Skip to content

TDHisto/Cyto - A software product used for managing medical… recallZ-1573-2023 · Technidata S.A. · initiated Mar 27, 2023

Class II Open, Classified All recalls posted in May 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Technidata S.A. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

In a specific use case, when printing labels for slides, some labels may display wrong information

FDA root cause category: Software design

What the company is doing

Techidata issued Medical Device Recall Urgent letter via email on 3/27/23. Letter states reason for recall, health risk and action to take: To prevent any occurrence of the issue, you must not close the current session, or open another one, or perform the printing of another label batch, until the label printing is finished. Note: printing is finished when the Printing dialog box closes. software modification was performed to prevent the risks mentioned above. We kindly ask you to fill the reply form below and return it to us at your earliest convenience.

Recall details

Recall number
Z-1573-2023
Recall class
Class II
Status
Open, Classified
Initiated by the company
Mar 27, 2023
Posted by the FDA
May 10, 2023
Recalling firm
Technidata S.A.
Quantity
1 unit
Product code
JQP
Distribution
US Nationwide distribution in the state of CA.

Product description

TDHisto/Cyto - A software product used for managing medical information in the field of laboratories performing histology, nongynecological and gynecological cytology and autopsy activities.

Code information (models, lots, serial numbers)

UDI-DI : 03770027519010 All TDHisto/Cyto software versions

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page