VENTRALIGHT recallZ-2241-2023 · Davol, Inc. · initiated Apr 25, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Davol, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Balloon may not inflate as intended. If the balloon does not inflate during laparoscopic ventral hernia repair this may result in a delay or prolongation of the procedure, inadequate mesh placement, erosion/migration, pain and mesh infection
FDA root cause category: Under Investigation by firm
What the company is doing
BD issued Urgent Medical Device Product Recall letter on 4/25/23 via FedEx 2-day and email. Letter states reason for recall, health risk and action to take: 1. Please immediately discontinue use of the affected product. Check all inventory locations within your facility and destroy all affected product in accordance with your facility s process. 2. Share this notification with all users of the affected product within your facility network to ensure they are also aware of this product recall. 3. If you purchased the affected product from a distributor, contact your distributor directly for further instructions and credit resolution. 4. Complete the attached Customer Response Form and return it to the BD contact noted on the form whether or not you have any of the affected product so that BD may acknowledge your receipt of this notification per FDA requirements. 5. Indicate on the response form the quantity of affected product identified at your facility and confirm that this product inventory was destroyed. 6. Please contact your BD representative if you require assistance with this process, or you may also contact BD. 1. BD will issue replacements for destroyed product to customers that purchased product directly from BD following receipt of the completed Customer Response Form. 2. BD is implementing actions to prevent recurrence of this product issue. Contact Information: If you require further assistance please contact: BD Contact Contact Information Areas of Support North American Regional Complaint Center Phone: 1-844-8BD- LIFE (1-844-823-5433) Say Recall when prompted M-F 8am - 5pm CT Email: [email protected] Recall questions, Product Complaints, Technical Questions
Recall details
- Recall number
- Z-2241-2023
- Status
- Open, Classified
- Initiated by the company
- Apr 25, 2023
- Posted by the FDA
- Jul 27, 2023
- Recalling firm
- Davol, Inc.
- Quantity
- 1188 units US; 5 units OUS
- Product code
- FTL
- FDA submissions
- K130968
- Distribution
- Worldwide distribution - US Nationwide and the countries of Belgium, Czech Republic, France, Germany, Ireland, Italy, Netherlands, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK, India, Hong Kong, Greece, Chile, and Colombia.
Product description
Ventralight ST 4" x 6" Ellipse with Echo PS Catalog Number: 5955460
Code information (models, lots, serial numbers)
UDI-DI: (01)00801741031724(17)240428(10) Lot Numbers: HUGS0481 HUGS0549 HUGS1019 HUGS2031 HUGS2079 HUGT1627 HUGS2025 HUGT1989
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- VENTRALIGHT 5955460Davol Inc. · Laparoscope, general & plastic surgery2 recalls91 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92213).
- Z-2242-2023 · Open, ClassifiedVentralight ST 4.5" Circle with Echo PS- A low profile, bioresorbable, coated, permanent mesh, with a pre-attached removable positioning system, designed for the reconstruction of …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page