Claria MRI CRT-D SureScan recallZ-1753-2023 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · initiated May 10, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA root cause category: Under Investigation by firm
What the company is doing
Medtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Recall details
- Recall number
- Z-1753-2023
- Status
- Open, Classified
- Initiated by the company
- May 10, 2023
- Posted by the FDA
- Jun 16, 2023
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Quantity
- 11794 units
- Product code
- NIK
- FDA submissions
- P010031S550
- Distribution
- US Nationwide - Worldwide Distribution
Product description
CRTD DTMA1D1 CLARIA MRI US DF1, Model Number DTMA1D1; Implantable Cardioverter Defibrillators
Code information (models, lots, serial numbers)
Show the full code information (142k characters)
GTIN 00643169837676, Lot Serial Numbers: RPT208454H, RPT211000H, RPT208258H, RPT211071H, RPT211077H, RPT210886H, RPT209224H, RPT211338H, RPT211343H, RPT211344H, RPT211347H, RPT208864H, RPT210392H, RPT204063H, RPT207540H, RPT210779H, RPT211477H, RPT210662H, RPT203810H, RPT204233H, RPT210695H, RPT204352H, RPT204713H, RPT205727H, RPT208222H, RPT208507H, RPT210829H, RPT211217H, RPT206139H, RPT207062H, RPT210688H, RPT210917H, RPT203993H, RPT205235H, RPT205768H, RPT206518H, RPT206833H, RPT209343H, RPT209843H, RPT205560H, RPT206519H, RPT210915H, RPT210978H, RPT207320H, RPT204196H, RPT205788H, RPT207588H, RPT209206H, RPT209796H, RPT210253H, RPT210444H, RPT210700H, RPT211542H, RPT211685H, RPT205575H, RPT205841H, RPT207926H, RPT204420H, RPT205180H, RPT204120H, RPT204133H, RPT209974H, RPT205340H, RPT209909H, RPT210452H, RPT211124H, RPT203859H, RPT205468H, RPT208257H, RPT208621H, RPT209674H, RPT211043H, RPT210842H, RPT211476H, RPT207094H, RPT208370H, RPT206004H, RPT206049H, RPT210092H, RPT210217H, RPT211406H, RPT204980H, RPT205294H, RPT208424H, RPT208705H, RPT209144H, RPT209238H, RPT209858H, RPT210083H, RPT210593H, RPT210747H, RPT210979H, RPT211230H, RPT211701H, RPT209577H, RPT209848H, RPT211158H, RPT205289H, RPT206148H, RPT209521H, RPT204605H, RPT206936H, RPT207571H, RPT209334H, RPT209373H, RPT204461H, RPT207429H, RPT209096H, RPT210964H, RPT211140H, RPT205903H, RPT208605H, RPT211566H, RPT206634H, RPT203640H, RPT203792H, RPT203793H, RPT203795H, RPT204568H, RPT204574H, RPT204577H, RPT204582H, RPT204584H, RPT204607H, RPT205619H, RPT205620H, RPT205622H, RPT205624H, RPT206785H, RPT207549H, RPT207550H, RPT207556H, RPT207561H, RPT207562H, RPT207566H, RPT207568H, RPT208590H, RPT208591H, RPT208597H, RPT208599H, RPT208601H, RPT209722H, RPT210663H, RPT210668H, RPT210670H, RPT208269H, RPT206355H, RPT208891H, RPT205697H, RPT205434H, RPT210488H, RPT204439H, RPT207232H, RPT208034H, RPT209462H, RPT203726H, RPT206932H, RPT207021H, RPT210596H, RPT211478H, RPT210913H, RPT203791H, RPT205433H, RPT206774H, RPT211487H, RPT209945H, RPT211316H, RPT204451H, RPT204615H, RPT204619H, RPT204621H, RPT205498H, RPT208563H, RPT208821H, RPT208827H, RPT208828H, RPT208831H, RPT208832H, RPT210307H, RPT210309H, RPT210310H, RPT210311H, RPT211219H, RPT209965H, RPT210101H, RPT210345H, RPT210552H, RPT207931H, RPT210890H, RPT205714H, RPT208626H, RPT211368H, RPT210287H, RPT204343H, RPT204808H, RPT205191H, RPT209632H, RPT208738H, RPT209129H, RPT210394H, RPT204415H, RPT207421H, RPT207989H, RPT210527H, RPT211328H, RPT205965H, RPT206335H, RPT206424H, RPT209985H, RPT210011H, RPT210059H, RPT211191H, RPT205364H, RPT206230H, RPT206334H, RPT206385H, RPT206976H, RPT207522H, RPT208545H, RPT208903H, RPT208906H, RPT208915H, RPT209293H, RPT209676H, RPT210254H, RPT210255H, RPT210532H, RPT210741H, RPT210809H, RPT211032H, RPT211035H, RPT205146H, RPT205305H, RPT205784H, RPT206448H, RPT206681H, RPT206777H, RPT206802H, RPT208470H, RPT209142H, RPT209337H, RPT210051H, RPT211377H, RPT211604H, RPT205145H, RPT209016H, RPT210282H, RPT210285H, RPT205369H, RPT206172H, RPT206278H, RPT206772H, RPT208248H, RPT208405H, RPT208707H, RPT209515H, RPT209587H, RPT210824H, RPT211039H, RPT211110H, RPT211489H, RPT211605H, RPT205266H, RPT208558H, RPT206676H, RPT204489H, RPT210845H, RPT204392H, RPT204589H, RPT204623H, RPT204638H, RPT205654H, RPT206658H, RPT206664H, RPT206666H, RPT207760H, RPT207761H, RPT207762H, RPT207763H, RPT209002H, RPT209004H, RPT209014H, RPT209023H, RPT209420H, RPT209430H, RPT209434H, RPT209457H, RPT205685H, RPT207304H, RPT208164H, RPT205392H, RPT206177H, RPT207338H, RPT207747H, RPT208247H, RPT209537H, RPT210805H, RPT208813H, RPT209285H, RPT210577H, RPT210578H, RPT204515H, RPT204726H, RPT206450H, RPT206533H, RPT207438H, RPT209572H, RPT209905H, RPT207529H, RPT204043H, RPT204860H, RPT205662H, RPT208007H, RPT208009H, RPT208638H, RPT209400H, RPT210870H, RPT205852H, RPT205935H, RPT205936H, RPT205939H, RPT205943H, RPT206796H, RPT206798H, RPT206799H, RPT206800H, RPT206807H, RPT206814H, RPT207836H, RPT207837H, RPT207838H, RPT208838H, RPT208961H, RPT209268H, RPT209619H, RPT206700H, RPT208088H, RPT204721H, RPT204737H, RPT207014H, RPT207015H, RPT208712H, RPT208713H, RPT210280H, RPT210281H, RPT211186H, RPT211187H, RPT211188H, RPT209672H, RPT204699H, RPT205123H, RPT205402H, RPT205873H, RPT206128H, RPT205863H, RPT206274H, RPT211137H, RPT203662H, RPT204188H, RPT204421H, RPT204882H, RPT205454H, RPT206091H, RPT206106H, RPT207587H, RPT208954H, RPT209100H, RPT210137H, RPT210161H, RPT210288H, RPT204812H, RPT208677H, RPT204042H, RPT205723H, RPT207300H, RPT203684H, RPT206793H, RPT206939H, RPT210564H, RPT210570H, RPT211335H, RPT211716H, RPT204436H, RPT204493H, RPT204495H, RPT205627H, RPT206764H, RPT206765H, RPT209347H, RPT209348H, RPT206447H, RPT209178H, RPT210867H, RPT204074H, RPT207657H, RPT204809H, RPT208524H, RPT209067H, RPT209328H, RPT205229H, RPT206473H, RPT206556H, RPT208066H, RPT206235H, RPT207004H, RPT204630H, RPT206883H, RPT206323H, RPT204127H, RPT204710H, RPT204761H, RPT206013H, RPT206684H, RPT207947H, RPT208090H, RPT208456H, RPT208612H, RPT208640H, RPT208904H, RPT209635H, RPT209648H, RPT210205H, RPT210272H, RPT210376H, RPT211465H, RPT211214H, RPT211037H, RPT208288H, RPT209022H, RPT204972H, RPT206184H, RPT207070H, RPT204192H, RPT204193H, RPT205078H, RPT205691H, RPT208072H, RPT209840H, RPT209919H, RPT210517H, RPT210945H, RPT208726H, RPT203877H, RPT206271H, RPT204791H, RPT206006H, RPT206503H, RPT208749H, RPT208963H, RPT204002H, RPT204920H, RPT204925H, RPT205717H, RPT206918H, RPT207887H, RPT209130H, RPT206954H, RPT208586H, RPT209492H, RPT210286H, RPT203638H, RPT204871H, RPT206039H, RPT208431H, RPT209088H, RPT208861H, RPT203652H, RPT203803H, RPT204702H, RPT204865H, RPT206464H, RPT207928H, RPT208553H, RPT208951H, RPT209338H, RPT209636H, RPT210209H, RPT210941H, RPT211602H, RPT211609H, RPT206308H, RPT206638H, RPT208546H, RPT210129H, RPT204465H, RPT205308H, RPT205405H, RPT206260H, RPT207538H, RPT208747H, RPT209898H, RPT210856H, RPT211069H, RPT211351H, RPT211394H, RPT205211H, RPT211449H, RPT209856H, RPT206333H, RPT208477H, RPT208957H, RPT210099H, RPT210132H, RPT203701H, RPT203946H, RPT208206H, RPT208867H, RPT209629H, RPT210937H, RPT211116H, RPT211129H, RPT211296H, RPT211707H, RPT203745H, RPT203831H, RPT203908H, RPT210678H, RPT211543H, RPT208781H, RPT209801H, RPT210565H, RPT208927H, RPT210654H, RPT210675H, RPT211472H, RPT204835H, RPT206571H, RPT207115H, RPT207628H, RPT206630H, RPT206988H, RPT207150H, RPT207535H, RPT211594H, RPT204749H, RPT209595H, RPT210122H, RPT203913H, RPT204205H, RPT205046H, RPT205047H, RPT206459H, RPT206465H, RPT204329H, RPT208776H, RPT209065H, RPT204912H, RPT206094H, RPT209003H, RPT207111H, RPT203646H, RPT203657H, RPT203687H, RPT203727H, RPT203887H, RPT204277H, RPT204361H, RPT204469H, RPT204757H, RPT204777H, RPT204903H, RPT204958H, RPT205209H, RPT205630H, RPT206015H, RPT207423H, RPT208000H, RPT208214H, RPT208300H, RPT208379H, RPT208651H, RPT208845H, RPT209159H, RPT209164H, RPT209309H, RPT209368H, RPT209461H, RPT209805H, RPT209880H, RPT210061H, RPT210146H, RPT210158H, RPT210169H, RPT210177H, RPT210401H, RPT210461H, RPT211048H, RPT211189H, RPT211522H, RPT203781H, RPT204782H, RPT204893H, RPT205270H, RPT205531H, RPT205566H, RPT205577H, RPT206035H, RPT206291H, RPT206345H, RPT206628H, RPT207359H, RPT207381H, RPT208354H, RPT209495H, RPT209593H, RPT209863H, RPT210744H, RPT210930H, RPT211301H, RPT211302H, RPT211313H, RPT211694H, RPT206781H, RPT209025H, RPT209445H, RPT210400H, RPT210424H, RPT204252H, RPT204863H, RPT204881H, RPT206407H, RPT204345H, RPT207292H, RPT207739H, RPT205036H, RPT205429H, RPT207189H, RPT207985H, RPT208531H, RPT208585H, RPT210764H, RPT211546H, RPT211551H, RPT205030H, RPT205160H, RPT205161H, RPT205224H, RPT205247H, RPT205425H, RPT205605H, RPT207032H, RPT207363H, RPT207420H, RPT207984H, RPT208041H, RPT208306H, RPT209273H, RPT210769H, RPT210989H, RPT210990H, RPT211329H, RPT211451H, RPT204132H, RPT204011H, RPT204833H, RPT205282H, RPT205313H, RPT206053H, RPT206343H, RPT206398H, RPT207660H, RPT208360H, RPT209175H, RPT209215H, RPT209924H, RPT210094H, RPT211509H, RPT207589H, RPT209083H, RPT209091H, RPT209093H, RPT211357H, RPT211622H, RPT211624H, RPT211630H, RPT211631H, RPT211637H, RPT204424H, RPT204825H, RPT204964H, RPT203958H, RPT204256H, RPT204410H, RPT204821H, RPT206880H, RPT206946H, RPT208161H, RPT208860H, RPT209704H, RPT209791H, RPT209986H, RPT210166H, RPT205185H, RPT207025H, RPT208779H, RPT210899H, RPT208681H, RPT208829H, RPT208897H, RPT210384H, RPT210858H, RPT210938H, RPT211590H, RPT204672H, RPT205568H, RPT206697H, RPT209920H, RPT210064H, RPT204294H, RPT204532H, RPT206499H, RPT207129H, RPT203898H, RPT204456H, RPT204775H, RPT205231H, RPT206261H, RPT208344H, RPT210539H, RPT203797H, RPT205876H, RPT208670H, RPT207167H, RPT209854H, RPT210296H, RPT210898H, RPT211117H, RPT205663H, RPT205664H, RPT205665H, RPT205666H, RPT205668H, RPT206606H, RPT206920H, RPT206921H, RPT206923H, RPT206924H, RPT206925H, RPT206927H, RPT206928H, RPT206934H, RPT207209H, RPT207352H, RPT207353H, RPT207580H, RPT207666H, RPT207667H, RPT207671H, RPT207827H, RPT207872H, RPT207876H, RPT207938H, RPT207975H, RPT210151H, RPT210191H, RPT210194H, RPT211072H, RPT211073H, RPT211076H, RPT204101H, RPT205242H, RPT205504H, RPT205897H, RPT206869H, RPT210454H, RPT210608H, RPT210866H, RPT211246H, RPT208425H, RPT209616H, RPT204442H, RPT206830H, RPT208416H, RPT208964H, RPT209227H, RPT210554H, RPT203647H, RPT204855H, RPT210248H, RPT211597H, RPT205761H, RPT207596H, RPT208499H, RPT208500H, RPT208709H, RPT205457H, RPT208871H, RPT211007H, RPT203894H, RPT204327H, RPT205057H, RPT205678H, RPT205893H, RPT207159H, RPT207289H, RPT208030H, RPT208084H, RPT208402H, RPT208582H, RPT209527H, RPT211378H, RPT203707H, RPT203714H, RPT205113H, RPT205134H, RPT205404H, RPT205414H, RPT205888H, RPT205975H, RPT206622H, RPT206885H, RPT207169H, RPT207172H, RPT207195H, RPT207752H, RPT208873H, RPT209877H, RPT210382H, RPT205032H, RPT206266H, RPT206968H, RPT211415H, RPT211516H, RPT208091H, RPT208410H, RPT209756H, RPT210355H, RPT210526H, RPT210658H, RPT210988H, RPT207301H, RPT206047H, RPT209053H, RPT209708H, RPT210957H, RPT211676H, RPT205572H, RPT210292H, RPT210303H, RPT210324H, RPT210861H, RPT211536H, RPT208128H, RPT209736H, RPT211326H, RPT204924H, RPT205262H, RPT207133H, RPT210487H, RPT205385H, RPT211303H, RPT209647H, RPT210618H, RPT208063H, RPT208375H, RPT209866H, RPT210131H, RPT210263H, RPT203718H, RPT204432H, RPT206205H, RPT206701H, RPT207230H, RPT207569H, RPT208842H, RPT210950H, RPT210999H, RPT211024H, RPT204827H, RPT207599H, RPT205063H, RPT205064H, RPT205124H, RPT205230H, RPT208602H, RPT211232H, RPT207340H, RPT208052H, RPT208778H, RPT208890H, RPT209524H, RPT210912H, RPT211322H, RPT203907H, RPT204762H, RPT204802H, RPT205670H, RPT206862H, RPT208750H, RPT208931H, RPT208977H, RPT209316H, RPT209888H, RPT209989H, RPT210201H, RPT210331H, RPT210366H, RPT210493H, RPT210986H, RPT211265H, RPT211430H, RPT211457H, RPT211635H, RPT211658H, RPT205776H, RPT210602H, RPT204373H, RPT204838H, RPT205126H, RPT205919H, RPT207110H, RPT207295H, RPT207544H, RPT208092H, RPT208418H, RPT208527H, RPT210914H, RPT211405H, RPT203715H, RPT206522H, RPT203984H, RPT204931H, RPT205204H, RPT206140H, RPT206406H, RPT207107H, RPT209857H, RPT206071H, RPT204935H, RPT205831H, RPT205834H, RPT205837H, RPT206905H, RPT206979H, RPT206984H, RPT208743H, RPT208744H, RPT208851H, RPT208852H, RPT209459H, RPT209463H, RPT209465H, RPT209468H, RPT209470H, RPT210159H, RPT210220H, RPT210222H, RPT211220H, RPT211223H, RPT211224H, RPT211229H, RPT203799H, RPT207904H, RPT204330H, RPT204884H, RPT206213H, RPT206241H, RPT206661H, RPT207314H, RPT207436H, RPT207437H, RPT207748H, RPT208741H, RPT208846H, RPT209611H, RPT209788H, RPT210005H, RPT211628H, RPT207954H, RPT211400H, RPT204308H, RPT204395H, RPT204633H, RPT204634H, RPT205053H, RPT205409H, RPT205508H, RPT205509H, RPT210405H, RPT210587H, RPT210620H, RPT211017H, RPT211067H, RPT211310H, RPT211714H, RPT211715H, RPT209060H, RPT210320H, RPT207247H, RPT208603H, RPT204021H, RPT207173H, RPT210514H, RPT208175H, RPT210728H, RPT206786H, RPT210980H, RPT203980H, RPT204301H, RPT204581H, RPT205008H, RPT206123H, RPT210471H, RPT206454H, RPT208207H, RPT208895H, RPT211063H, RPT205274H, RPT205465H, RPT208318H, RPT208480H, RPT208857H, RPT210072H, RPT210693H, RPT211367H, RPT204792H, RPT211376H, RPT207987H, RPT209570H, RPT209952H, RPT205