Skip to content

Micro-Access Tearaway Introducer Kit recallZ-1968-2023 · Galt Medical Corporation · initiated May 8, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Galt Medical Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Sheath introducer assemblies could potentially, due to improper storage conditions, have a luer connection portion of the dilator hub that cracks or dislodges during use, which renders the dilator component unusable, and can require the use of a replacement device.

FDA root cause category: Process design

What the company is doing

On 5/5/23, recall notices were sent to customers who were asked to do the following: 1) Identify and segregate the affected lot(s) that are in your possession as well as those in the possession of your end-users. 2) Complete and return the field correction reply form to [email protected] 2) Return affected product to the recalling firm. Customers with questions can email: [email protected]

Recall details

Recall number
Z-1968-2023
Status
Open, Classified
Initiated by the company
May 8, 2023
Posted by the FDA
Jun 15, 2023
Recalling firm
Galt Medical Corporation
Quantity
490
Product code
DYB
FDA submissions
K153533
Distribution
Worldwide - US Nationwide distribution in the states of PA, UT, FL, PA, OH, GA, MA, NV, CA, NY, NJ and the countries of UK, AU, TR, FR, CH, NO.

Product description

Tearaway Introducer Kit, REF: KIT-010-12, KIT-010-17, KIT-900-16

Code information (models, lots, serial numbers)

REF/UDI-DI/LOT (Expiration): KIT-010-12/00841268105195/S23104607(2/05/2027), S22356870(1/11/2027); KIT-010-17/00841268105249/S23083725(2/05/2027), S22231321(9/08/2026), S22328872(1/18/2027); KIT-900-16/S22308637(1/04/2027)

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 92322).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page