Draeger Softbed Resuscitaire and/or Birthing Room Warmer… recallZ-1992-2023 · Draeger Medical Systems, Inc. · initiated May 16, 2023
Class II Open, Classified All recalls posted in June 2023
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Draeger Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Resuscitaire Infant Radiant Warmer with the optional scale could display inaccurate weight values.
FDA root cause category: Device Design
What the company is doing
An Urgent - Medical Device Correction dated May 2023 was sent to customers. Actions to be taken: Please ensure that all users of the above-mentioned products and other persons within your organization are made aware of this Important Safety Notice. If you have provided the products to third parties, please forward a copy of this information. Users can mitigate this risk by performing the visual check, test procedure and take appropriate actions based on the Attachment provided in this notice. Complete and return the enclosed Customer Acknowledgement and Response Form . We are continuing to evaluate this issue and will advise the affected customers when we have an alternative solution. Visual Check for Correct Mattress Placement: To ensure proper operation of the scale, please check to confirm there is a clearance between the mattress cover and the corner blocks and side panels of the Resuscitaire. 1. Ensure mattress with cover is installed in the Resuscitaire bassinet. 2. Observe edges of mattress for clearance with corner blocks and side panels, see Figure 2. 3. If there is interference, then the scale may not provide accurate weight measurements. It is recommended to use an independent scale for therapeutic decisions and treatment Test Procedure If it is unclear from the visual check if the mattress cover interferes with the corner blocks and/or side panels, impacting scale function, please perform the following test: a. Place a known calibrated weight on the center mattress (with mattress cover installed). A 5kg weight is recommended in the instructions for use. b. Execute weighing procedure and record weight. c. Repeat weighing procedure 5-10 times. d. Compare measurements. i. If measurements are within 5 grams of calibrated 5kg weight, mattress assembly is acceptable. ii. If measurements deviate more than +/- 5 grams from the calibrated 5kg weight, it is recommended to use an alternate weighing scale Contact: We regret any inconvenience this may cause.
Recall details
- Recall number
- Z-1992-2023
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- May 16, 2023
- Posted by the FDA
- Jun 22, 2023
- Recalling firm
- Draeger Medical Systems, Inc.
- Quantity
- 936 units
- Product code
- FMT
- FDA submissions
- K120642
- Distribution
- US Nationwide distribution including in the states of AL, AR, AZ, CA, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO MS, MT, NC, NE, NH, NJ, NV, NY, OH, OR, PA, PR, SD, TN, TX, VT, WA, and WI.
Product description
Draeger Softbed Resuscitaire and/or Birthing Room Warmer with Scale [Model(s): RW82VHA-1, RW82VHA-1C, RW82VHA-1D, RW82-1, RW82-1C, RW82-1D, WBR82-1 and WBR82-1C]
Code information (models, lots, serial numbers)
All units shipped from January 20, 2020 through March 1, 2023; MU20503 (Resuscitaire) UDI - 04049098000215 MU20512 (Birthing Room Warmer - BRW) UDI - 04049098058001
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Resuscitaire Radiant Warmer (RW82, RW82 VHA) MU20503Draeger Medical Systems, Inc. · Warmer, Infant Radiant1 recall3 deaths15 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.