312H, RPT206212H, RPT207023H, RPT207986H, RPT208784H, RPT209029H, RPT210339H, RPT207315H, RPT206626H, RPT206640H, RPT206648H, RPT207533H, RPT209131H, RPT209372H, RPT207477H, RPT207734H, RPT208215H, RPT208885H, RPT209272H, RPT209386H, RPT209581H, RPT209747H, RPT211122H, RPT211276H, RPT211455H, RPT211686H, RPT211704H, RPT203850H, RPT204032H, RPT204502H, RPT204824H, RPT205397H, RPT206198H, RPT206433H, RPT208244H, RPT208294H, RPT208653H, RPT209166H, RPT209600H, RPT209930H, RPT210065H, RPT210323H, RPT210806H, RPT210954H, RPT211283H, RPT211410H, RPT211480H, RPT211507H, RPT211691H, RPT209776H, RPT211467H, RPT203813H, RPT205168H, RPT205985H, RPT205989H, RPT208666H, RPT208921H, RPT209606H, RPT210975H, RPT211350H, RPT211550H, RPT204034H, RPT208120H, RPT209340H, RPT209596H, RPT209768H, RPT204078H, RPT204377H, RPT204406H, RPT204409H, RPT205393H, RPT205394H, RPT206738H, RPT206742H, RPT207608H, RPT207610H, RPT208541H, RPT208544H, RPT209410H, RPT210421H, RPT210765H, RPT205203H, RPT205384H, RPT208382H, RPT208956H, RPT210245H, RPT204457H, RPT206650H, RPT206871H, RPT207027H, RPT209397H, RPT209401H, RPT211231H, RPT211235H, RPT211236H, RPT211237H, RPT206852H, RPT208349H, RPT208371H, RPT208449H, RPT211639H, RPT204829H, RPT211636H, RPT210963H, RPT205181H, RPT207104H, RPT209914H, RPT210313H, RPT211153H, RPT203807H, RPT204813H, RPT205569H, RPT208158H, RPT210606H, RPT210619H, RPT203867H, RPT203869H, RPT203872H, RPT203882H, RPT203884H, RPT204939H, RPT204943H, RPT205179H, RPT206854H, RPT206855H, RPT206857H, RPT207328H, RPT207337H, RPT207508H, RPT208401H, RPT209222H, RPT211240H, RPT211243H, RPT211244H, RPT211248H, RPT211253H, RPT211348H, RPT209865H, RPT210021H, RPT211464H, RPT210607H, RPT204407H, RPT206020H, RPT207035H, RPT207387H, RPT207441H, RPT208937H, RPT211570H, RPT206038H, RPT210484H, RPT206221H, RPT206545H, RPT208756H, RPT211439H, RPT211528H, RPT211555H, RPT207999H, RPT209875H, RPT210580H, RPT209455H, RPT211058H, RPT207331H, RPT203650H, RPT204344H, RPT204806H, RPT208979H, RPT209360H, RPT209484H, RPT209617H, RPT210891H, RPT205222H, RPT205374H, RPT205556H, RPT205576H, RPT205588H, RPT205598H, RPT207086H, RPT207153H, RPT208073H, RPT209847H, RPT210379H, RPT210626H, RPT203749H, RPT204448H, RPT204728H, RPT204730H, RPT205188H, RPT205221H, RPT205455H, RPT205530H, RPT206249H, RPT206585H, RPT207629H, RPT207925H, RPT208129H, RPT208523H, RPT208537H, RPT208753H, RPT208987H, RPT209125H, RPT209208H, RPT209244H, RPT209407H, RPT209421H, RPT209526H, RPT209534H, RPT209539H, RPT209970H, RPT210017H, RPT210149H, RPT210252H, RPT210256H, RPT210403H, RPT210756H, RPT210766H, RPT210803H, RPT210840H, RPT211198H, RPT211264H, RPT211380H, RPT205582H, RPT204693H, RPT206065H, RPT206158H, RPT210998H, RPT207639H, RPT208234H, RPT211482H, RPT205564H, RPT207255H, RPT210749H, RPT205092H, RPT209260H, RPT204091H, RPT205286H, RPT207609H, RPT204199H, RPT205748H, RPT206143H, RPT207265H, RPT211345H, RPT205151H, RPT205403H, RPT206197H, RPT207188H, RPT207269H, RPT207274H, RPT207370H, RPT208174H, RPT208365H, RPT209214H, RPT209389H, RPT210652H, RPT211010H, RPT211131H, RPT211341H, RPT211403H, RPT211443H, RPT211508H, RPT211512H, RPT211569H, RPT207207H, RPT205599H, RPT209669H, RPT204805H, RPT204807H, RPT206132H, RPT206219H, RPT206580H, RPT206759H, RPT206760H, RPT208702H, RPT208703H, RPT209310H, RPT209396H, RPT211082H, RPT211083H, RPT204149H, RPT204154H, RPT204163H, RPT204237H, RPT204376H, RPT204481H, RPT205070H, RPT205071H, RPT205159H, RPT205291H, RPT205898H, RPT205952H, RPT206199H, RPT206536H, RPT206849H, RPT207084H, RPT207462H, RPT207511H, RPT207541H, RPT208351H, RPT210085H, RPT210759H, RPT209662H, RPT203918H, RPT204259H, RPT204282H, RPT206319H, RPT208121H, RPT208224H, RPT208642H, RPT209956H, RPT204368H, RPT207085H, RPT208737H, RPT204753H, RPT209807H, RPT203849H, RPT210318H, RPT210550H, RPT204866H, RPT204894H, RPT205152H, RPT205725H, RPT207520H, RPT208377H, RPT208780H, RPT209808H, RPT205382H, RPT205826H, RPT205828H, RPT209466H, RPT211025H, RPT204279H, RPT204969H, RPT204999H, RPT209192H, RPT209200H, RPT210380H, RPT206994H, RPT207050H, RPT203886H, RPT203944H, RPT204020H, RPT204732H, RPT204816H, RPT205311H, RPT205442H, RPT207065H, RPT208107H, RPT208615H, RPT209231H, RPT209963H, RPT211041H, RPT205524H, RPT207543H, RPT203703H, RPT203895H, RPT203929H, RPT204012H, RPT204885H, RPT206500H, RPT208094H, RPT208268H, RPT209671H, RPT210511H, RPT211470H, RPT203760H, RPT204840H, RPT204842H, RPT204874H, RPT205688H, RPT205693H, RPT206938H, RPT207009H, RPT207161H, RPT207311H, RPT208098H, RPT208201H, RPT208204H, RPT208406H, RPT210571H, RPT210807H, RPT210813H, RPT211319H, RPT211441H, RPT203676H, RPT207096H, RPT207995H, RPT208929H, RPT203696H, RPT206200H, RPT208134H, RPT208420H, RPT210321H, RPT210503H, RPT211404H, RPT207916H, RPT207137H, RPT207362H, RPT207379H, RPT207213H, RPT204876H, RPT207355H, RPT208742H, RPT210919H, RPT210944H, RPT211607H, RPT203988H, RPT204179H, RPT205174H, RPT206654H, RPT208384H, RPT208540H, RPT209077H, RPT210960H, RPT206392H, RPT203959H, RPT207130H, RPT211081H, RPT207007H, RPT208947H, RPT210045H, RPT210949H, RPT211309H, RPT204164H, RPT205149H, RPT205218H, RPT205422H, RPT209405H, RPT209820H, RPT210742H, RPT210880H, RPT203680H, RPT203848H, RPT205258H, RPT205336H, RPT206046H, RPT206401H, RPT206443H, RPT207058H, RPT207547H, RPT208868H, RPT209645H, RPT209816H, RPT210024H, RPT204831H, RPT204423H, RPT209991H, RPT210271H, RPT210689H, RPT204612H, RPT204635H, RPT205370H, RPT211159H, RPT211533H, RPT211292H, RPT204586H, RPT205091H, RPT205750H, RPT205963H, RPT209773H, RPT211666H, RPT209489H, RPT210056H, RPT203819H, RPT205460H, RPT205516H, RPT205718H, RPT207367H, RPT208125H, RPT209582H, RPT210697H, RPT211166H, RPT204411H, RPT205079H, RPT205647H, RPT206088H, RPT208144H, RPT208484H, RPT209758H, RPT210947H, RPT204476H, RPT205641H, RPT206445H, RPT210465H, RPT207344H, RPT207478H, RPT209613H, RPT209614H, RPT210773H, RPT210776H, RPT210777H, RPT205283H, RPT204488H, RPT211427H, RPT210360H, RPT210594H, RPT211361H, RPT204652H, RPT204654H, RPT204655H, RPT204656H, RPT204660H, RPT204662H, RPT204664H, RPT204665H, RPT206601H, RPT207810H, RPT207813H, RPT207816H, RPT207825H, RPT207828H, RPT208557H, RPT208559H, RPT208561H, RPT208564H, RPT208568H, RPT209080H, RPT209775H, RPT209873H, RPT210513H, RPT203992H, RPT205395H, RPT205398H, RPT206790H, RPT207327H, RPT208005H, RPT209301H, RPT209769H, RPT203665H, RPT203948H, RPT204688H, RPT205842H, RPT206941H, RPT207422H, RPT208077H, RPT208892H, RPT209043H, RPT209476H, RPT207983H, RPT209713H, RPT210346H, RPT211399H, RPT206609H, RPT206747H, RPT210347H, RPT206526H, RPT209741H, RPT209839H, RPT205228H, RPT207171H, RPT209823H, RPT211115H, RPT211209H, RPT204480H, RPT205746H, RPT208469H, RPT209862H, RPT210031H, RPT210088H, RPT211119H, RPT204201H, RPT204371H, RPT205085H, RPT205300H, RPT205483H, RPT205613H, RPT205770H, RPT205849H, RPT205867H, RPT206297H, RPT206351H, RPT206402H, RPT206550H, RPT206623H, RPT207109H, RPT207643H, RPT208080H, RPT208082H, RPT208097H, RPT209300H, RPT209345H, RPT209751H, RPT209752H, RPT209826H, RPT209929H, RPT210023H, RPT210265H, RPT210422H, RPT210757H, RPT210761H, RPT207393H, RPT207411H, RPT209109H, RPT209981H, RPT206704H, RPT208667H, RPT209059H, RPT209496H, RPT209564H, RPT211365H, RPT211118H, RPT203876H, RPT204303H, RPT204477H, RPT204856H, RPT204985H, RPT205019H, RPT205413H, RPT205906H, RPT205914H, RPT205986H, RPT206289H, RPT206440H, RPT206492H, RPT206789H, RPT206801H, RPT207271H, RPT207277H, RPT207501H, RPT207505H, RPT208337H, RPT208519H, RPT208549H, RPT208583H, RPT208665H, RPT209148H, RPT209423H, RPT209944H, RPT210337H, RPT210397H, RPT210497H, RPT211167H, RPT211447H, RPT211498H, RPT211661H, RPT208858H, RPT205158H, RPT205832H, RPT206277H, RPT206303H, RPT209015H, RPT209330H, RPT210408H, RPT210659H, RPT205400H, RPT205681H, RPT207322H, RPT210329H, RPT211047H, RPT204682H, RPT204686H, RPT205731H, RPT207576H, RPT209289H, RPT210012H, RPT210336H, RPT210653H, RPT211042H, RPT204243H, RPT205389H, RPT205686H, RPT206724H, RPT208137H, RPT208923H, RPT208996H, RPT207400H, RPT209640H, RPT204391H, RPT206733H, RPT207154H, RPT204125H, RPT204911H, RPT205265H, RPT205778H, RPT205779H, RPT205783H, RPT206912H, RPT206915H, RPT208146H, RPT208195H, RPT208332H, RPT208555H, RPT208724H, RPT208725H, RPT203833H, RPT205193H, RPT207459H, RPT207812H, RPT208914H, RPT208967H, RPT208998H, RPT209504H, RPT209506H, RPT210116H, RPT210328H, RPT210338H, RPT210359H, RPT210482H, RPT210876H, RPT210970H, RPT211282H, RPT209812H, RPT202723H, RPT203720H, RPT203737H, RPT203853H, RPT205003H, RPT205061H, RPT205137H, RPT205652H, RPT206710H, RPT207766H, RPT208045H, RPT208870H, RPT210426H, RPT210449H, RPT210682H, RPT211355H, RPT205492H, RPT206699H, RPT206173H, RPT207640H, RPT211665H, RPT203852H, RPT204245H, RPT206722H, RPT207433H, RPT207448H, RPT210616H, RPT211049H, RPT209651H, RPT204569H, RPT204570H, RPT204576H, RPT204578H, RPT204580H, RPT207526H, RPT207527H, RPT209113H, RPT210195H, RPT210196H, RPT210197H, RPT210198H, RPT210702H, RPT204389H, RPT205043H, RPT205586H, RPT205913H, RPT205947H, RPT205996H, RPT206131H, RPT207346H, RPT208157H, RPT208710H, RPT209326H, RPT209482H, RPT210274H, RPT211616H, RPT206711H, RPT206953H, RPT207383H, RPT207442H, RPT206340H, RPT207495H, RPT208486H, RPT210387H, RPT210900H, RPT211342H, RPT204210H, RPT205667H, RPT208047H, RPT208367H, RPT209620H, RPT205551H, RPT205553H, RPT209270H, RPT209598H, RPT209827H, RPT210312H, RPT210510H, RPT211277H, RPT211572H, RPT203669H, RPT208376H, RPT208427H, RPT209124H, RPT207059H, RPT207350H, RPT210559H, RPT209467H, RPT209980H, RPT203885H, RPT205478H, RPT206218H, RPT207963H, RPT208727H, RPT208847H, RPT209021H, RPT209622H, RPT209779H, RPT209906H, RPT210547H, RPT211387H, RPT205379H, RPT205587H, RPT206105H, RPT206548H, RPT207103H, RPT209196H, RPT210160H, RPT203977H, RPT204770H, RPT205227H, RPT205578H, RPT206017H, RPT206024H, RPT207944H, RPT208131H, RPT208981H, RPT209282H, RPT209505H, RPT209552H, RPT209583H, RPT209845H, RPT210785H, RPT211307H, RPT211584H, RPT209815H, RPT210180H, RPT203821H, RPT203823H, RPT204126H, RPT205022H, RPT205074H, RPT209561H, RPT205606H, RPT207903H, RPT206469H, RPT207977H, RPT208085H, RPT209177H, RPT209327H, RPT209706H, RPT209774H, RPT210074H, RPT210499H, RPT210512H, RPT210817H, RPT208505H, RPT204018H, RPT208672H, RPT210353H, RPT203901H, RPT204934H, RPT206678H, RPT208467H, RPT209161H, RPT211254H, RPT203783H, RPT207405H, RPT210495H, RPT211556H, RPT211286H, RPT205466H, RPT204970H, RPT206301H, RPT208690H, RPT208928H, RPT209295H, RPT210417H, RPT208276H, RPT208459H, RPT209136H, RPT209035H, RPT202720H, RPT204440H, RPT205067H, RPT205412H, RPT207347H, RPT209281H, RPT209284H, RPT210468H, RPT210549H, RPT211356H, RPT211690H, RPT207310H, RPT205051H, RPT205069H, RPT208221H, RPT208237H, RPT208327H, RPT208445H, RPT208538H, RPT208580H, RPT208887H, RPT209171H, RPT210500H, RPT211133H, RPT211425H, RPT211483H, RPT211485H, RPT206403H, RPT206856H, RPT208694H, RPT209711H, RPT209746H, RPT209772H, RPT209793H, RPT210463H, RPT211222H, RPT206228H, RPT206495H, RPT206152H, RPT206620H, RPT208502H, RPT209217H, RPT210540H, RPT203917H, RPT206859H, RPT208122H, RPT211453H, RPT203768H, RPT204454H, RPT204795H, RPT208668H, RPT210212H, RPT211036H, RPT207193H, RPT211360H, RPT205406H, RPT209836H, RPT209885H, RPT210521H, RPT210889H, RPT204333H, RPT206698H, RPT208105H, RPT208212H, RPT208439H, RPT209325H, RPT209953H, RPT210633H, RPT205951H, RPT205974H, RPT208342H, RPT209052H, RPT210171H, RPT211318H, RPT203966H, RPT205648H, RPT208093H, RPT208096H, RPT209239H, RPT209262H, RPT209393H, RPT211379H, RPT203860H, RPT203874H, RPT206253H, RPT206615H, RPT206850H, RPT206851H, RPT207864H, RPT208754H, RPT204338H, RPT205976H, RPT206003H, RPT208761H, RPT208762H, RPT209243H, RPT210270H, RPT210273H, RPT204891H, RPT204892H, RPT204896H, RPT209571H, RPT209680H, RPT209683H, RPT211132H, RPT211233H, RPT211606H, RPT205073H, RPT207740H, RPT205477H, RPT207454H, RPT209332H, RPT207308H, RPT209852H, RPT210987H, RPT208475H, RPT211031H, RPT203987H, RPT205267H, RPT206894H, RPT207254H, RPT205484H, RPT210276H, RPT203897H, RPT203905H, RPT207425H, RPT210375H, RPT210520H, RPT210874H, RPT211468H, RPT204313H, RPT208369H, RPT208770H, RPT211271H, RPT206113H, RPT207022H, RPT207095H, RPT207413H, RPT208033H, RPT208788H, RPT208789H, RPT209778H, RPT210037H, RPT210535H, RPT210740H, RPT204610H, RPT208229H, RPT209654H, RPT211087H, RPT211098H, RPT211448H, RPT203754H, RPT204346H, RPT205288H, RPT205362H, RPT206194H, RPT207506H, RPT211018H, RPT204094H, RPT209172H, RPT211107H, RPT210896H, RPT204413H, RPT208230H, RPT207270H, RPT208312H, RPT210152H, RPT203637H, RPT204401H, RPT204952H, RPT208886H, RPT208970H, RPT209061H, RPT209137H, RPT211006H, RPT211359H, RPT211437H, RPT204209H, RPT203932H, RPT205130H, RPT205485H, RPT206563H, RPT207057H, RPT209185H, RPT209247H, RPT209250H, RPT203664H, RPT203757H, RPT203864H, RPT204103H, RPT205254H, RPT205292H, RPT205349H, RPT205438H, RPT205439H, RPT205608H, RPT205946H, RPT205973H, RPT206174H, RPT206397H, RPT206669H, RPT207898H, RPT208153H, RPT208978H, RPT209493H, RPT209717H, RPT210407H, RPT210609H, RPT204241H, RPT204388H, RPT205711H, RPT206307H, RPT206534H, RPT207002H, RPT207063H, RPT209321H, RPT210354H, RPT210625H, RPT210985H, RPT211014H, RPT204181H, RPT204183H, RPT204185