ET III Nano-HA fixture System ET NH Dental Implant Model #… recallZ-1604-2023 · Hiossen Inc. · initiated Apr 6, 2023
Class II Open, Classified All recalls posted in May 2023
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Hiossen Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Mis-labeled length of the dental implant on the Recall Implant's package. Dental implant size D5.0 x L8.5mm, labeled as D5.0 x L10mm
FDA root cause category: Labeling Change Control
What the company is doing
Hoissen Inc. initial contact verbally through sales rep. on April 7, 2023 and by "URGENT: MEDICAL DEVICE RECALL" Letter via Mail & email on 04/19/2023. Letter states reason for recall, health risk and action to take: 1. Do not use the Recall Implants. 2. Quarantine (separate) the Recall Implants in a different location, away from normal stock room. 3. Immediately fill out the Acknowledgement and Receipt Form* (Fax: 267-795-7031 or Email: [email protected]) *Response is required regardless if the Recall Implants have already been return or implanted. 4. Returned the Recall Implant(s) as Unusable Returns to Hiossen QA** **Contact your sales representative for assistance. 5. Replacement devices are available for delivery. If you need any further information or support concerning this recall/issue, please reach HIOSSEN: Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time. Tel: 267-795-7031 Email: [email protected]
Recall details
- Recall number
- Z-1604-2023
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Apr 6, 2023
- Posted by the FDA
- May 18, 2023
- Recalling firm
- Hiossen Inc.
- Quantity
- 287 units
- Product code
- DZE
- FDA submissions
- K151626
- Distribution
- US Nationwide Distribution
Product description
ET III Nano-HA fixture System ET NH Dental Implant Model # ET3R5010B -Intended for Dental Implant Catalog #: AET3R5010B
Code information (models, lots, serial numbers)
UDI-DI: 00814913023717 Lot Number: H1E22J263
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ET III Fixture NH Regular Pre-Mounted AET3R5010BHiossen, Inc. · Implant, Endosseous, Root-Form1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.