CODMAN recallZ-1614-2023 · Integra LifeSciences Corp. · initiated Apr 11, 2023

Class II Open, Classified All recalls posted in May 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Integra LifeSciences Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Released with discoloration (rust) on the Drill Chuck and inner chuck of Codman Cranial Hand Drill may result in inflammation, local toxicity, systemic toxicity, sensitivity, revision surgery, procedural delay, inconvenience to user, and/or infection.

FDA root cause category: Under Investigation by firm

What the company is doing

Codman issued Urgent Medical Device Recall Letter on 4/11/23 via Fed'X. Letter states reason for recall, health risk and action to take: 1. If you do have units of the affected product listed in Table 1, remove them immediately from service. 2. If you do have affected product, check the box on the enclosed form I do have affected product. Record the total quantity of the affected product that you have. 3. If you do not have affected product, check the box, I do not have affected product. 4. Complete the attached Acknowledgement Form and return to [email protected] i. or FAX to 1-609-750-4220. 5. Keep a copy of the form for your records. 6. When your form is received, and it is noted that you have affected product, Customer Service will contact you and provide a Return Material Authorization (RMA) number and directions to return the affected product. A credit or alternative replacement can be placed for the quantity noted on the form. Receipt of this form ensures that Integra has achieved a level of effectiveness in communicating this information. We recommend you also maintain a copy of this notification and signed copy of the acknowledgement form for your records. Regulatory agencies such as the FDA perform audits of field actions of this nature to verify that customers have been notified and understand the nature of the field action. Should you have any questions regarding these instructions, please contact your Integra Sales Representative or Customer Service: Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: [email protected]

Recall details

Recall number
Z-1614-2023
Recall class
Class II
Status
Open, Classified
Initiated by the company
Apr 11, 2023
Posted by the FDA
May 19, 2023
Recalling firm
Integra LifeSciences Corp.
Quantity
337 units
Product code
GWM
FDA submissions
K914479
Distribution
PA Foreign: Australia, Belgium

Product description

Codman Cranial Hand Drill- Indicated when a craniotomy is required for placement of an intracranial pressure (ICP) monitoring device and/or cerebrospinal fluid drainage device. Part Number: 82-6607

Code information (models, lots, serial numbers)

UDI-DI: 10381780520337 Lot Numbers: 6808685, 6563961, 6568711, 6568709, 6568710

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

8 devices cleared under K914479 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.