H, RPT204628H, RPT204629H, RPT205472H, RPT205475H, RPT205476H, RPT207415H, RPT207417H, RPT207419H, RPT209191H, RPT210013H, RPT210040H, RPT210351H, RPT210664H, RPT210666H, RPT210828H, RPT206063H, RPT208387H, RPT204674H, RPT205172H, RPT205499H, RPT205781H, RPT206036H, RPT206259H, RPT206262H, RPT206279H, RPT206530H, RPT207325H, RPT208103H, RPT209248H, RPT209304H, RPT209311H, RPT209331H, RPT209925H, RPT209928H, RPT209959H, RPT204310H, RPT206022H, RPT206154H, RPT206463H, RPT206562H, RPT208919H, RPT209162H, RPT211548H, RPT204397H, RPT203854H, RPT204014H, RPT204059H, RPT204145H, RPT204263H, RPT204341H, RPT204711H, RPT205059H, RPT205219H, RPT206021H, RPT206566H, RPT207275H, RPT207336H, RPT207348H, RPT207489H, RPT207513H, RPT208151H, RPT208526H, RPT208701H, RPT208797H, RPT208983H, RPT209010H, RPT210069H, RPT210087H, RPT210294H, RPT210501H, RPT210646H, RPT210713H, RPT203686H, RPT204064H, RPT204217H, RPT204244H, RPT205281H, RPT206002H, RPT208099H, RPT208227H, RPT208566H, RPT209070H, RPT209806H, RPT204818H, RPT205786H, RPT205787H, RPT205792H, RPT205927H, RPT206806H, RPT207083H, RPT207235H, RPT207237H, RPT207239H, RPT207240H, RPT207241H, RPT208219H, RPT208223H, RPT208646H, RPT208980H, RPT209545H, RPT210192H, RPT210203H, RPT210295H, RPT210628H, RPT210738H, RPT211375H, RPT204956H, RPT204967H, RPT205183H, RPT207318H, RPT207408H, RPT208340H, RPT208917H, RPT209580H, RPT209782H, RPT210553H, RPT210841H, RPT210844H, RPT211501H, RPT204148H, RPT204200H, RPT204238H, RPT206138H, RPT206510H, RPT208393H, RPT209749H, RPT210716H, RPT211576H, RPT204909H, RPT205672H, RPT205673H, RPT206694H, RPT206712H, RPT206720H, RPT206723H, RPT206726H, RPT206731H, RPT209794H, RPT209861H, RPT209864H, RPT209887H, RPT209889H, RPT209890H, RPT210486H, RPT210821H, RPT210826H, RPT210977H, RPT211089H, RPT203748H, RPT207536H, RPT208700H, RPT206226H, RPT208143H, RPT210437H, RPT211653H, RPT205464H, RPT206031H, RPT208036H, RPT209544H, RPT209685H, RPT204077H, RPT205432H, RPT205631H, RPT207603H, RPT209797H, RPT209931H, RPT211371H, RPT204028H, RPT204268H, RPT204846H, RPT205528H, RPT206992H, RPT207873H, RPT208571H, RPT208759H, RPT210025H, RPT210651H, RPT210767H, RPT211108H, RPT204567H, RPT205890H, RPT203725H, RPT205002H, RPT205550H, RPT205785H, RPT206206H, RPT206947H, RPT207166H, RPT208775H, RPT210115H, RPT204273H, RPT204826H, RPT204851H, RPT205875H, RPT206498H, RPT207091H, RPT207170H, RPT207626H, RPT207653H, RPT208267H, RPT208518H, RPT208592H, RPT209592H, RPT209731H, RPT210669H, RPT209516H, RPT209555H, RPT204366H, RPT208119H, RPT210720H, RPT205324H, RPT206863H, RPT208438H, RPT208853H, RPT210531H, RPT204153H, RPT206841H, RPT207991H, RPT208686H, RPT209149H, RPT209395H, RPT209443H, RPT209958H, RPT210246H, RPT210247H, RPT210940H, RPT211091H, RPT211092H, RPT211095H, RPT211096H, RPT211100H, RPT211101H, RPT211103H, RPT211109H, RPT211471H, RPT203926H, RPT204079H, RPT204190H, RPT204202H, RPT206834H, RPT207040H, RPT207896H, RPT208536H, RPT208833H, RPT209151H, RPT209335H, RPT205098H, RPT205184H, RPT205501H, RPT205754H, RPT205886H, RPT207163H, RPT203695H, RPT204321H, RPT206680H, RPT210126H, RPT210368H, RPT203904H, RPT203936H, RPT204435H, RPT204764H, RPT205407H, RPT205734H, RPT205982H, RPT206811H, RPT206812H, RPT207250H, RPT207874H, RPT207879H, RPT207883H, RPT207884H, RPT209103H, RPT209107H, RPT209111H, RPT209114H, RPT209122H, RPT209123H, RPT210157H, RPT210162H, RPT210167H, RPT210168H, RPT210175H, RPT210178H, RPT210183H, RPT210185H, RPT210186H, RPT211288H, RPT204072H, RPT204105H, RPT204206H, RPT204946H, RPT205153H, RPT208357H, RPT208479H, RPT209232H, RPT209249H, RPT205018H, RPT206732H, RPT206735H, RPT211298H, RPT211097H, RPT203896H, RPT205418H, RPT207965H, RPT208808H, RPT204431H, RPT204781H, RPT205570H, RPT206521H, RPT208126H, RPT208618H, RPT211090H, RPT203940H, RPT204234H, RPT204449H, RPT205109H, RPT205995H, RPT206073H, RPT206986H, RPT207959H, RPT209923H, RPT210269H, RPT210545H, RPT210655H, RPT210847H, RPT205431H, RPT210848H, RPT204278H, RPT204845H, RPT205259H, RPT205451H, RPT206776H, RPT208419H, RPT210763H, RPT204928H, RPT203788H, RPT204729H, RPT204800H, RPT205094H, RPT205363H, RPT205479H, RPT206632H, RPT206717H, RPT206845H, RPT207390H, RPT207604H, RPT207627H, RPT208051H, RPT208639H, RPT208662H, RPT209546H, RPT209547H, RPT210864H, RPT204257H, RPT204425H, RPT204832H, RPT208935H, RPT206458H, RPT210780H, RPT210905H, RPT206714H, RPT207395H, RPT203996H, RPT204632H, RPT204673H, RPT206844H, RPT206889H, RPT206983H, RPT207122H, RPT209817H, RPT210133H, RPT204136H, RPT204478H, RPT205001H, RPT205119H, RPT205902H, RPT206025H, RPT207516H, RPT203830H, RPT206543H, RPT207614H, RPT208889H, RPT209983H, RPT211618H, RPT211667H, RPT207480H, RPT208497H, RPT209287H, RPT211577H, RPT205373H, RPT208156H, RPT206090H, RPT206409H, RPT207612H, RPT208609H, RPT209165H, RPT209802H, RPT204104H, RPT204283H, RPT204285H, RPT205175H, RPT206163H, RPT209047H, RPT211044H, RPT211466H, RPT203994H, RPT204115H, RPT204595H, RPT205299H, RPT205315H, RPT206815H, RPT206817H, RPT206818H, RPT206819H, RPT206820H, RPT206822H, RPT206824H, RPT210004H, RPT210699H, RPT208040H, RPT210585H, RPT203734H, RPT209588H, RPT205116H, RPT205629H, RPT209790H, RPT210207H, RPT204880H, RPT206929H, RPT208878H, RPT210086H, RPT211383H, RPT206391H, RPT206887H, RPT209819H, RPT206070H, RPT206122H, RPT208140H, RPT208464H, RPT209529H, RPT205789H, RPT206057H, RPT206549H, RPT206659H, RPT207950H, RPT208100H, RPT209460H, RPT209892H, RPT210967H, RPT211026H, RPT211513H, RPT204305H, RPT208095H, RPT208273H, RPT209754H, RPT205157H, RPT206180H, RPT203985H, RPT203995H, RPT204055H, RPT204218H, RPT204517H, RPT205348H, RPT205633H, RPT206107H, RPT206195H, RPT207134H, RPT208293H, RPT203654H, RPT204487H, RPT204923H, RPT205329H, RPT206066H, RPT208210H, RPT209008H, RPT209792H, RPT210084H, RPT210460H, RPT204117H, RPT204858H, RPT204864H, RPT211531H, RPT211064H, RPT204902H, RPT205055H, RPT205419H, RPT205593H, RPT207045H, RPT210014H, RPT211515H, RPT210768H, RPT205049H, RPT205276H, RPT207632H, RPT208266H, RPT209251H, RPT209860H, RPT209878H, RPT205314H, RPT205316H, RPT205319H, RPT209341H, RPT209342H, RPT211314H, RPT204320H, RPT204462H, RPT205005H, RPT205915H, RPT208433H, RPT208819H, RPT208949H, RPT209883H, RPT204731H, RPT204733H, RPT204555H, RPT207463H, RPT209605H, RPT210078H, RPT211270H, RPT211352H, RPT206037H, RPT207913H, RPT207915H, RPT209510H, RPT209666H, RPT209954H, RPT204166H, RPT211241H, RPT205261H, RPT205757H, RPT208664H, RPT204797H, RPT206275H, RPT203827H, RPT204961H, RPT205189H, RPT205452H, RPT205698H, RPT205763H, RPT205765H, RPT205767H, RPT205769H, RPT205773H, RPT205887H, RPT205969H, RPT205998H, RPT206034H, RPT206048H, RPT206187H, RPT206349H, RPT206370H, RPT206373H, RPT206381H, RPT206405H, RPT206523H, RPT206542H, RPT206544H, RPT206569H, RPT206594H, RPT206707H, RPT207396H, RPT207507H, RPT207509H, RPT207510H, RPT207617H, RPT208177H, RPT208629H, RPT208862H, RPT209387H, RPT209388H, RPT209679H, RPT210211H, RPT210213H, RPT210792H, RPT210925H, RPT211079H, RPT211084H, RPT211094H, RPT211099H, RPT211105H, RPT211262H, RPT211445H, RPT211504H, RPT211575H, RPT211656H, RPT204559H, RPT206050H, RPT206060H, RPT206061H, RPT206074H, RPT207672H, RPT207924H, RPT207940H, RPT207941H, RPT207942H, RPT209716H, RPT205353H, RPT208075H, RPT211050H, RPT210390H, RPT203840H, RPT206629H, RPT206491H, RPT208658H, RPT205704H, RPT208968H, RPT210819H, RPT207216H, RPT208396H, RPT209075H, RPT210799H, RPT205176H, RPT207416H, RPT207469H, RPT210746H, RPT211114H, RPT211560H, RPT208683H, RPT206660H, RPT208240H, RPT209824H, RPT210641H, RPT211291H, RPT211496H, RPT203661H, RPT203692H, RPT203873H, RPT204292H, RPT204500H, RPT207409H, RPT210181H, RPT208884H, RPT210946H, RPT210973H, RPT205352H, RPT204017H, RPT204081H, RPT204597H, RPT204598H, RPT204599H, RPT204600H, RPT204604H, RPT206100H, RPT207818H, RPT207821H, RPT208068H, RPT209255H, RPT209785H, RPT209881H, RPT209894H, RPT209899H, RPT210076H, RPT204645H, RPT205430H, RPT204178H, RPT204223H, RPT205330H, RPT205480H, RPT206462H, RPT207428H, RPT209494H, RPT209513H, RPT209871H, RPT203942H, RPT205864H, RPT208729H, RPT208768H, RPT210260H, RPT203839H, RPT203947H, RPT204359H, RPT204760H, RPT205883H, RPT205884H, RPT207664H, RPT210788H, RPT204189H, RPT205013H, RPT207615H, RPT207997H, RPT203847H, RPT204945H, RPT205035H, RPT207087H, RPT209313H, RPT209339H, RPT206110H, RPT208552H, RPT209901H, RPT210509H, RPT204707H, RPT204868H, RPT205719H, RPT205722H, RPT206111H, RPT206528H, RPT206908H, RPT207644H, RPT210822H, RPT204110H, RPT210969H, RPT203698H, RPT205448H, RPT206734H, RPT206842H, RPT207755H, RPT208633H, RPT209163H, RPT209350H, RPT209907H, RPT211164H, RPT206121H, RPT211456H, RPT204275H, RPT205934H, RPT206444H, RPT207523H, RPT208378H, RPT210528H, RPT211259H, RPT204386H, RPT206387H, RPT209968H, RPT206596H, RPT208542H, RPT210982H, RPT211354H, RPT210483H, RPT211045H, RPT204649H, RPT204897H, RPT205310H, RPT205987H, RPT206096H, RPT206294H, RPT206368H, RPT206917H, RPT207976H, RPT210434H, RPT211002H, RPT211019H, RPT204165H, RPT205021H, RPT208962H, RPT204211H, RPT205983H, RPT207465H, RPT207514H, RPT207559H, RPT208145H, RPT208217H, RPT208795H, RPT208909H, RPT209681H, RPT210038H, RPT210107H, RPT210530H, RPT211256H, RPT211598H, RPT204869H, RPT210752H, RPT204954H, RPT206124H, RPT207449H, RPT208767H, RPT209891H, RPT211022H, RPT211450H, RPT211497H, RPT211713H, RPT204418H, RPT204419H, RPT207494H, RPT208714H, RPT208715H, RPT208717H, RPT208718H, RPT208719H, RPT209967H, RPT209971H, RPT209973H, RPT211323H, RPT205753H, RPT206624H, RPT208517H, RPT208888H, RPT203934H, RPT205990H, RPT206265H, RPT207811H, RPT208339H, RPT208381H, RPT208721H, RPT208836H, RPT209253H, RPT209941H, RPT210398H, RPT210491H, RPT210965H, RPT211652H, RPT208411H, RPT210477H, RPT204240H, RPT207625H, RPT208628H, RPT208752H, RPT205517H, RPT208539H, RPT209682H, RPT210621H, RPT210701H, RPT210994H, RPT204428H, RPT205337H, RPT205940H, RPT206570H, RPT210165H, RPT210470H, RPT210831H, RPT211263H, RPT211411H, RPT203767H, RPT204857H, RPT205033H, RPT205226H, RPT205565H, RPT205589H, RPT206005H, RPT208117H, RPT210586H, RPT211070H, RPT207101H, RPT208171H, RPT211066H, RPT205638H, RPT203673H, RPT204883H, RPT205715H, RPT208818H, RPT210533H, RPT210928H, RPT206574H, RPT210036H, RPT210832H, RPT210860H, RPT211692H, RPT203815H, RPT203951H, RPT203990H, RPT206302H, RPT206456H, RPT206564H, RPT206987H, RPT208364H, RPT208805H, RPT208806H, RPT210804H, RPT204948H, RPT206633H, RPT211034H, RPT203780H, RPT203721H, RPT203758H, RPT208437H, RPT209519H, RPT207090H, RPT207342H, RPT207512H, RPT209765H, RPT209770H, RPT210130H, RPT207470H, RPT207957H, RPT208482H, RPT210435H, RPT210458H, RPT211038H, RPT206393H, RPT206400H, RPT206651H, RPT206673H, RPT207074H, RPT207443H, RPT208496H, RPT209011H, RPT209294H, RPT209351H, RPT210063H, RPT210184H, RPT210343H, RPT210745H, RPT211514H, RPT211695H, RPT210770H, RPT210812H, RPT211317H, RPT206537H, RPT208409H, RPT209317H, RPT207019H, RPT206367H, RPT203717H, RPT204990H, RPT205271H, RPT206559H, RPT207891H, RPT208169H, RPT208336H, RPT208392H, RPT208520H, RPT209567H, RPT210523H, RPT210537H, RPT210877H, RPT211369H, RPT203814H, RPT209734H, RPT211436H, RPT205502H, RPT205595H, RPT205791H, RPT206179H, RPT207158H, RPT208216H, RPT208323H, RPT209597H, RPT209825H, RPT209939H, RPT211052H, RPT211349H, RPT211526H, RPT203862H, RPT208900H, RPT209673H, RPT203659H, RPT203762H, RPT203818H, RPT204022H, RPT204033H, RPT204152H, RPT204354H, RPT204437H, RPT204766H, RPT205421H, RPT205645H, RPT205732H, RPT205972H, RPT206561H, RPT206639H, RPT206780H, RPT206792H, RPT206853H, RPT206990H, RPT207258H, RPT207306H, RPT208331H, RPT208495H, RPT208859H, RPT209032H, RPT209542H, RPT209943H, RPT210230H, RPT210262H, RPT210393H, RPT210464H, RPT210714H, RPT210797H, RPT210825H, RPT211433H, RPT211321H, RPT205323H, RPT206018H, RPT208608H, RPT209126H, RPT211390H, RPT211422H, RPT211429H, RPT211657H, RPT209634H, RPT204177H, RPT204247H, RPT204271H, RPT204398H, RPT207994H, RPT208911H, RPT208969H, RPT209390H, RPT211227H, RPT211336H, RPT211337H, RPT211611H, RPT203919H, RPT203964H, RPT204940H, RPT204965H, RPT205014H, RPT205751H, RPT206280H, RPT207011H, RPT208252H, RPT209207H, RPT209280H, RPT211705H, RPT203851H, RPT203941H, RPT205458H, RPT205879H, RPT204151H, RPT204718H, RPT204720H, RPT204722H, RPT204724H, RPT204725H, RPT206476H, RPT206481H, RPT206483H, RPT206485H, RPT206489H, RPT208673H, RPT208848H, RPT208849H, RPT209903H, RPT210299H, RPT210475H, RPT210476H, RPT210478H, RPT205585H, RPT206461H, RPT206552H, RPT211389H, RPT204267H, RPT207278H, RPT206551H, RPT205682H, RPT206383H, RPT208620H, RPT209194H, RPT210515H, RPT208113H, RPT208035H, RPT208046H, RPT208236H, RPT208429H, RPT209917H, RPT210027H, RPT210604H, RPT210611H, RPT211413H, RPT203933H, RPT204754H, RPT204872H, RPT205253H, RPT205759H, RPT208656H, RPT208659H, RPT208660H, RPT210176H, RPT204261H, RPT205170H, RPT210904H, RPT207655H, RPT208444H, RPT210373H, RPT210927H, RPT209876H, RPT211136H, RPT211561H, RPT209844H, RPT208577H, RPT208769H, RPT209488H, RPT210605H, RPT211506H, RPT203679H, RPT207020H, RPT209979H, RPT211004H, RPT211334H, RPT206557H, RPT209627H, RPT205424H, RPT205609H, RPT206607H, RPT206608H, RPT206611H, RPT210284H, RPT207152H, RPT208270H, RPT210588H, RPT205518H, RPT205519H, RPT206891H, RPT206892H, RPT208733H, RPT210029H, RPT210548H, RPT204470H, RPT205554H, RPT209771H, RPT205840H, RPT206998H, RPT208790H, RPT208844H, RPT208394H, RPT203730H, RPT205125H, RPT205497H, RPT206242H, RPT206897H, RPT206899H, RPT209446H, RPT210109H, RPT211681H, RPT203981H, RPT208825H, RPT208992H, RPT211614H, RPT206354H, RPT204486H, RPT209108H, RPT211617H, RPT205368H, RPT207333H, RPT209787H, RPT205696H, RPT207912H, RPT210075H, RPT204496H, RPT206665H, RPT206677H, RPT207097H, RPT207252H, RPT209012H, RPT209780H, RPT210062H, RPT211225H, RPT204706H, RPT205671H, RPT205824H, RPT207332H, RPT207444H, RPT207646H, RPT207648H, RPT208231H, RPT208986H, RPT208997H, RPT209001H, RPT209479H, RPT209480H, RPT209538H, RPT211179H, RPT211202H, RPT211549H, RPT211642H, RPT210283H, RPT203782H, RPT204123H, RPT204751H, RPT205917H, RPT207052H, RPT208015H, RPT208395H, RPT209641H, RPT211226H, RPT211416H, RPT207114H, RPT207249H, RPT207633H, RPT207637H, RPT207638H, RPT210881H, RPT210892H, RPT207432H, RPT207793H, RPT208675H, RPT209668H, RPT210385H, RPT208127H, RPT204771H, RPT209101H, RPT209102H, RPT210297H, RPT210943H, RPT211149H, RPT211150H, RPT211151H, RPT208278H, RPT208560H, RPT208631H, RPT209176H, RPT209427H, RPT209621H, RPT209993H, RPT210015H, RPT209098H, RPT209099H, RPT209322H, RPT209895H, RPT209900H, RPT209982H, RPT210147H, RPT210685H, RPT210778H, RPT211160H, RPT211162H, RPT211163H, RPT211165H, RPT211538H, RPT211659H, RPT211698H, RPT203923H, RPT205128H, RPT205532H, RPT206101H, RPT207450H, RPT207468H, RPT208913H, RPT209380H, RPT209414H, RPT210067H, RPT211003H, RPT211474H, RPT211535H, RPT208314H, RPT209063H, RPT203974H, RPT204088H, RPT204097H, RPT204142H, RPT204231H, RPT204379H, RPT204405H, RPT205239H, RPT205297H, RPT206872H, RPT207687H, RPT210425H, RPT210707H, RPT211391H, RPT211612H, RPT211684H, RPT206475H, RPT206516H, RPT211654H, RPT206708H, RPT203763H, RPT205416H, RPT205949H, RPT203708H, RPT204778H, RPT206089H, RPT207558H, RPT208533H, RPT209690H, RPT209757H, RPT210170H, RPT204108H, RPT204175H, RPT204180H, RPT204254H, RPT204472H, RPT204596H, RPT205469H, RPT205515H, RPT205623H, RPT205628H, RPT205771H, RPT206505H, RPT206507H, RPT207392H, RPT207988H, RPT208296H, RPT209233H, RPT209760H, RPT209874H, RPT210704H, RPT210966H, RPT203704H, RPT205060H, RPT209896H, RPT209933H, RPT210344H, RPT210350H, RPT204438H, RPT210567H, RPT211664H, RPT203911H, RPT204162H, RPT204548H, RPT205108H, RPT205111H, RPT205244H, RPT205955H, RPT206614H, RPT206965H, RPT206967H, RPT207008H, RPT208010H, RPT208116H, RPT208386H, RPT208408H, RPT209984H, RPT210489H, RPT210760H, RPT211228H, RPT211340H, RPT209750H, RPT203798H, RPT204471H, RPT204776H, RPT205356H, RPT205854H, RPT206304H, RPT207261H, RPT207546H, RPT208335H, RPT208971H, RPT209497H, RPT209697H, RPT209725H, RPT209320H, RPT204460H, RPT204750H, RPT207431H, RPT210070H, RPT210436H, RPT204396H, RPT209312H, RPT204246H, RPT207634H, RPT210718H, RPT204756H, RPT205023H, RPT205860H, RPT209922H, RPT210781H, RPT211106H, RPT205563H, RPT209859H, RPT210111H, RPT204550H, RPT204551H, RPT204552H, RPT204553H, RPT204554H, RPT204556H, RPT204557H, RPT204558H, RPT204560H, RPT204561H, RPT204562H, RPT206411H, RPT206412H, RPT206414H, RPT206415H, RPT206417H, RPT206418H, RPT206420H, RPT206421H, RPT206422H, RPT207869H, RPT207875H, RPT207880H, RPT207888H, RPT207893H, RPT207899H, RPT207901H, RPT207921H, RPT208355H, RPT208356H, RPT208358H, RPT208359H, RPT208361H, RPT208362H, RPT208939H, RPT205136H, RPT205339H, RPT208786H, RPT209748H, RPT210717H, RPT211130H, RPT204734H, RPT204941H, RPT206167H, RPT206350H, RPT209152H, RPT209188H, RPT208283H, RPT208414H, RPT204095H, RPT206404H, RPT206951H, RPT207018H, RPT211414H, RPT206160H, RPT206758H, RPT208548H, RPT209503H, RPT204363H, RPT205843H, RPT208773H, RPT208918H, RPT209073H, RPT209913H, RPT210438H, RPT210519H, RPT210623H, RPT210677H, RPT210871H, RPT211275H, RPT211521H, RPT205885H, RPT207806H, RPT209046H, RPT210481H, RPT206937H, RPT208290H, RPT204403H, RPT207280H, RPT208315H, RPT208443H, RPT208856H, RPT209210H, RPT209649H, RPT209661H, RPT209853H, RPT209961H, RPT210423H, RPT210574H, RPT210751H, RPT211001H, RPT203837H, RPT204443H, RPT206670H, RPT209578H, RPT204220H, RPT204287H, RPT209594H, RPT203689H, RPT206683H, RPT208148H, RPT209508H, RPT209500H, RPT204504H, RPT206155H, RPT206514H, RPT208407H, RPT209314H, RPT209411H, RPT209783H, RPT209784H, RPT210706H, RPT203671H, RPT209988H, RPT204862H, RPT209688H, RPT208877H, RPT208950H, RPT208985H, RPT209449H, RPT209652H, RPT204666H, RPT206472H, RPT208142H, RPT208865H, RPT210232H, RPT211494H, RPT211647H, RPT211660H, RPT211687H, RPT211693H, RPT211706H, RPT205037H, RPT209786H, RPT210613H, RPT203683H, RPT203775H, RPT203973H, RPT205213H, RPT206337H, RPT207451H, RPT207641H, RPT207642H, RPT210456H, RPT205024H, RPT206201H, RPT208373H, RPT208984H, RPT209799H, RPT209822H, RPT210485H, RPT210529H, RPT210786H, RPT211503H, RPT203845H, RPT203846H, RPT204506H, RPT206993H, RPT209189H, RPT209511H, RPT210305H, RPT211490H, RPT206296H, RPT207735H, RPT208260H, RPT208840H, RPT210859H, RPT204433H, RPT206388H, RPT209491H, RPT203938H, RPT204053H, RPT204226H, RPT206408H, RPT208064H, RPT211011H, RPT211579H, RPT208108H, RPT209315H, RPT209442H, RPT208067H, RPT208011H, RPT211592H, RPT210239H, RPT205994H, RPT207453H, RPT203878H, RPT204530H, RPT206531H, RPT206737H, RPT208008H, RPT203784H, RPT204794H, RPT208679H, RPT209307H, RPT211156H, RPT211157H, RPT205960H, RPT206040H, RPT208894H, RPT210172H, RPT210432H, RPT204276H, RPT204414H, RPT206283H, RPT210404H, RPT210561H, RPT211060H, RPT207046H, RPT203930H, RPT210041H, RPT210054H, RPT210853H, RPT205625H, RPT203736H, RPT203747H, RPT204337H, RPT204473H, RPT204769H, RPT207149H, RPT209366H, RPT206126H, RPT207165H, RPT207245H, RPT210836H, RPT203801H, RPT206619H, RPT208291H, RPT208734H, RPT203826H, RPT205333H, RPT205358H, RPT208543H, RPT208569H, RPT204208H, RPT205302H, RPT211251H, RPT210121H, RPT210374H, RPT211261H, RPT204215H, RPT205009H, RPT205408H, RPT207500H, RPT210082H, RPT211398H, RPT204785H, RPT208792H, RPT203834H, RPT203835H, RPT203838H, RPT203841H, RPT205922H, RPT205926H, RPT205928H, RPT206513H, RPT207492H, RPT211486H, RPT203732H, RPT204977H, RPT207923H, RPT208004H, RPT208732H, RPT211475H, RPT204492H, RPT205555H, RPT208343H, RPT208625H, RPT209009H, RPT209603H, RPT209693H, RPT206682H, RPT204861H, RPT209392H, RPT209709H, RPT210680H, RPT210710H, RPT204453H, RPT204458H, RPT204463H, RPT205900H, RPT205904H, RPT205907H, RPT205909H, RPT205910H, RPT205911H, RPT205918H, RPT206272H, RPT206826H, RPT206828H, RPT206829H, RPT206831H, RPT207770H, RPT207771H, RPT207772H, RPT207773H, RPT207776H, RPT208197H, RPT208627H, RPT208630H, RPT208739H, RPT208740H, RPT208748H, RPT209147H, RPT209246H, RPT209408H, RPT209409H, RPT209413H, RPT209415H, RPT209416H, RPT209418H, RPT211086H, RPT207391H, RPT210703H, RPT210833H, RPT209085H, RPT205862H, RPT207071H, RPT209721H, RPT211495H, RPT203785H, RPT204119H, RPT204139H, RPT205225H, RPT205334H, RPT205774H, RPT206244H, RPT206427H, RPT207206H, RPT209027H, RPT209170H, RPT209557H, RPT209869H, RPT210557H, RPT210560H, RPT211029H, RPT204383H, RPT205082H, RPT205095H, RPT205420H, RPT205855H, RPT206477H, RPT206997H, RPT207430H, RPT207930H, RPT208111H, RPT208441H, RPT208498H, RPT208791H, RPT209112H, RPT210448H, RPT210627H, RPT211121H, RPT207356H, RPT204643H, RPT204644H, RPT206210H, RPT206239H, RPT206254H, RPT206255H, RPT206256H, RPT206645H, RPT207794H, RPT208086H, RPT208815H, RPT208816H, RPT208817H, RPT209220H, RPT209376H, RPT209377H, RPT209378H, RPT209379H, RPT209382H, RPT209383H, RPT210835H, RPT211239H, RPT211242H, RPT211418H, RPT211565H, RPT211573H, RPT205217H, RPT205220H, RPT203729H, RPT203796H, RPT203800H, RPT204016H, RPT204019H, RPT204035H, RPT204037H, RPT204039H, RPT204040H, RPT204041H, RPT204044H, RPT204057H, RPT204070H, RPT204096H, RPT205187H, RPT206072H, RPT206582H, RPT206586H, RPT206590H, RPT206591H, RPT206592H, RPT206597H, RPT207600H, RPT207790H, RPT208178H, RPT208179H, RPT208180H, RPT208181H, RPT208182H, RPT208183H, RPT208186H, RPT208188H, RPT208189H, RPT208191H, RPT208192H, RPT208193H, RPT208196H, RPT208199H, RPT208200H, RPT208205H, RPT208452H, RPT208567H, RPT208616H, RPT208617H, RPT208622H, RPT208632H, RPT208634H, RPT208635H, RPT210647H, RPT211148H, RPT211172H, RPT211174H, RPT211175H, RPT211177H, RPT211178H, RPT211431H, RPT211435H, RPT211625H, RPT204339H, RPT209643H, RPT203967H, RPT205293H, RPT206346H, RPT206419H, RPT207297H, RPT210542H, RPT210692H, RPT211258H, RPT211663H, RPT205705H, RPT207389H, RPT208483H, RPT210440H, RPT204639H, RPT205163H, RPT207434H, RPT209485H, RPT209486H, RPT210330H, RPT210333H, RPT211446H, RPT209738H, RPT210582H, RPT205118H, RPT210538H, RPT211204H, RPT211492H, RPT204302H, RPT205127H, RPT205129H, RPT206861H, RPT208412H, RPT211709H, RPT204265H, RPT204944H, RPT205338H, RPT205690H, RPT205941H, RPT206077H, RPT207371H, RPT208139H, RPT209288H, RPT209471H, RPT210466H, RPT210753H, RPT210875H, RPT210878H, RPT210009H, RPT204573H, RPT205491H, RPT208019H, RPT204049H, RPT205186H, RPT211651H, RPT207305H, RPT207472H, RPT205304H, RPT205580H, RPT208803H, RPT210936H, RPT205335H, RPT208311H, RPT208265H, RPT209278H, RPT210188H, RPT210997H, RPT211559H, RPT203935H, RPT203939H, RPT204426H, RPT204657H, RPT204658H, RPT204661H, RPT205200H, RPT206410H, RPT206689H, RPT206692H, RPT206696H, RPT207920H, RPT207935H, RPT207937H, RPT207939H, RPT207961H, RPT207962H, RPT208341H, RPT208511H, RPT208975H, RPT209279H, RPT209584H, RPT209995H, RPT210231H, RPT210236H, RPT210267H, RPT211173H, RPT211185H, RPT211200H, RPT211221H, RPT204100H, RPT206245H, RPT207968H, RPT208521H, RPT210494H, RPT210551H, RPT210583H, RPT204084H, RPT208123H, RPT210071H, RPT205155H, RPT206718H, RPT207148H, RPT208989H, RPT204106H, RPT209957H, RPT209992H, RPT210113H, RPT210450H, RPT209241H, RPT211059H, RPT204879H, RPT206016H, RPT210983H, RPT211280H, RPT203999H, RPT204111H, RPT206587H, RPT207175H, RPT207767H, RPT210849H, RPT203779H, RPT204546H, RPT206716H, RPT208109H, RPT208228H, RPT204216H, RPT205846H, RPT209474H, RPT210163H, RPT208820H, RPT208322H, RPT209203H, RPT211463H, RPT203719H, RPT204497H, RPT204499H, RPT205848H, RPT206803H, RPT208722H, RPT208926H, RPT210150H, RPT211632H, RPT203778H, RPT204227H, RPT205388H, RPT207488H, RPT209385H, RPT209566H, RPT210648H, RPT210895H, RPT204170H, RPT205177H, RPT206451H, RPT207082H, RPT208880H, RPT209202H, RPT209318H, RPT209447H, RPT209579H, RPT210010H, RPT210518H, RPT204834H, RPT205245H, RPT207313H, RPT209653H, RPT205100H, RPT206041H, RPT207160H, RPT207324H, RPT209086H, RPT210018H, RPT210356H, RPT210447H, RPT206612H, RPT206613H, RPT206618H, RPT206621H, RPT210505H, RPT207447H, RPT207474H, RPT208168H, RPT208213H, RPT208245H, RPT208303H, RPT208999H, RPT209267H, RPT210098H, RPT211627H, RPT207479H, RPT205621H, RPT205076H, RPT205870H, RPT205871H, RPT206185H, RPT209276H, RPT210771H, RPT210974H, RPT204093H, RPT206336H, RPT206616H, RPT209344H, RPT209950H, RPT210562H, RPT203809H, RPT204676H, RPT204908H, RPT205150H, RPT205241H, RPT206434H, RPT209105H, RPT209576H, RPT210907H, RPT211562H, RPT204264H, RPT205361H, RPT206330H, RPT208908H, RPT209370H, RPT209483H, RPT209554H, RPT206900H, RPT206933H, RPT208822H, RPT208823H, RPT209018H, RPT210002H, RPT210748H, RPT208924H, RPT211127H, RPT205851H, RPT208160H, RPT210136H, RPT205142H, RPT208208H, RPT208799H, RPT209715H, RPT210100H, RPT210903H, RPT211581H, RPT204355H, RPT208330H, RPT209333H, RPT204130H, RPT206316H, RPT204789H, RPT207368H, RPT203743H, RPT206646H, RPT204155H, RPT209402H, RPT208676H, RPT204993H, RPT206225H, RPT207675H, RPT211145H, RPT211146H, RPT211147H, RPT211510H, RPT211511H, RPT204422H, RPT210732H, RPT206784H, RPT210660H, RPT204983H, RPT205634H, RPT206457H, RPT206474H, RPT208235H, RPT209308H, RPT205104H, RPT209568H, RPT208525H, RPT211278H, RPT211308H, RPT203716H, RPT207910H, RPT209140H, RPT210431H, RPT211056H, RPT211061H, RPT211401H, RPT211547H, RPT211643H, RPT210837H, RPT205260H, RPT206246H, RPT206453H, RPT206890H, RPT207483H, RPT207969H, RPT208298H, RPT208916H, RPT209932H, RPT210743H, RPT210772H, RPT205590H, RPT205596H, RPT205597H, RPT207365H, RPT208422H, RPT209230H, RPT209428H, RPT207952H, RPT210453H, RPT210787H, RPT211346H, RPT211563H, RPT203733H, RPT205390H, RPT205446H, RPT205950H, RPT206133H, RPT207108H, RPT207288H, RPT208124H, RPT209229H, RPT206991H, RPT205269H, RPT203706H, RPT206019H, RPT206171H, RPT207597H, RPT208952H, RPT209732H, RPT210049H, RPT205604H, RPT207993H, RPT208301H, RPT205617H, RPT207326H, RPT207698H, RPT207708H, RPT207745H, RPT210383H, RPT210599H, RPT210719H, RPT210724H, RPT210730H, RPT211412H, RPT207519H, RPT210206H, RPT209591H, RPT210314H, RPT204116H, RPT204351H, RPT204758H, RPT206539H, RPT207147H, RPT208417H, RPT209720H, RPT211053H, RPT209842H, RPT211267H, RPT211294H, RPT211370H, RPT204484H, RPT204490H, RPT205707H, RPT205901H, RPT206326H, RPT206575H, RPT207183H, RPT207635H, RPT208277H, RPT208787H, RPT209499H, RPT209536H, RPT210124H, RPT210135H, RPT210142H, RPT210155H, RPT210156H, RPT210244H, RPT211008H, RPT204293H, RPT209686H, RPT204173H, RPT205042H, RPT205198H, RPT205199H, RPT205839H, RPT206312H, RPT206315H, RPT206318H, RPT206339H, RPT208837H, RPT208841H, RPT209761H, RPT209762H, RPT209803H, RPT209804H, RPT207361H, RPT209656H, RPT208308H, RPT204790H, RPT205084H, RPT203776H, RPT204709H, RPT204735H, RPT204736H, RPT204767H, RPT204772H, RPT204788H, RPT207399H, RPT207863H, RPT209116H, RPT209119H, RPT209219H, RPT210042H, RPT210043H, RPT210058H, RPT206247H, RPT210301H, RPT204225H, RPT204332H, RPT205117H, RPT205275H, RPT205375H, RPT205376H, RPT206690H, RPT207267H, RPT207534H, RPT208810H, RPT208988H, RPT209867H, RPT209935H, RPT210420H, RPT210617H, RPT210897H, RPT210993H, RPT211192H, RPT211247H, RPT211421H, RPT211454H, RPT205607H, RPT209811H, RPT209240H, RPT209541H, RPT209962H, RPT211519H, RPT205600H, RPT205741H, RPT206430H, RPT206668H, RPT208159H, RPT208299H, RPT208751H, RPT209481H, RPT209565H, RPT210522H, RPT211530H, RPT204696H, RPT205144H, RPT206009H, RPT206436H, RPT206568H, RPT206573H, RPT206647H, RPT206705H, RPT207039H, RPT207127H, RPT207398H, RPT207466H, RPT207647H, RPT207783H, RPT207791H, RPT208334H, RPT208366H, RPT208388H, RPT208450H, RPT208685H, RPT208800H, RPT208843H, RPT208875H, RPT208953H, RPT209056H, RPT209062H, RPT209090H, RPT209120H, RPT209263H, RPT209699H, RPT209700H, RPT209897H, RPT209910H, RPT209936H, RPT210032H, RPT210066H, RPT210128H, RPT210139H, RPT210433H, RPT210498H, RPT210683H, RPT210722H, RPT211112H, RPT211182H, RPT211306H, RPT211386H, RPT211500H, RPT211638H, RPT204594H, RPT206208H, RPT206379H, RPT207552H, RPT209882H, RPT211020H, RPT211374H, RPT210857H, RPT206795H, RPT207858H, RPT207862H, RPT207996H, RPT208509H, RPT208689H, RPT208693H, RPT210789H, RPT211284H, RPT211287H, RPT208016H, RPT211589H, RPT205970H, RPT209744H, RPT209755H, RPT207351H, RPT203920H, RPT203921H, RPT203922H, RPT206884H, RPT207287H, RPT208650H, RPT208655H, RPT211363H, RPT211364H, RPT209212H, RPT210629H, RPT203812H, RPT204060H, RPT204089H, RPT205533H, RPT206064H, RPT206520H, RPT207737H, RPT209868H, RPT210148H, RPT210289H, RPT206572H, RPT207214H, RPT210681H, RPT210932H, RPT210649H, RPT210733H, RPT205133H, RPT210007H, RPT209553H, RPT208132H, RPT208774H, RPT211578H, RPT204467H, RPT205534H, RPT206032H, RPT206432H, RPT206794H, RPT207524H, RPT209127H, RPT209256H, RPT209658H, RPT209710H, RPT211193H, RPT210818H, RPT204109H, RPT204304H, RPT204306H, RPT205101H, RPT205190H, RPT205326H, RPT205868H, RPT206029H, RPT206043H, RPT206162H, RPT207077H, RPT207248H, RPT208793H, RPT210016H, RPT211332H, RPT209444H, RPT207218H, RPT208138H, RPT210279H, RPT210516H, RPT211183H, RPT211324H, RPT203746H, RPT203863H, RPT203865H, RPT205386H, RPT211596H, RPT203972H, RPT209612H, RPT210712H, RPT211697H, RPT206116H, RPT207917H, RPT207186H, RPT207439H, RPT209893H, RPT204118H, RPT204524H, RPT206881H, RPT210268H, RPT204212H, RPT208758H, RPT205822H, RPT206788H, RPT206791H, RPT207808H, RPT210322H, RPT210734H, RPT204348H, RPT211331H, RPT206182H, RPT206183H, RPT208925H, RPT209960H, RPT210507H, RPT205372H, RPT205380H, RPT206227H, RPT208078H, RPT208706H, RPT210888H, RPT204009H, RPT205295H, RPT205355H, RPT206023H, RPT206540H, RPT206541H, RPT208698H, RPT208777H, RPT208994H, RPT209121H, RPT209198H, RPT209265H, RPT209299H, RPT209490H, RPT209551H, RPT209886H, RPT210598H, RPT210790H, RPT211027H, RPT211366H, RPT211388H, RPT211440H, RPT211603H, RPT211677H, RPT211708H, RPT205889H, RPT208243H, RPT204228H, RPT209143H, RPT210117H, RPT206189H, RPT206331H, RPT206471H, RPT208297H, RPT208893H, RPT208896H, RPT210352H, RPT206719H, RPT208032H, RPT206298H, RPT208554H, RPT208578H, RPT205669H, RPT208976H, RPT203945H, RPT205450H, RPT206775H, RPT207016H, RPT207079H, RPT207132H, RPT207334H, RPT207458H, RPT208317H, RPT208446H, RPT209117H, RPT209522H, RPT211134H, RPT211154H, RPT211682H, RPT206565H, RPT208736H, RPT209209H, RPT210650H, RPT209733H, RPT210108H, RPT207934H, RPT209667H, RPT209831H, RPT206470H, RPT209190H, RPT207349H, RPT207908H, RPT208434H, RPT208606H, RPT207650H, RPT204191H, RPT204222H, RPT204260H, RPT204989H, RPT205026H, RPT205858H, RPT206252H, RPT206502H, RPT206637H, RPT207075H, RPT208282H, RPT208604H, RPT205041H, RPT208657H, RPT210597H, RPT203744H, RPT210462H, RPT210926H, RPT206204H, RPT207303H, RPT209104H, RPT210762H, RPT210922H, RPT210948H, RPT204400H, RPT205000H, RPT206511H, RPT206922H, RPT203740H, RPT203741H, RPT207128H, RPT209154H, RPT205233H, RPT205325H, RPT205401H, RPT206553H, RPT206652H, RPT207286H, RPT207404H, RPT207914H, RPT209440H, RPT210399H, RPT211269H, RPT204853H, RPT211062H, RPT207497H, RPT208390H, RPT206168H, RPT211417H, RPT208136H, RPT210332H, RPT204150H, RPT204266H, RPT204937H, RPT204963H, RPT205318H, RPT205331H, RPT206359H, RPT209291H, RPT209464H, RPT210568H, RPT210612H, RPT210815H, RPT211689H, RPT205453H, RPT207330H, RPT208636H, RPT203756H, RPT204048H, RPT204384H, RPT205106H, RPT205344H, RPT206076H, RPT206078H, RPT207118H, RPT207323H, RPT208304H, RPT209696H, RPT209821H, RPT209951H, RPT211699H, RPT205752H, RPT205756H, RPT205896H, RPT206395H, RPT207120H, RPT207602H, RPT207662H, RPT208031H, RPT208328H, RPT208426H, RPT210644H, RPT204270H, RPT204280H, RPT210901H, RPT211600H, RPT203774H, RPT203963H, RPT204533H, RPT204695H, RPT206192H, RPT206284H, RPT206413H, RPT206468H, RPT207316H, RPT205874H, RPT203888H, RPT206251H, RPT210302H, RPT209764H, RPT206188H, RPT206293H, RPT208960H, RPT210106H, RPT211218H, RPT211327H, RPT211206H, RPT211208H, RPT209829H, RPT211300H, RPT204317H, RPT209254H, RPT210308H, RPT206695H, RPT206169H, RPT206027H, RPT207030H, RPT208333H, RPT208059H, RPT208933H, RPT208936H, RPT209082H, RPT210601H, RPT210396H, RPT210524H, RPT204566H, RPT205993H, RPT206644H, RPT208623H, RPT209759H, RPT210395H, RPT210415H, RPT210935H, RPT204910H, RPT206864H, RPT209994H, RPT210541H, RPT210691H, RPT211553H, RPT208735H, RPT204822H, RPT204947H, RPT205561H, RPT207121H, RPT209089H, RPT210090H, RPT211402H, RPT208798H, RPT205529H, RPT210118H, RPT205077H, RPT208039H, RPT208295H, RPT210715H, RPT211257H, RPT208995H, RPT210959H, RPT204491H, RPT210105H, RPT204982H, RPT205287H, RPT204647H, RPT204859H, RPT207126H, RPT211442H, RPT211650H, RPT205249H, RPT204774H, RPT209729H, RPT211595H, RPT204748H, RPT209040H, RPT209226H, RPT210348H, RPT210971H, RPT204899H, RPT204353H, RPT206425H, RPT204248H, RPT204793H, RPT211128H, RPT210592H, RPT211392H, RPT209837H, RPT210956H, RPT204430H, RPT205020H, RPT205208H, RPT205583H, RPT206108H, RPT207093H, RPT207156H, RPT209095H, RPT209543H, RPT210174H, RPT210187H, RPT210306H, RPT211139H, RPT205138H, RPT206234H, RPT206625H, RPT206877H, RPT206878H, RPT208274H, RPT209367H, RPT209934H, RPT210326H, RPT211181H, RPT203902H, RPT203903H, RPT206000H, RPT206098H, RPT207797H, RPT207799H, RPT207803H, RPT209138H, RPT210589H, RPT203722H, RPT204974H, RPT206480H, RPT208325H, RPT210884H, RPT206236H, RPT206603H, RPT207124H, RPT209048H, RPT209324H, RPT210258H, RPT204668H, RPT204670H, RPT204681H, RPT204685H, RPT205660H, RPT206384H, RPT206858H, RPT206873H, RPT206874H, RPT207652H, RPT208641H, RPT208644H, RPT208699H, RPT210739H, RPT205167H, RPT205114H, RPT210802H, RPT208530H, RPT210855H, RPT207038H, RPT205206H, RPT205687H, RPT208302H, RPT208352H, RPT208397H, RPT208398H, RPT208941H, RPT209007H, RPT209024H, RPT209041H, RPT209186H, RPT209297H, RPT209660H, RPT209813H, RPT210341H, RPT210446H, RPT210885H, RPT211393H, RPT211586H, RPT205981H, RPT211143H, RPT204013H, RPT204678H, RPT204679H, RPT211629H, RPT205087H, RPT203691H, RPT204146H, RPT205058H, RPT205383H, RPT208782H, RPT210939H, RPT208883H, RPT210962H, RPT211088H, RPT211583H, RPT205978H, RPT206532H, RPT208682H, RPT204719H, RPT207119H, RPT203688H, RPT203759H, RPT207376H, RPT204274H, RPT208173H, RPT209201H, RPT210089H, RPT211325H, RPT211672H, RPT204740H, RPT205601H, RPT205937H, RPT206429H, RPT206600H, RPT206602H, RPT206604H, RPT207955H, RPT211481H, RPT211588H, RPT210238H, RPT211212H, RPT208872H, RPT208973H, RPT211005H, RPT211023H, RPT211458H, RPT211523H, RPT211524H, RPT211552H, RPT211655H, RPT204689H, RPT206688H, RPT206693H, RPT207060H, RPT207284H, RPT208940H, RPT209221H, RPT209644H, RPT210143H, RPT211571H, RPT208413H, RPT208669H, RPT211135H, RPT211396H, RPT204147H, RPT206157H, RPT208022H, RPT208804H, RPT210852H, RPT211285H, RPT211304H, RPT205573H, RPT208695H, RPT204929H, RPT204953H, RPT204996H, RPT205040H, RPT206129H, RPT210942H, RPT208147H, RPT204887H, RPT204050H, RPT208461H, RPT206901H, RPT210869H, RPT211645H, RPT204312H, RPT204755H, RPT208654H, RPT209841H, RPT204197H, RPT205277H, RPT206026H, RPT206084H, RPT206085H, RPT207068H, RPT208671H, RPT211462H, RPT211683H, RPT205833H, RPT207138H, RPT209473H, RPT210808H, RPT210810H, RPT210814H, RPT208238H, RPT209879H, RPT208383H, RPT204642H, RPT206691H, RPT204311H, RPT204314H, RPT204316H, RPT204324H, RPT204325H, RPT204328H, RPT204347H, RPT204356H, RPT204357H, RPT204360H, RPT205708H, RPT205712H, RPT205721H, RPT205724H, RPT205726H, RPT206501H, RPT206504H, RPT206506H, RPT207622H, RPT207623H, RPT207624H, RPT209353H, RPT209356H, RPT209357H, RPT209361H, RPT209363H, RPT209364H, RPT210235H, RPT210242H, RPT210243H, RPT210249H, RPT210257H, RPT210266H, RPT211249H, RPT211266H, RPT204322H, RPT204803H, RPT206389H, RPT208404H, RPT208869H, RPT209763H, RPT204365H, RPT211055H, RPT203899H, RPT204579H, RPT205614H, RPT206783H, RPT209118H, RPT209533H, RPT206950H, RPT209275H, RPT211210H, RPT204837H, RPT206882H, RPT207098H, RPT207291H, RPT207798H, RPT207800H, RPT207804H, RPT207805H, RPT207820H, RPT207822H, RPT207831H, RPT207839H, RPT207845H, RPT207848H, RPT207856H, RPT207929H, RPT207932H, RPT207933H, RPT208079H, RPT208083H, RPT208839H, RPT209371H, RPT210430H, RPT210614H, RPT210631H, RPT211281H, RPT211434H, RPT211499H, RPT211525H, RPT204616H, RPT204618H, RPT204636H, RPT205467H, RPT206706H, RPT203772H, RPT204340H, RPT205357H, RPT205365H, RPT205367H, RPT206314H, RPT206320H, RPT206321H, RPT206322H, RPT207072H, RPT207139H, RPT207140H, RPT207141H, RPT207142H, RPT207144H, RPT207145H, RPT207146H, RPT207847H, RPT207849H, RPT207857H, RPT209019H, RPT209573H, RPT210227H, RPT210325H, RPT210327H, RPT210721H, RPT210795H, RPT211168H, RPT211213H, RPT211532H, RPT211534H, RPT211619H, RPT211626H, RPT211662H, RPT211668H, RPT203971H, RPT203700H, RPT204402H, RPT204507H, RPT204508H, RPT204510H, RPT204511H, RPT204513H, RPT204516H, RPT204518H, RPT204519H, RPT204520H, RPT204521H, RPT204522H, RPT204523H, RPT204527H, RPT204535H, RPT204536H, RPT204537H, RPT204538H, RPT204541H, RPT204542H, RPT204543H, RPT204907H, RPT204932H, RPT205027H, RPT205028H, RPT205263H, RPT205695H, RPT205700H, RPT205701H, RPT205958H, RPT205961H, RPT206134H, RPT206135H, RPT206137H, RPT206144H, RPT206145H, RPT206149H, RPT206151H, RPT206338H, RPT206341H, RPT206342H, RPT206347H, RPT206348H, RPT206352H, RPT206356H, RPT206357H, RPT206358H, RPT206361H, RPT206362H, RPT206366H, RPT206369H, RPT206372H, RPT206374H, RPT206376H, RPT206740H, RPT206743H, RPT206744H, RPT206745H, RPT206748H, RPT206749H, RPT206750H, RPT206753H, RPT206754H, RPT206756H, RPT206757H, RPT207948H, RPT208249H, RPT208250H, RPT208253H, RPT208254H, RPT208261H, RPT208478H, RPT208489H, RPT208508H, RPT208573H, RPT208574H, RPT208575H, RPT208576H, RPT208589H, RPT209074H, RPT209141H, RPT209145H, RPT209150H, RPT209153H, RPT209155H, RPT209157H, RPT209158H, RPT209216H, RPT209432H, RPT209435H, RPT209436H, RPT209439H, RPT209453H, RPT209456H, RPT209458H, RPT210044H, RPT210048H, RPT210052H, RPT210214H, RPT210215H, RPT210216H, RPT210218H, RPT210223H, RPT210362H, RPT210576H, RPT210630H, RPT210634H, RPT210636H, RPT210637H, RPT210639H, RPT210640H, RPT210783H, RPT210791H, RPT210793H, RPT210796H, RPT210798H, RPT210800H, RPT210801H, RPT210846H, RPT210921H, RPT210961H, RPT211075H, RPT211111H, RPT211195H, RPT211196H, RPT211197H, RPT211199H, RPT211201H, RPT211711H, RPT203668H, RPT205387H, RPT205924H, RPT206056H, RPT206300H, RPT206449H, RPT204290H, RPT205482H, RPT208155H, RPT210290H, RPT210590H, RPT210823H, RPT205317H, RPT211574H, RPT208020H, RPT209058H, RPT209475H, RPT209477H, RPT211599H, RPT211648H, RPT210496H, RPT208165H, RPT208796H, RPT211333H, RPT210656H, RPT210976H, RPT210894H, RPT211353H, RPT209702H, RPT210095H, RPT210097H, RPT210372H, RPT210377H, RPT210378H, RPT210555H, RPT210981H, RPT211046H, RPT211527H, RPT211529H, RPT211679H, RPT211680H, RPT211203H, RPT206686H, RPT209528H, RPT209550H, RPT210534H, RPT210210H, RPT211205H, RPT211207H, RPT210755H, RPT211615H, RPT207113H, RPT207658H, RPT207970H, RPT208029H, RPT208512H, RPT208922H, RPT208938H, RPT209000H, RPT209168H, RPT209169H, RPT209179H, RPT209182H, RPT209195H, RPT209257H, RPT209258H, RPT209259H, RPT209264H, RPT209266H, RPT209283H, RPT209623H, RPT209625H, RPT209655H, RPT209657H, RPT209663H, RPT209664H, RPT209692H, RPT209695H, RPT210000H, RPT210001H, RPT210020H, RPT210060H, RPT210224H, RPT210546H, RPT210696H, RPT210725H, RPT210726H, RPT210736H, RPT210737H, RPT211407H, RPT211408H, RPT211484H, RPT211544H, RPT211557H, RPT211582H, RPT211587H, RPT211671H; GTIN 00763000059552, Lot Serial Numbers: