Skip to content

MEDLINE THERAHONEY GEL, 100% LEPTOSPERMUM SCOPARIUM HONEY… recallZ-1982-2023 · MEDLINE INDUSTRIES, LP - Northfield · initiated May 3, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medline issued a recall on TheraHoney Wound Gel due to the manufacturer notifying Medline of a sterilization dose audit failure on the wound gel and not augmenting the dose after failure per ISO 11137.

FDA root cause category: Under Investigation by firm

What the company is doing

Medline issued a recall notice to its consignees on 05/03/2023 by first calls mail and email The notice explained the problem with the device, risk and requested the following: "REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers listed on the recall portal. Destroy affected product.2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-23-059 Recall Code: RXXXXXXX 3.Upon completion of the form, please destroy affected product. Your account will receive credit once the response form is submitted. 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Have the customers document and destroy any affected product. You should include your customers quantities on your response form."

Recall details

Recall number
Z-1982-2023
Status
Open, Classified
Initiated by the company
May 3, 2023
Posted by the FDA
Jun 21, 2023
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
851,328 units
Product code
FRO
FDA submissions
K131796
Distribution
Worldwide distribution - US Nationwide and the countries of SINGAPORE, DOMINICAN REPUBLIC, UNITED ARAB EMIRATES, DUBAI, CHINA, PANAMA, CANADA, SAUDI ARABIA, BERMUDA, COLOMBIA, MAURITIUS, KUWAIT.

Product description

MEDLINE THERAHONEY GEL, 100% LEPTOSPERMUM SCOPARIUM HONEY, a) REF MNK0005, b) REF MNK0015 wound gel

Code information (models, lots, serial numbers)

a) REF MNK0005: GTIN 30884389161566, Lot Numbers: LMPTH0189, LMPTH0191, LMPTH0193, LMPTH0195, LMPTH0197, LMPTH0199, LMPTH0201, LMPTH0203, LMPTH0205, LMPTH0207, LMPTH0211, LMPTH0213; b) REF MNK0015: GTIN 40884389154558, Lot Numbers: LMPTH0188, LMPTH0190, LMPTH0192, LMPTH0196, LMPTH0198, LMPTH0200, LMPTH0202, LMPTH0204, LMPTH0206, LMPTH0210, LMPTH0210B, LMPTH0212

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates