PushTracker E2 component utilized with the SmartDrive MX2+… recallZ-2004-2023 · Permobil · initiated May 10, 2023
Class II Open, Classified All recalls posted in June 2023
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Permobil or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.
FDA root cause category: Software design
What the company is doing
Permobil began notifying customers via in-person and through letters on 05/10/2023. Customers were instructed to examine their device and update their software to SmartDrive MX2+ Version 1.1.00, to acknowledge the update was conducted and the notification letter was understood through the field action portal, notify customers if the affected devices were further distributed. They were also instructed to discontinue using the affected device while waiting for the correction.
Recall details
- Recall number
- Z-2004-2023
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- May 10, 2023
- Posted by the FDA
- Jun 26, 2023
- Recalling firm
- Permobil
- Quantity
- 18,905 units
- Product code
- ITI
- FDA submissions
- K151199
- Distribution
- Worldwide distribution.
Product description
PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application
Code information (models, lots, serial numbers)
All units with software versions 1.0.00 to 1.0.04
Other devices under the same FDA submission
2 devices cleared under K151199 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- SmartDrive® MX2+Max Mobility, LLC · Wheelchair, Powered858 reports
- SmartDrive® MX2Max Mobility, LLC · Wheelchair, Powered6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92250).
- Z-2005-2023 · Open, ClassifiedPushTracker E3 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page