RPT604986S, RPT605274S, RPT605732S, RPT606733S, RPT605040S, RPT603073S, RPT604908S, RPT603563S, RPT605339S, RPT605852S, RPT607151S, RPT607153S, RPT607154S, RPT607150S, RPT607152S, RPT604978S, RPT605942S, RPT605284S, RPT605304S, RPT605984S, RPT606129S, RPT606753S, RPT607326S, RPT604084S, RPT606295S, RPT600220S, RPT604415S, RPT605206S, RPT605812S, RPT605861S, RPT606406S, RPT601014S, RPT603770S, RPT604675S, RPT604693S, RPT604431S, RPT607570S, RPT605691S, RPT605823S, RPT606328S, RPT600197S, RPT600364S, RPT600365S, RPT601001S, RPT601271S, RPT601937S, RPT602030S, RPT602590S, RPT602591S, RPT602617S, RPT602618S, RPT603194S, RPT603195S, RPT603927S, RPT603960S, RPT603971S, RPT603972S, RPT604051S, RPT604052S, RPT604406S, RPT602996S, RPT603137S, RPT603145S, RPT603171S, RPT603715S, RPT603897S, RPT603994S, RPT603443S, RPT605317S, RPT605318S, RPT607011S, RPT602720S, RPT602776S, RPT603187S, RPT603630S, RPT600728S, RPT601732S, RPT602092S, RPT605006S, RPT603222S, RPT605073S, RPT600270S, RPT600640S, RPT601623S, RPT602430S, RPT602558S, RPT603061S, RPT603107S, RPT603114S, RPT603749S, RPT603986S, RPT604662S, RPT605307S, RPT606466S, RPT601933S, RPT602836S, RPT603569S, RPT606695S, RPT605208S, RPT605409S, RPT606299S, RPT606581S, RPT600922S, RPT601040S, RPT601476S, RPT601584S, RPT602491S, RPT602557S, RPT603414S, RPT603441S, RPT603480S, RPT604306S, RPT604405S, RPT606193S, RPT606536S, RPT607068S, RPT604706S, RPT604918S, RPT604948S, RPT605046S, RPT605388S, RPT605506S, RPT605572S, RPT605915S, RPT606260S, RPT607700S, RPT600455S, RPT600509S, RPT601110S, RPT601212S, RPT601360S, RPT601512S, RPT601825S, RPT601834S, RPT601887S, RPT602247S, RPT602324S, RPT602881S, RPT602959S, RPT603085S, RPT603099S, RPT603964S, RPT604070S, RPT604260S, RPT604387S, RPT607072S, RPT601797S, RPT602640S, RPT603138S, RPT604149S, RPT605086S, RPT600685S, RPT606874S, RPT606875S, RPT600375S, RPT603278S, RPT605752S, RPT600467S, RPT601868S, RPT603082S, RPT602222S, RPT605690S, RPT605997S, RPT600501S, RPT600991S, RPT601699S, RPT602481S, RPT602876S, RPT603008S, RPT603103S, RPT604247S, RPT600406S, RPT601254S, RPT605124S, RPT605171S, RPT605192S, RPT606284S, RPT607284S, RPT600924S, RPT603410S, RPT604443S, RPT607282S, RPT601580S, RPT604177S, RPT600341S, RPT604977S, RPT605280S, RPT605398S, RPT605474S, RPT605569S, RPT605731S, RPT606276S, RPT606741S, RPT606829S, RPT607055S, RPT607404S, RPT600241S, RPT600271S, RPT601032S, RPT601102S, RPT601245S, RPT601351S, RPT601534S, RPT602363S, RPT602370S, RPT603141S, RPT603317S, RPT603331S, RPT604168S, RPT604176S, RPT607389S, RPT607499S, RPT606906S, RPT600404S, RPT600670S, RPT600904S, RPT600959S, RPT601682S, RPT602223S, RPT605982S, RPT601205S, RPT602170S, RPT606823S, RPT603421S, RPT602621S, RPT604418S, RPT601379S, RPT600566S, RPT602733S, RPT602780S, RPT600735S, RPT600200S, RPT602501S, RPT602835S, RPT603123S, RPT604388S, RPT605429S, RPT600957S, RPT603935S, RPT604437S, RPT604514S, RPT606001S, RPT600722S, RPT600724S, RPT600725S, RPT600572S, RPT600573S, RPT600624S, RPT605036S, RPT605775S, RPT606068S, RPT603854S, RPT606524S, RPT605352S, RPT602735S, RPT603179S, RPT603711S, RPT601054S, RPT602383S, RPT604826S, RPT600367S, RPT602879S, RPT603988S, RPT604106S, RPT606503S, RPT600042S, RPT601311S, RPT601627S, RPT602980S, RPT603087S, RPT603337S, RPT603822S, RPT600730S, RPT600830S, RPT600382S, RPT604598S, RPT601321S, RPT601234S, RPT604666S, RPT600340S, RPT601400S, RPT604663S, RPT606473S, RPT606799S, RPT607333S, RPT600527S, RPT601599S, RPT602833S, RPT603446S, RPT603574S, RPT605533S, RPT603299S, RPT604057S, RPT606765S, RPT602761S, RPT603105S, RPT603827S, RPT603908S, RPT605295S, RPT605296S, RPT605681S, RPT601663S, RPT604780S, RPT600862S, RPT604279S, RPT605694S, RPT605764S, RPT600619S, RPT600855S, RPT601498S, RPT601642S, RPT601811S, RPT602443S, RPT602616S, RPT604242S, RPT604385S, RPT604390S, RPT604511S, RPT605058S, RPT605300S, RPT601416S, RPT603170S, RPT603573S, RPT606583S, RPT605265S, RPT600011S, RPT603703S, RPT602915S, RPT605220S, RPT605458S, RPT605740S, RPT605903S, RPT606196S, RPT606470S, RPT607031S, RPT607066S, RPT607083S, RPT607424S, RPT607425S, RPT607435S, RPT607453S, RPT600704S, RPT601062S, RPT601068S, RPT601204S, RPT601947S, RPT604018S, RPT604042S, RPT606571S, RPT606734S, RPT607365S, RPT600315S, RPT600463S, RPT601320S, RPT602084S, RPT604792S, RPT606407S, RPT600669S, RPT602310S, RPT603678S, RPT600517S, RPT605762S, RPT607234S, RPT600635S, RPT600715S, RPT601208S, RPT601526S, RPT602258S, RPT602756S, RPT602766S, RPT602778S, RPT603548S, RPT604154S, RPT607233S, RPT607486S, RPT603608S, RPT603674S, RPT601288S, RPT604945S, RPT600392S, RPT601100S, RPT604620S, RPT605673S, RPT601749S, RPT602604S, RPT604050S, RPT604130S, RPT604756S, RPT605183S, RPT605444S, RPT606812S, RPT600944S, RPT601025S, RPT603186S, RPT604370S, RPT604699S, RPT602482S, RPT600130S, RPT600133S, RPT600135S, RPT600136S, RPT600137S, RPT600138S, RPT600139S, RPT600140S, RPT600142S, RPT605130S, RPT605638S, RPT605645S, RPT605887S, RPT605894S, RPT607411S, RPT602942S, RPT602943S, RPT603876S, RPT600560S, RPT600312S, RPT600313S, RPT600555S, RPT600673S, RPT600674S, RPT601151S, RPT601877S, RPT602129S, RPT602308S, RPT604407S, RPT604408S, RPT606980S, RPT607040S, RPT605895S, RPT606065S, RPT606303S, RPT606850S, RPT606934S, RPT600317S, RPT600346S, RPT601776S, RPT602234S, RPT602798S, RPT603637S, RPT604248S, RPT604630S, RPT604997S, RPT605595S, RPT605597S, RPT606102S, RPT606273S, RPT600494S, RPT601039S, RPT601901S, RPT602407S, RPT602525S, RPT602659S, RPT602830S, RPT603016S, RPT603282S, RPT603501S, RPT603502S, RPT603724S, RPT604099S, RPT604542S, RPT604563S, RPT604566S, RPT605193S, RPT605326S, RPT603658S, RPT603934S, RPT604575S, RPT604577S, RPT604578S, RPT606388S, RPT606389S, RPT606390S, RPT606391S, RPT606392S, RPT606393S, RPT607203S, RPT607204S, RPT607205S, RPT600903S, RPT601263S, RPT601264S, RPT601265S, RPT601266S, RPT602545S, RPT602546S, RPT604187S, RPT607202S, RPT607206S, RPT604868S, RPT604869S, RPT605515S, RPT605745S, RPT605844S, RPT605846S, RPT605847S, RPT606174S, RPT606933S, RPT607028S, RPT601516S, RPT601528S, RPT601715S, RPT602503S, RPT602504S, RPT603120S, RPT603316S, RPT603817S, RPT604401S, RPT604539S, RPT604827S, RPT605264S, RPT605611S, RPT602068S, RPT604485S, RPT605932S, RPT602583S, RPT605026S, RPT606416S, RPT606474S, RPT601617S, RPT602016S, RPT602059S, RPT603493S, RPT603901S, RPT603947S, RPT604047S, RPT604615S, RPT605343S, RPT605930S, RPT600543S, RPT600680S, RPT603128S, RPT600513S, RPT601111S, RPT601611S, RPT602036S, RPT605505S, RPT602303S, RPT602602S, RPT602216S, RPT603396S, RPT600767S, RPT606032S, RPT604567S, RPT604568S, RPT605048S, RPT605316S, RPT605614S, RPT606071S, RPT606073S, RPT606118S, RPT606274S, RPT606447S, RPT606556S, RPT606640S, RPT607017S, RPT600569S, RPT601577S, RPT601711S, RPT602638S, RPT603204S, RPT603302S, RPT603667S, RPT603710S, RPT606431S, RPT607317S, RPT600103S, RPT600104S, RPT604683S, RPT605560S, RPT605946S, RPT606435S, RPT606749S, RPT600981S, RPT602991S, RPT604658S, RPT601997S, RPT602886S, RPT600528S, RPT600529S, RPT604672S, RPT606285S, RPT606616S, RPT604709S, RPT600190S, RPT602081S, RPT603378S, RPT603672S, RPT601686S, RPT603505S, RPT606021S, RPT606621S, RPT606909S, RPT607149S, RPT607449S, RPT600387S, RPT602054S, RPT603098S, RPT604140S, RPT604151S, RPT606403S, RPT604610S, RPT604614S, RPT604714S, RPT604731S, RPT604989S, RPT605034S, RPT605114S, RPT606943S, RPT606961S, RPT606977S, RPT601456S, RPT601567S, RPT601589S, RPT601777S, RPT602000S, RPT602240S, RPT602415S, RPT602476S, RPT602788S, RPT602819S, RPT603117S, RPT603656S, RPT603816S, RPT604125S, RPT603707S, RPT605358S, RPT605522S, RPT605851S, RPT605966S, RPT600666S, RPT601372S, RPT603510S, RPT603720S, RPT605413S, RPT602173S, RPT604773S, RPT604779S, RPT606972S, RPT607119S, RPT604328S, RPT602811S, RPT600953S, RPT602034S, RPT603944S, RPT603294S, RPT602845S, RPT603796S, RPT604790S, RPT600337S, RPT602796S, RPT604519S, RPT607370S, RPT601620S, RPT601964S, RPT602232S, RPT602814S, RPT604053S, RPT604067S, RPT604116S, RPT607159S, RPT600030S, RPT606533S, RPT600483S, RPT605580S, RPT603453S, RPT606116S, RPT600508S, RPT604595S, RPT604596S, RPT606272S, RPT604493S, RPT606831S, RPT607687S, RPT604145S, RPT607684S, RPT605406S, RPT605606S, RPT606283S, RPT607170S, RPT606171S, RPT606659S, RPT606830S, RPT607142S, RPT600967S, RPT601359S, RPT601746S, RPT603275S, RPT604435S, RPT606742S, RPT601174S, RPT604109S, RPT605229S, RPT605485S, RPT605588S, RPT606429S, RPT601842S, RPT602239S, RPT602810S, RPT602841S, RPT603215S, RPT604088S, RPT603340S, RPT604653S, RPT605501S, RPT605725S, RPT606591S, RPT601783S, RPT602786S, RPT604516S, RPT606990S, RPT603675S, RPT604716S, RPT604338S, RPT603223S, RPT603882S, RPT604374S, RPT607402S, RPT601037S, RPT601440S, RPT602773S, RPT600954S, RPT601357S, RPT602033S, RPT605340S, RPT601557S, RPT602424S, RPT602909S, RPT604483S, RPT601639S, RPT604840S, RPT606489S, RPT601237S, RPT605539S, RPT607314S, RPT607395S, RPT602505S, RPT602652S, RPT603092S, RPT601666S, RPT605149S, RPT605159S, RPT604272S, RPT606493S, RPT606562S, RPT600344S, RPT601654S, RPT601988S, RPT604487S, RPT603750S, RPT601080S, RPT602723S, RPT603604S, RPT605433S, RPT605574S, RPT606332S, RPT605435S, RPT606958S, RPT602477S, RPT603586S, RPT604638S, RPT606438S, RPT604261S, RPT604985S, RPT600653S, RPT600752S, RPT602527S, RPT606287S, RPT606889S, RPT601338S, RPT601341S, RPT605965S, RPT604507S, RPT604623S, RPT604927S, RPT605119S, RPT605242S, RPT605445S, RPT605766S, RPT605799S, RPT605889S, RPT605892S, RPT605998S, RPT606259S, RPT606338S, RPT606339S, RPT606494S, RPT606709S, RPT607106S, RPT607444S, RPT600025S, RPT600453S, RPT600871S, RPT601327S, RPT602834S, RPT603069S, RPT603078S, RPT603147S, RPT603306S, RPT603435S, RPT603464S, RPT603677S, RPT603777S, RPT603778S, RPT603779S, RPT603780S, RPT603831S, RPT603861S, RPT603896S, RPT603990S, RPT604049S, RPT604103S, RPT604129S, RPT604323S, RPT607683S, RPT600409S, RPT600590S, RPT602335S, RPT603180S, RPT604027S, RPT604115S, RPT604280S, RPT604299S, RPT603789S, RPT605365S, RPT605385S, RPT600821S, RPT600873S, RPT601153S, RPT601752S, RPT602962S, RPT603738S, RPT604671S, RPT600801S, RPT600230S, RPT605294S, RPT605950S, RPT600605S, RPT603291S, RPT603295S, RPT604148S, RPT605212S, RPT606130S, RPT606937S, RPT603355S, RPT604619S, RPT604621S, RPT604629S, RPT604799S, RPT605021S, RPT605078S, RPT605127S, RPT605338S, RPT605771S, RPT605772S, RPT605773S, RPT605774S, RPT606402S, RPT606404S, RPT606405S, RPT606809S, RPT601042S, RPT601043S, RPT601451S, RPT601458S, RPT601847S, RPT602153S, RPT602595S, RPT603721S, RPT606740S, RPT607034S, RPT607633S, RPT604624S, RPT604725S, RPT605480S, RPT606508S, RPT606998S, RPT600654S, RPT600667S, RPT601345S, RPT603361S, RPT607634S, RPT604659S, RPT604858S, RPT606185S, RPT606197S, RPT606598S, RPT606904S, RPT606918S, RPT606995S, RPT607235S, RPT607236S, RPT602488S, RPT603634S, RPT605158S, RPT603400S, RPT604302S, RPT604847S, RPT605028S, RPT605622S, RPT604631S, RPT602367S, RPT605199S, RPT607438S, RPT607563S, RPT601083S, RPT604048S, RPT605913S, RPT600706S, RPT604035S, RPT601712S, RPT602565S, RPT606755S, RPT600815S, RPT601578S, RPT602394S, RPT602593S, RPT602768S, RPT603104S, RPT603547S, RPT604055S, RPT607302S, RPT602373S, RPT602641S, RPT602664S, RPT604584S, RPT600245S, RPT601845S, RPT605109S, RPT601335S, RPT602020S, RPT603650S, RPT603570S, RPT605911S, RPT601439S, RPT601669S, RPT603997S, RPT601229S, RPT605806S, RPT605807S, RPT605808S, RPT605809S, RPT605810S, RPT605971S, RPT605972S, RPT605974S, RPT606305S, RPT606306S, RPT606307S, RPT606308S, RPT607039S, RPT607046S, RPT607048S, RPT601417S, RPT601419S, RPT601420S, RPT601421S, RPT601422S, RPT601423S, RPT601540S, RPT601816S, RPT601821S, RPT601826S, RPT601827S, RPT602304S, RPT602305S, RPT602306S, RPT602307S, RPT603457S, RPT603458S, RPT603459S, RPT603824S, RPT603825S, RPT603826S, RPT607038S, RPT607041S, RPT607042S, RPT607043S, RPT607044S, RPT607045S, RPT607047S, RPT607092S, RPT607172S, RPT607173S, RPT607174S, RPT607175S, RPT607176S, RPT607177S, RPT607178S, RPT607179S, RPT607180S, RPT607104S, RPT604959S, RPT606107S, RPT607120S, RPT607165S, RPT607190S, RPT600948S, RPT601319S, RPT602256S, RPT603221S, RPT603866S, RPT604225S, RPT607526S, RPT605629S, RPT606970S, RPT602174S, RPT603368S, RPT603970S, RPT604334S, RPT604804S, RPT605680S, RPT603801S, RPT604313S, RPT601003S, RPT604133S, RPT604579S, RPT604890S, RPT605354S, RPT605382S, RPT605562S, RPT606179S, RPT606180S, RPT601342S, RPT601643S, RPT601913S, RPT601921S, RPT602342S, RPT602459S, RPT603413S, RPT604022S, RPT604171S, RPT604188S, RPT604189S, RPT603718S, RPT605524S, RPT605687S, RPT605702S, RPT606585S, RPT600060S, RPT600547S, RPT601115S, RPT601548S, RPT603519S, RPT603828S, RPT604194S, RPT605155S, RPT605293S, RPT606235S, RPT606782S, RPT603614S, RPT603684S, RPT603685S, RPT603948S, RPT603949S, RPT603959S, RPT603963S, RPT604238S, RPT604462S, RPT607706S, RPT600496S, RPT601818S, RPT602541S, RPT602986S, RPT607373S, RPT605258S, RPT600366S, RPT600422S, RPT601618S, RPT602860S, RPT604226S, RPT604373S, RPT602131S, RPT602520S, RPT605516S, RPT603878S, RPT602075S, RPT604641S, RPT604665S, RPT604839S, RPT605039S, RPT605573S, RPT605853S, RPT606024S, RPT606258S, RPT607528S, RPT600283S, RPT601178S, RPT601510S, RPT601910S, RPT602246S, RPT602632S, RPT602654S, RPT602827S, RPT602969S, RPT603193S, RPT603451S, RPT603989S, RPT601192S, RPT606092S, RPT606873S, RPT601748S, RPT602445S, RPT605451S, RPT605906S, RPT606087S, RPT600380S, RPT600601S, RPT602923S, RPT605796S, RPT606242S, RPT600700S, RPT601658S, RPT600267S, RPT601190S, RPT602082S, RPT602255S, RPT602750S, RPT603511S, RPT604249S, RPT604681S, RPT605222S, RPT605452S, RPT605490S, RPT605827S, RPT603951S, RPT604043S, RPT604060S, RPT604414S, RPT606976S, RPT607299S, RPT607316S, RPT604237S, RPT604922S, RPT604984S, RPT600493S, RPT603835S, RPT605170S, RPT607337S, RPT607521S, RPT600079S, RPT600249S, RPT