you and provide a Return Material Authorization (RMA) number and directions to return the affected product. A credit or alternative replacement can be placed for the quantity noted on the form. Receipt of this form ensures that Integra has achieved a level of effectiveness in communicating this information. We recommend you also maintain a copy of this notification and signed copy of the acknowledgement form for your records. Regulatory agencies such as the FDA perform audits of field actions of this nature to verify that customers have been notified and understand the nature of the field action. Should you have any questions regarding these instructions, please contact your Integra Sales Representative or Customer Service: Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: [email protected]
Recall details
- Recall number
- Z-1614-2023
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Apr 11, 2023
- Posted by the FDA
- May 19, 2023
- Recalling firm
- Integra LifeSciences Corp.
- Quantity
- 337 units
- Product code
- GWM
- FDA submissions
- K914479
- Distribution
- PA Foreign: Australia, Belgium
Product description
Codman Cranial Hand Drill- Indicated when a craniotomy is required for placement of an intracranial pressure (ICP) monitoring device and/or cerebrospinal fluid drainage device. Part Number: 82-6607
Code information (models, lots, serial numbers)
UDI-DI: 10381780520337 Lot Numbers: 6808685, 6563961, 6568711, 6568709, 6568710
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CODMAN® 826607Integra Lifesciences Production Corporation · Device, monitoring, intracranial pressure1 recall2 reports
Product lines
Other devices under the same FDA submission
8 devices cleared under K914479 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Codman Microsensor® 826631Integra Lifesciences Production Corporation · Device, monitoring, intracranial pressure75 reports
- CODMAN® 826614Integra Lifesciences Production Corporation · Device, monitoring, intracranial pressure1 recall1 report
- CODMAN® 826616Integra Lifesciences Production Corporation · Device, monitoring, intracranial pressure1 recall0 reports
- CODMAN MICROSENSOR 82-6631Cerenovus, Inc. · DEVICE, MONITORING, INTRACRANIAL PRESSURE14 reports
- Codman Microsensor® 826638Integra Lifesciences Production Corporation · Device, monitoring, intracranial pressure11 reports
- CODMAN MICROSENSOR 82-6638Cerenovus, Inc. · DEVICE, MONITORING, INTRACRANIAL PRESSURE6 reports
- CODMAN® 826612Integra Lifesciences Production Corporation · Device, monitoring, intracranial pressure1 report
- Codman Microsensor® 826632Integra Lifesciences Production Corporation · Device, monitoring, intracranial pressure1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page