Draeger Softbed Resuscitaire and/or Birthing Room Warmer… recallZ-1992-2023 · Draeger Medical Systems, Inc. · initiated May 16, 2023

Class II Open, Classified All recalls posted in June 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draeger Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Resuscitaire Infant Radiant Warmer with the optional scale could display inaccurate weight values.

FDA root cause category: Device Design

What the company is doing

An Urgent - Medical Device Correction dated May 2023 was sent to customers. Actions to be taken: Please ensure that all users of the above-mentioned products and other persons within your organization are made aware of this Important Safety Notice. If you have provided the products to third parties, please forward a copy of this information. Users can mitigate this risk by performing the visual check, test procedure and take appropriate actions based on the Attachment provided in this notice. Complete and return the enclosed Customer Acknowledgement and Response Form . We are continuing to evaluate this issue and will advise the affected customers when we have an alternative solution. Visual Check for Correct Mattress Placement: To ensure proper operation of the scale, please check to confirm there is a clearance between the mattress cover and the corner blocks and side panels of the Resuscitaire. 1. Ensure mattress with cover is installed in the Resuscitaire bassinet. 2. Observe edges of mattress for clearance with corner blocks and side panels, see Figure 2. 3. If there is interference, then the scale may not provide accurate weight measurements. It is recommended to use an independent scale for therapeutic decisions and treatment Test Procedure If it is unclear from the visual check if the mattress cover interferes with the corner blocks and/or side panels, impacting scale function, please perform the following test: a. Place a known calibrated weight on the center mattress (with mattress cover installed). A 5kg weight is recommended in the instructions for use. b. Execute weighing procedure and record weight. c. Repeat weighing procedure 5-10 times. d. Compare measurements. i. If measurements are within 5 grams of calibrated 5kg weight, mattress assembly is acceptable. ii. If measurements deviate more than +/- 5 grams from the calibrated 5kg weight, it is recommended to use an alternate weighing scale Contact: We regret any inconvenience this may cause.

Recall details

Recall number
Z-1992-2023
Recall class
Class II
Status
Open, Classified
Initiated by the company
May 16, 2023
Posted by the FDA
Jun 22, 2023
Recalling firm
Draeger Medical Systems, Inc.
Quantity
936 units
Product code
FMT
FDA submissions
K120642
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO MS, MT, NC, NE, NH, NJ, NV, NY, OH, OR, PA, PR, SD, TN, TX, VT, WA, and WI.

Product description

Draeger Softbed Resuscitaire and/or Birthing Room Warmer with Scale [Model(s): RW82VHA-1, RW82VHA-1C, RW82VHA-1D, RW82-1, RW82-1C, RW82-1D, WBR82-1 and WBR82-1C]

Code information (models, lots, serial numbers)

All units shipped from January 20, 2020 through March 1, 2023; MU20503 (Resuscitaire) UDI - 04049098000215 MU20512 (Birthing Room Warmer - BRW) UDI - 04049098058001