600449S, RPT600909S, RPT601366S, RPT601751S, RPT601919S, RPT602045S, RPT603784S, RPT603785S, RPT603247S, RPT600481S, RPT602042S, RPT606455S, RPT603424S, RPT603596S, RPT606775S, RPT606620S, RPT600765S, RPT603205S, RPT605527S, RPT605559S, RPT607102S, RPT607281S, RPT604964S, RPT605112S, RPT605609S, RPT602804S, RPT603404S, RPT603407S, RPT603432S, RPT603602S, RPT603669S, RPT604463S, RPT604481S, RPT604743S, RPT604763S, RPT605556S, RPT605621S, RPT605706S, RPT606221S, RPT606678S, RPT607073S, RPT607472S, RPT607518S, RPT600247S, RPT600250S, RPT600710S, RPT600750S, RPT600829S, RPT601700S, RPT601892S, RPT602328S, RPT602447S, RPT603009S, RPT603049S, RPT603163S, RPT603397S, RPT603739S, RPT603776S, RPT603783S, RPT603833S, RPT604195S, RPT604495S, RPT604496S, RPT604720S, RPT605188S, RPT605205S, RPT605297S, RPT605424S, RPT605683S, RPT605751S, RPT606693S, RPT606927S, RPT606951S, RPT602484S, RPT603202S, RPT603296S, RPT603332S, RPT604253S, RPT604501S, RPT604503S, RPT604547S, RPT602446S, RPT600351S, RPT601143S, RPT602019S, RPT604381S, RPT607451S, RPT604783S, RPT606089S, RPT607334S, RPT600574S, RPT602598S, RPT603965S, RPT601024S, RPT605223S, RPT606613S, RPT601086S, RPT601602S, RPT601608S, RPT602167S, RPT602168S, RPT602537S, RPT602771S, RPT600086S, RPT606472S, RPT600796S, RPT601183S, RPT601297S, RPT601404S, RPT601650S, RPT601994S, RPT602180S, RPT602646S, RPT603028S, RPT603821S, RPT605142S, RPT601938S, RPT602270S, RPT604428S, RPT605786S, RPT605902S, RPT603844S, RPT604893S, RPT604894S, RPT605203S, RPT605204S, RPT605551S, RPT605555S, RPT606109S, RPT606451S, RPT606703S, RPT606704S, RPT601770S, RPT601902S, RPT601957S, RPT601959S, RPT601962S, RPT603930S, RPT604011S, RPT604012S, RPT604475S, RPT604476S, RPT604477S, RPT605245S, RPT605541S, RPT605548S, RPT605550S, RPT606226S, RPT601765S, RPT601956S, RPT603273S, RPT603274S, RPT603371S, RPT603512S, RPT603732S, RPT605921S, RPT606639S, RPT606959S, RPT603333S, RPT603473S, RPT603683S, RPT603752S, RPT607398S, RPT602680S, RPT602717S, RPT603546S, RPT606017S, RPT606986S, RPT607290S, RPT601185S, RPT601789S, RPT602372S, RPT604809S, RPT605367S, RPT602519S, RPT602703S, RPT603742S, RPT606595S, RPT604782S, RPT605493S, RPT607259S, RPT602196S, RPT606507S, RPT605422S, RPT606953S, RPT600530S, RPT601079S, RPT601235S, RPT601830S, RPT602513S, RPT602622S, RPT602829S, RPT603370S, RPT603767S, RPT606342S, RPT602013S, RPT604676S, RPT605000S, RPT605116S, RPT605811S, RPT606409S, RPT606764S, RPT606766S, RPT606947S, RPT607005S, RPT607067S, RPT607069S, RPT607620S, RPT600293S, RPT600838S, RPT600859S, RPT602176S, RPT602400S, RPT602585S, RPT602781S, RPT603072S, RPT603594S, RPT603641S, RPT604916S, RPT600436S, RPT602451S, RPT604036S, RPT604703S, RPT605247S, RPT601119S, RPT602359S, RPT602442S, RPT603101S, RPT603815S, RPT603892S, RPT604245S, RPT604502S, RPT601873S, RPT604836S, RPT605143S, RPT606792S, RPT603865S, RPT603909S, RPT605716S, RPT605392S, RPT603857S, RPT604753S, RPT600996S, RPT600997S, RPT602563S, RPT602564S, RPT603377S, RPT603585S, RPT603593S, RPT604304S, RPT607008S, RPT605907S, RPT603798S, RPT605840S, RPT602502S, RPT605336S, RPT601601S, RPT602069S, RPT602070S, RPT604017S, RPT606006S, RPT601027S, RPT602002S, RPT602161S, RPT604375S, RPT604497S, RPT605387S, RPT606106S, RPT606635S, RPT601833S, RPT602300S, RPT605233S, RPT604718S, RPT605589S, RPT606028S, RPT606282S, RPT607064S, RPT607122S, RPT607212S, RPT601363S, RPT602460S, RPT602713S, RPT603487S, RPT603912S, RPT604192S, RPT604484S, RPT604543S, RPT604135S, RPT604200S, RPT606088S, RPT606178S, RPT603992S, RPT603050S, RPT603284S, RPT604005S, RPT605410S, RPT604251S, RPT605914S, RPT603987S, RPT604932S, RPT605096S, RPT601239S, RPT604944S, RPT605341S, RPT600462S, RPT604009S, RPT604518S, RPT603267S, RPT606866S, RPT601059S, RPT606824S, RPT603575S, RPT603644S, RPT604293S, RPT605131S, RPT605999S, RPT606218S, RPT606290S, RPT606618S, RPT606828S, RPT601683S, RPT604490S, RPT606280S, RPT605088S, RPT605093S, RPT605150S, RPT607294S, RPT601640S, RPT602175S, RPT603121S, RPT603336S, RPT603693S, RPT604137S, RPT602130S, RPT602792S, RPT603153S, RPT603154S, RPT604008S, RPT604471S, RPT604510S, RPT605167S, RPT605057S, RPT605154S, RPT606098S, RPT606430S, RPT606515S, RPT600050S, RPT600272S, RPT600719S, RPT600899S, RPT601044S, RPT601401S, RPT602926S, RPT603076S, RPT604327S, RPT604489S, RPT607309S, RPT604545S, RPT605071S, RPT605800S, RPT606132S, RPT603395S, RPT603811S, RPT601519S, RPT605139S, RPT601373S, RPT607375S, RPT604821S, RPT606056S, RPT602734S, RPT601755S, RPT603168S, RPT605279S, RPT605758S, RPT605990S, RPT606038S, RPT606458S, RPT600768S, RPT601387S, RPT601600S, RPT601760S, RPT601880S, RPT601900S, RPT602057S, RPT602060S, RPT602995S, RPT603627S, RPT603676S, RPT606155S, RPT606487S, RPT600039S, RPT604861S, RPT601911S, RPT604159S, RPT601812S, RPT605953S, RPT603300S, RPT601675S, RPT600426S, RPT607251S, RPT604570S, RPT604933S, RPT605051S, RPT605053S, RPT605070S, RPT605241S, RPT605768S, RPT606531S, RPT606706S, RPT606804S, RPT606852S, RPT606907S, RPT607252S, RPT602154S, RPT602169S, RPT602524S, RPT604254S, RPT604436S, RPT607249S, RPT605404S, RPT606069S, RPT606296S, RPT606347S, RPT607220S, RPT607295S, RPT603325S, RPT604576S, RPT604592S, RPT604599S, RPT604942S, RPT605013S, RPT605721S, RPT606216S, RPT606249S, RPT606262S, RPT606279S, RPT606370S, RPT600193S, RPT600418S, RPT600499S, RPT600568S, RPT600614S, RPT600622S, RPT601095S, RPT601186S, RPT601207S, RPT601252S, RPT601260S, RPT601615S, RPT601625S, RPT601895S, RPT602774S, RPT602922S, RPT603185S, RPT603324S, RPT603508S, RPT603625S, RPT603655S, RPT603725S, RPT604257S, RPT604550S, RPT607642S, RPT604673S, RPT600033S, RPT602115S, RPT606135S, RPT601214S, RPT603881S, RPT605390S, RPT600782S, RPT601049S, RPT602067S, RPT602126S, RPT603911S, RPT604379S, RPT602746S, RPT603540S, RPT603688S, RPT602022S, RPT602164S, RPT602441S, RPT605179S, RPT605646S, RPT603853S, RPT600036S, RPT600552S, RPT605176S, RPT606254S, RPT606327S, RPT607510S, RPT601005S, RPT601716S, RPT601722S, RPT601975S, RPT602418S, RPT602475S, RPT602705S, RPT602793S, RPT602794S, RPT602817S, RPT602952S, RPT602954S, RPT603208S, RPT603636S, RPT606214S, RPT606396S, RPT607514S, RPT605750S, RPT600519S, RPT600125S, RPT600608S, RPT601590S, RPT600080S, RPT601226S, RPT603177S, RPT603702S, RPT607247S, RPT604972S, RPT604973S, RPT604974S, RPT601073S, RPT603800S, RPT603921S, RPT604354S, RPT601761S, RPT605141S, RPT600861S, RPT601738S, RPT605115S, RPT606126S, RPT604806S, RPT605305S, RPT607214S, RPT607215S, RPT601130S, RPT601285S, RPT601352S, RPT601504S, RPT602012S, RPT602050S, RPT602090S, RPT602869S, RPT602968S, RPT603428S, RPT604193S, RPT607269S, RPT606312S, RPT604488S, RPT607448S, RPT607394S, RPT603606S, RPT602217S, RPT601575S, RPT602584S, RPT603323S, RPT603922S, RPT605002S, RPT605110S, RPT605705S, RPT605788S, RPT606156S, RPT600092S, RPT600585S, RPT601780S, RPT602263S, RPT602600S, RPT603216S, RPT603931S, RPT604163S, RPT604341S, RPT602072S, RPT602574S, RPT603848S, RPT604104S, RPT605535S, RPT605564S, RPT605577S, RPT606067S, RPT601221S, RPT602614S, RPT602624S, RPT602625S, RPT602629S, RPT602916S, RPT602945S, RPT603167S, RPT603288S, RPT605374S, RPT601374S, RPT604786S, RPT605777S, RPT605778S, RPT603209S, RPT604705S, RPT604883S, RPT605005S, RPT605346S, RPT605704S, RPT604326S, RPT604332S, RPT605427S, RPT605820S, RPT604532S, RPT604930S, RPT605383S, RPT603834S, RPT605767S, RPT601807S, RPT606058S, RPT600286S, RPT601986S, RPT602276S, RPT602449S, RPT603580S, RPT604079S, RPT600414S, RPT604905S, RPT600676S, RPT603804S, RPT606094S, RPT604970S, RPT600823S, RPT601120S, RPT601448S, RPT601470S, RPT601605S, RPT601678S, RPT601764S, RPT602209S, RPT602290S, RPT602421S, RPT602427S, RPT602458S, RPT602920S, RPT604527S, RPT604919S, RPT604950S, RPT606767S, RPT606974S, RPT604065S, RPT604434S, RPT600740S, RPT601129S, RPT603017S, RPT603018S, RPT604068S, RPT604482S, RPT605214S, RPT605885S, RPT606966S, RPT601492S, RPT602554S, RPT602578S, RPT603966S, RPT604025S, RPT604072S, RPT604196S, RPT604208S, RPT600852S, RPT605118S, RPT605197S, RPT600384S, RPT602197S, RPT604774S, RPT605126S, RPT606690S, RPT600398S, RPT600652S, RPT600751S, RPT601006S, RPT601057S, RPT601624S, RPT601781S, RPT601879S, RPT602500S, RPT603309S, RPT603663S, RPT604330S, RPT605161S, RPT605285S, RPT605507S, RPT605848S, RPT605849S, RPT607025S, RPT607158S, RPT600412S, RPT600969S, RPT601034S, RPT601199S, RPT601737S, RPT601946S, RPT601985S, RPT602066S, RPT602420S, RPT602859S, RPT602982S, RPT603060S, RPT603074S, RPT603360S, RPT603374S, RPT603416S, RPT603842S, RPT605084S, RPT605797S, RPT606147S, RPT606527S, RPT606541S, RPT607121S, RPT607418S, RPT607419S, RPT601088S, RPT601213S, RPT601251S, RPT601688S, RPT602327S, RPT602667S, RPT602698S, RPT602823S, RPT603066S, RPT603698S, RPT604150S, RPT604457S, RPT604558S, RPT606708S, RPT607417S, RPT603670S, RPT604990S, RPT605259S, RPT606728S, RPT606919S, RPT603236S, RPT601289S, RPT601334S, RPT604759S, RPT604761S, RPT605024S, RPT605025S, RPT605029S, RPT605072S, RPT605463S, RPT606266S, RPT606292S, RPT601195S, RPT601905S, RPT602575S, RPT602878S, RPT602960S, RPT603455S, RPT603961S, RPT604111S, RPT604509S, RPT605052S, RPT605780S, RPT606707S, RPT600705S, RPT602706S, RPT602848S, RPT606250S, RPT600186S, RPT603184S, RPT603589S, RPT604312S, RPT605014S, RPT606586S, RPT600472S, RPT606898S, RPT603127S, RPT604386S, RPT602041S, RPT604602S, RPT600696S, RPT600714S, RPT600181S, RPT600182S, RPT606137S, RPT603198S, RPT600773S, RPT600791S, RPT601630S, RPT601852S, RPT602029S, RPT602264S, RPT602487S, RPT603746S, RPT604114S, RPT604259S, RPT604710S, RPT605303S, RPT605891S, RPT601809S, RPT602214S, RPT606714S, RPT603805S, RPT603810S, RPT603829S, RPT604625S, RPT605032S, RPT605215S, RPT605633S, RPT601158S, RPT601425S, RPT601849S, RPT601850S, RPT602770S, RPT603463S, RPT603621S, RPT604888S, RPT605306S, RPT605320S, RPT606410S, RPT604452S, RPT605292S, RPT605309S, RPT605785S, RPT606148S, RPT606545S, RPT606590S, RPT606626S, RPT606855S, RPT607133S, RPT603642S, RPT603884S, RPT604096S, RPT605674S, RPT606157S, RPT602791S, RPT603841S, RPT605359S, RPT606341S, RPT606480S, RPT606694S, RPT607254S, RPT603571S, RPT603981S, RPT604320S, RPT607431S, RPT604895S, RPT606033S, RPT600820S, RPT600822S, RPT601223S, RPT601690S, RPT601766S, RPT602073S, RPT602253S, RPT602724S, RPT602789S, RPT603020S, RPT604965S, RPT604881S, RPT604300S, RPT606344S, RPT604733S, RPT603379S, RPT606190S, RPT607359S, RPT600625S, RPT603723S, RPT604120S, RPT606353S, RPT606310S, RPT606655S, RPT602185S, RPT602567S, RPT603727S, RPT603849S, RPT604233S, RPT600681S, RPT600946S, RPT602570S, RPT602936S, RPT603365S, RPT603939S, RPT605986S, RPT600785S, RPT602134S, RPT604142S, RPT604608S, RPT601446S, RPT603902S, RPT600553S, RPT600631S, RPT600918S, RPT601019S, RPT601076S, RPT601152S, RPT601304S, RPT601709S, RPT601736S, RPT601757S, RPT601918S, RPT602388S, RPT602486S, RPT602550S, RPT602553S, RPT604265S, RPT604266S, RPT604267S, RPT603819S, RPT605262S, RPT605263S, RPT601389S, RPT605692S, RPT606650S, RPT606677S, RPT606835S, RPT600166S, RPT602111S, RPT602273S, RPT602381S, RPT602660S, RPT602944S, RPT603534S, RPT603595S, RPT603609S, RPT603610S, RPT604091S, RPT605082S, RPT601011S, RPT606668S, RPT606035S, RPT605979S, RPT606908S, RPT605632S, RPT605801S, RPT606605S, RPT606913S, RPT603556S, RPT600984S, RPT603864S, RPT603891S, RPT605873S, RPT606408S, RPT607422S, RPT601463S, RPT602143S, RPT602192S, RPT600723S, RPT601128S, RPT602730S, RPT604214S, RPT604309S, RPT605185S, RPT605584S, RPT605594S, RPT606815S, RPT607018S, RPT607308S, RPT602543S, RPT604353S, RPT604363S, RPT602408S, RPT602409S, RPT605643S, RPT602867S, RPT603052S, RPT604622S, RPT605248S, RPT605355S, RPT600629S, RPT602422S, RPT604557S, RPT600600S, RPT605217S, RPT604122S, RPT606070S, RPT603112S, RPT603760S, RPT605022S, RPT605372S, RPT607386S, RPT604735S, RPT604860S, RPT606597S, RPT601184S, RPT603942S, RPT601257S, RPT601576S, RPT601786S, RPT602700S, RPT603823S, RPT601480S, RPT602288S, RPT602901S, RPT604729S, RPT604910S, RPT605023S, RPT600075S, RPT602007S, RPT603268S, RPT603269S, RPT603270S, RPT603276S, RPT603277S, RPT604308S, RPT606954S, RPT607147S, RPT605007S, RPT605448S, RPT604682S, RPT606454S, RPT606625S, RPT607711S, RPT604739S, RPT605055S, RPT605886S, RPT605985S, RPT606452S, RPT606779S, RPT607385S, RPT601075S, RPT602872S, RPT603118S, RPT600236S, RPT601685S, RPT604232S, RPT607001S, RPT600019S, RPT600576S, RPT602292S, RPT603504S, RPT604013S, RPT606486S, RPT606584S, RPT601768S, RPT601820S, RPT602559S, RPT602808S, RPT603281S, RPT604092S, RPT604158S, RPT604528S, RPT600844S, RPT601494S, RPT603576S, RPT605462S, RPT606436S, RPT601402S, RPT605085S, RPT605117S, RPT605579S, RPT606698S, RPT600437S, RPT600534S, RPT600729S, RPT601058S, RPT601242S, RPT604831S, RPT600072S, RPT602599S, RPT605513S, RPT605928S, RPT602496S, RPT602573S, RPT604107S, RPT604117S, RPT604156S, RPT600609S, RPT602254S, RPT603578S, RPT600602S, RPT600603S, RPT600604S, RPT600647S, RPT600779S, RPT601527S, RPT601775S, RPT601831S, RPT601876S, RPT601893S, RPT601999S, RPT603197S, RPT605927S, RPT606052S, RPT605590S, RPT600217S, RPT600870S, RPT601340S, RPT603941S, RPT600331S, RPT603791S, RPT606009S, RPT606993S, RPT602655S, RPT603054S, RPT606337S, RPT605926S, RPT606220S, RPT606604S, RPT600347S, RPT600772S, RPT600923S, RPT601030S, RPT601330S, RPT602754S, RPT602858S, RPT603132S, RPT603362S, RPT604372S, RPT602802S, RPT605459S, RPT603918S, RPT600875S, RPT601167S, RPT601490S, RPT602693S, RPT603093S, RPT604877S, RPT600559S, RPT604678S, RPT604885S, RPT605135S, RPT605219S, RPT605958S, RPT606818S, RPT606851S, RPT606926S, RPT607601S, RPT607659S, RPT601447S, RPT601756S, RPT602023S, RPT602431S, RPT602670S, RPT602800S, RPT604146S, RPT604433S, RPT604535S, RPT601081S, RPT606136S, RPT607136S, RPT602665S, RPT606251S, RPT603308S, RPT604465S, RPT604971S, RPT605012S, RPT605377S, RPT605379S, RPT605805S, RPT606206S, RPT606244S, RPT606246S, RPT606318S, RPT607002S, RPT601381S, RPT601872S, RPT601923S, RPT602055S, RPT602213S, RPT602745S, RPT603183S, RPT603895S, RPT603898S, RPT604311S, RPT606319S, RPT607241S, RPT607628S, RPT605416S, RPT605419S, RPT605428S, RPT605763S, RPT605791S, RPT600211S, RPT600707S, RPT600711S, RPT601159S, RPT601253S, RPT601354S, RPT602532S, RPT602658S, RPT602957S, RPT604174S, RPT607397S, RPT604967S, RPT605371S, RPT605373S, RPT606085S, RPT600464S, RPT601028S, RPT601735S, RPT601763S, RPT601968S, RPT602385S, RPT603095S, RPT603859S, RPT604349S, RPT605260S, RPT602619S, RPT604651S, RPT604941S, RPT605162S, RPT605899S, RPT605924S, RPT606018S, RPT606841S, RPT607193S, RPT607433S, RPT600570S, RPT600664S, RPT600839S, RPT603149S, RPT603356S, RPT603373S, RPT603420S, RPT603713S, RPT604574S, RPT605394S, RPT605761S, RPT605963S, RPT606010S, RPT606891S, RPT601116S, RPT601269S, RPT602639S, RPT602887S, RPT602906S, RPT602998S, RPT603521S, RPT603792S, RPT607555S, RPT607161S, RPT601487S, RPT602329S, RPT603265S, RPT603426S, RPT605361S, RPT605747S, RPT606291S, RPT604573S, RPT605182S, RPT605741S, RPT606983S, RPT600413S, RPT602489S, RPT603315S, RPT603664S, RPT606289S, RPT606471S, RPT606827S, RPT600093S, RPT603607S, RPT602392S, RPT602594S, RPT605699S, RPT607544S, RPT607593S, RPT601177S, RPT601258S, RPT602515S, RPT603159S, RPT603334S, RPT606510S, RPT601411S, RPT603682S, RPT605335S, RPT603633S, RPT604352S, RPT601810S, RPT602150S, RPT605408S, RPT605557S, RPT606101S, RPT606756S, RPT607257S, RPT600965S, RPT601233S, RPT602038S, RPT602166S, RPT607189S, RPT604891S, RPT600257S, RPT600296S, RPT600616S, RPT601488S, RPT601890S, RPT603014S, RPT600226S, RPT600737S, RPT600781S, RPT601084S, RPT601455S, RPT601502S, RPT601906S, RPT601907S, RPT607059S, RPT607060S, RPT600432S, RPT601721S, RPT604670S, RPT604685S, RPT606696S, RPT601035S, RPT601660S, RPT601828S, RPT601971S, RPT602686S, RPT604409S, RPT600532S, RPT605503S, RPT606133S, RPT606217S, RPT606642S, RPT607132S, RPT600448S, RPT601668S, RPT601673S, RPT601679S, RPT602274S, RPT603786S, RPT603787S, RPT603812S, RPT605637S, RPT603479S, RPT605411S, RPT605530S, RPT600649S, RPT600776S, RPT601819S, RPT604400S, RPT600760S, RPT602053S, RPT604712S, RPT600964S, RPT600972S, RPT603998S, RPT604278S, RPT603799S, RPT602202S, RPT605719S, RPT605729S, RPT605049S, RPT605934S, RPT600533S, RPT601728S, RPT602813S, RPT603562S, RPT604252S, RPT605933S, RPT606042S, RPT606463S, RPT604445S, RPT600546S, RPT600564S, RPT601273S, RPT601491S, RPT600345S, RPT601166S, RPT603189S, RPT603245S, RPT603386S, RPT603769S, RPT603832S, RPT604069S, RPT604173S, RPT605235S, RPT604691S, RPT606004S, RPT601672S, RPT604118S, RPT603339S, RPT600187S, RPT601087S, RPT601496S, RPT603671S, RPT603697S, RPT600628S, RPT600836S, RPT603660S, RPT601550S, RPT603536S, RPT600995S, RPT602928S, RPT603747S, RPT604687S, RPT605782S, RPT606007S, RPT607356S, RPT600085S, RPT600213S, RPT600790S, RPT600960S, RPT601230S, RPT601240S, RPT601691S, RPT601753S, RPT602014S, RPT602218S, RPT602249S, RPT602261S, RPT602669S, RPT603646S, RPT604451S, RPT607513S, RPT600708S, RPT604962S, RPT605412S, RPT605492S, RPT605591S, RPT600098S, RPT600689S, RPT601074S, RPT601248S, RPT601773S, RPT601878S, RPT601983S, RPT602049S, RPT602096S, RPT602535S, RPT602569S, RPT602571S, RPT602762S, RPT603310S, RPT603312S, RPT603349S, RPT603353S, RPT603478S, RPT603863S, RPT604524S, RPT607535S, RPT606164S, RPT606204S, RPT606876S, RPT606929S, RPT600880S, RPT601963S, RPT602695S, RPT602851S, RPT603068S, RPT604538S, RPT605670S, RPT606060S, RPT601243S, RPT601296S, RPT602008S, RPT602244S, RPT602471S, RPT602648S, RPT602904S, RPT603235S, RPT604039S, RPT604549S, RPT605697S, RPT605698S, RPT605875S, RPT605881S, RPT602063S, RPT602737S, RPT603142S, RPT603381S, RPT603587S, RPT604157S, RPT604723S, RPT604994S, RPT605941S, RPT604316S, RPT600597S, RPT605455S, RPT605461S, RPT605198S, RPT605447S, RPT605450S, RPT605553S, RPT606440S, RPT607162S, RPT607339S, RPT600563S, RPT600692S, RPT601317S, RPT601459S, RPT601514S, RPT601645S, RPT602028S, RPT602238S, RPT602765S, RPT602902S, RPT602994S, RPT603301S, RPT603622S, RPT603850S, RPT605465S, RPT607677S, RPT604897S, RPT604898S, RPT606302S, RPT606915S, RPT607329S, RPT600755S, RPT604430S, RPT600489S, RPT600792S, RPT601004S, RPT601077S, RPT601206S, RPT603588S, RPT603591S, RPT604023S, RPT604061S, RPT604063S, RPT604361S, RPT604410S, RPT604425S, RPT604440S, RPT604534S, RPT606519S, RPT604667S, RPT607463S, RPT606516S, RPT606582S, RPT606646S, RPT602601S, RPT602603S, RPT603090S, RPT603239S, RPT604339S, RPT604609S, RPT605454S, RPT606097S, RPT606833S, RPT600246S, RPT600326S, RPT600327S, RPT600447S, RPT600598S, RPT600798S, RPT600828S, RPT600866S, RPT601094S, RPT601124S, RPT601308S, RPT601915S, RPT602037S, RPT602158S, RPT602671S, RPT602738S, RPT602755S, RPT603113S, RPT603814S, RPT603936S, RPT604002S, RPT604771S, RPT605432S, RPT606773S, RPT601050S, RPT601353S, RPT602064S, RPT602228S, RPT602266S, RPT602397S, RPT602512S, RPT602727S, RPT602914S, RPT603058S, RPT603096S, RPT603928S, RPT604512S, RPT605593S, RPT605630S, RPT601063S, RPT600858S, RPT601386S, RPT602155S, RPT602647S, RPT604960S, RPT600212S, RPT600284S, RPT600385S, RPT600766S, RPT601275S, RPT601298S, RPT601301S, RPT601714S, RPT601936S, RPT602003S, RPT602177S, RPT602609S, RPT602395S, RPT605122S, RPT606573S, RPT605819S, RPT604646S, RPT604647S, RPT604752S, RPT605357S, RPT605512S, RPT605784S, RPT606162S, RPT606671S, RPT606854S, RPT602518S, RPT602812S, RPT602828S, RPT602941S, RPT603191S, RPT603940S, RPT604010S, RPT604346S, RPT606964S, RPT607223S, RPT607293S, RPT607238S, RPT605405S, RPT604848S, RPT606672S, RPT605446S, RPT606040S, RPT606049S, RPT600096S, RPT600254S, RPT601867S, RPT602797S, RPT603234S, RPT603553S, RPT603856S, RPT604305S, RPT604310S, RPT604843S, RPT605332S, RPT605378S, RPT600442S, RPT603991S, RPT604054S, RPT604175S, RPT604179S, RPT604222S, RPT604255S, RPT606863S, RPT606912S, RPT600219S, RPT600783S, RPT601131S, RPT601914S, RPT601996S, RPT602227S, RPT602374S, RPT603645S, RPT600309S, RPT600310S, RPT601604S, RPT601972S, RPT602275S, RPT606037S, RPT606082S, RPT602839S, RPT604745S, RPT605686S, RPT603399S, RPT604432S, RPT603887S, RPT604824S, RPT600339S, RPT601103S, RPT601541S, RPT601542S, RPT601692S, RPT602241S, RPT603889S, RPT604368S, RPT603001S, RPT606075S, RPT601215S, RPT604798S, RPT600397S, RPT606676S, RPT604797S, RPT603952S, RPT604082S, RPT604478S, RPT606384S, RPT605540S, RPT607571S, RPT600159S, RPT601727S, RPT607208S, RPT602380S, RPT607321S, RPT605391S, RPT606638S, RPT600255S, RPT600732S, RPT600733S, RPT600736S, RPT605640S, RPT605641S, RPT600607S, RPT600757S, RPT600761S, RPT604704S, RPT604832S, RPT605770S, RPT605822S, RPT606045S, RPT606551S, RPT606572S, RPT601433S, RPT601870S, RPT602163S, RPT602555S, RPT602633S, RPT602637S, RPT602805S, RPT603044S, RPT603385S, RPT603391S, RPT603860S, RPT606476S, RPT607473S, RPT607474S, RPT607670S, RPT605202S, RPT603813S, RPT605056S, RPT605578S, RPT605727S, RPT606537S, RPT600041S, RPT600475S, RPT601380S, RPT602837S, RPT602838S, RPT603043S, RPT604182S, RPT603433S, RPT604652S, RPT604732S, RPT604884S, RPT605244S, RPT605249S, RPT605256S, RPT605466S, RPT605760S, RPT606110S, RPT606142S, RPT606143S, RPT606144S, RPT607075S, RPT602077S, RPT602078S, RPT602250S, RPT602293S, RPT603869S, RPT603957S, RPT604007S, RPT607074S, RPT607186S, RPT607187S, RPT605017S, RPT605201S, RPT605483S, RPT600851S, RPT601225S, RPT601227S, RPT602551S, RPT604943S, RPT605254S, RPT606163S, RPT606202S, RPT604000S, RPT604056S, RPT605253S, RPT602540S, RPT602709S, RPT602772S, RPT603924S, RPT602024S, RPT603629S, RPT600582S, RPT602816S, RPT603766S, RPT604131S, RPT600051S, RPT605843S, RPT606324S, RPT607617S, RPT605031S, RPT604264S, RPT604808S, RPT600491S, RPT601070S, RPT601337S, RPT602736S, RPT603212S, RPT603305S, RPT601718S, RPT602432S, RPT605169S, RPT605187S, RPT605234S, RPT605748S, RPT605938S, RPT605955S, RPT606461S, RPT606462S, RPT600015S, RPT600078S, RPT600242S, RPT601112S, RPT601189S, RPT601376S, RPT601413S, RPT601606S, RPT601734S, RPT601940S, RPT602285S, RPT602530S, RPT602552S, RPT602726S, RPT602763S, RPT603214S, RPT603345S, RPT604058S, RPT604219S, RPT604508S, RPT604856S, RPT602146S, RPT602582S, RPT602799S, RPT604450S, RPT606630S, RPT605030S, RPT605898S, RPT603603S, RPT603616S, RPT600232S, RPT600350S, RPT601482S, RPT600906S, RPT606784S, RPT605426S, RPT606603S, RPT603133S, RPT607231S, RPT604814S, RPT600421S, RPT601800S, RPT602753S, RPT605015S, RPT607232S, RPT607239S, RPT607240S, RPT605567S, RPT605568S, RPT605714S, RPT606504S, RPT606739S, RPT600224S, RPT600567S, RPT600993S, RPT601270S, RPT601445S, RPT603293S, RPT603327S, RPT604307S, RPT603926S, RPT606721S, RPT607324S, RPT606198S, RPT601909S, RPT603357S, RPT600744S, RPT602677S, RPT602678S, RPT603151S, RPT603190S, RPT603231S, RPT604605S, RPT606847S, RPT607376S, RPT601472S, RPT603680S, RPT605347S, RPT605475S, RPT605571S, RPT605759S, RPT605798S, RPT606523S, RPT606736S, RPT607194S, RPT600678S, RPT603886S, RPT604209S, RPT604395S, RPT607195S, RPT601121S, RPT601247S, RPT605561S, RPT600677S, RPT602613S, RPT607037S, RPT602516S, RPT605081S, RPT600318S, RPT601002S, RPT604669S, RPT600461S, RPT601117S, RPT603440S, RPT606446S, RPT604767S, RPT600043S, RPT603124S, RPT604207S, RPT600277S, RPT604523S, RPT605059S, RPT605468S, RPT606300S, RPT606705S, RPT603915S, RPT606719S, RPT605431S, RPT606100S, RPT602231S, RPT605065S, RPT606697S, RPT602826S, RPT603768S, RPT605960S, RPT602511S, RPT603449S, RPT606278S, RPT606492S, RPT605470S, RPT601414S, RPT601434S, RPT601884S, RPT602224S, RPT602233S, RPT605828S, RPT606170S, RPT600191S, RPT603059S, RPT604073S, RPT606152S, RPT603023S, RPT601457S, RPT604366S, RPT606669S, RPT600333S, RPT606682S, RPT602132S, RPT603520S, RPT600451S, RPT603287S, RPT605585S, RPT605587S, RPT605607S, RPT606679S, RPT605190S, RPT605956S, RPT606963S, RPT607135S, RPT604987S, RPT605289S, RPT606066S, RPT606084S, RPT606086S, RPT606540S, RPT606675S, RPT601211S, RPT602419S, RPT603249S, RPT604337S, RPT604377S, RPT604394S, RPT607721S, RPT603712S, RPT603714S, RPT605418S, RPT603389S, RPT604896S, RPT606450S, RPT606557S, RPT600764S, RPT602277S, RPT602472S, RPT602806S, RPT603303S, RPT603375S, RPT604380S, RPT606920S, RPT607076S, RPT606700S, RPT605400S, RPT601072S, RPT607297S, RPT607581S, RPT607708S, RPT601593S, RPT602757S, RPT604421S, RPT600336S, RPT601096S, RPT601866S, RPT605322S, RPT605323S, RPT600291S, RPT603417S, RPT607227S, RPT607228S, RPT607274S, RPT600465S, RPT601021S, RPT601307S, RPT603084S, RPT603326S, RPT604642S, RPT605186S, RPT600145S, RPT600658S, RPT601793S, RPT602485S, RPT605514S, RPT605722S, RPT606149S, RPT606188S, RPT606568S, RPT601256S, RPT603903S, RPT604632S, RPT604899S, RPT605087S, RPT605098S, RPT605145S, RPT605146S, RPT605508S, RPT605602S, RPT605839S, RPT606074S, RPT606128S, RPT606265S, RPT606546S, RPT606547S, RPT606548S, RPT606549S, RPT606617S, RPT600395S, RPT600497S, RPT600902S, RPT601277S, RPT601585S, RPT601588S, RPT601778S, RPT601779S, RPT601857S, RPT601894S, RPT601899S, RPT602079S, RPT602093S, RPT602205S, RPT602493S, RPT603037S, RPT603161S, RPT603162S, RPT603230S, RPT603632S, RPT603673S, RPT603694S, RPT603695S, RPT603705S, RPT603706S, RPT603728S, RPT603736S, RPT603962S, RPT604087S, RPT604138S, RPT604412S, RPT604427S, RPT607405S, RPT603220S, RPT603913S, RPT600827S, RPT603352S, RPT604626S, RPT604627S, RPT604628S, RPT604643S, RPT604654S, RPT604880S, RPT605380S, RPT605830S, RPT605980S, RPT606311S, RPT606314S, RPT606670S, RPT601147S, RPT601393S, RPT601394S, RPT601555S, RPT601556S, RPT601861S, RPT601950S, RPT601961S, RPT602636S, RPT602853S, RPT602894S, RPT602897S, RPT602987S, RPT603405S, RPT603406S, RPT603782S, RPT604136S, RPT604169S, RPT606313S, RPT607094S, RPT607095S, RPT607096S, RPT607097S, RPT607098S, RPT607099S, RPT606848S, RPT602751S, RPT603335S, RPT603393S, RPT603599S, RPT603755S, RPT603781S, RPT604167S, RPT604281S, RPT604351S, RPT604371S, RPT604793S, RPT605121S, RPT606172S, RPT600474S, RPT600758S, RPT602846S, RPT605994S, RPT604692S, RPT607408S, RPT600416S, RPT604581S, RPT603408S, RPT603691S, RPT604834S, RPT605310S, RPT600665S, RPT600943S, RPT602822S, RPT606666S, RPT604892S, RPT606117S, RPT600949S, RPT601392S, RPT604721S, RPT605080S, RPT605520S, RPT605542S, RPT606053S, RPT606090S, RPT606443S, RPT606726S, RPT606727S, RPT606750S, RPT606751S, RPT607221S, RPT600998S, RPT602189S, RPT602547S, RPT602924S, RPT603320S, RPT603444S, RPT603893S, RPT604515S, RPT601917S, RPT605147S, RPT601250S, RPT602230S, RPT602344S, RPT602399S, RPT602403S, RPT602620S, RPT604416S, RPT600013S, RPT604940S, RPT602251S, RPT603764S, RPT604448S, RPT604838S, RPT602099S, RPT605399S, RPT600423S, RPT600476S, RPT601026S, RPT601202S, RPT603157S, RPT606081S, RPT606091S, RPT603419S, RPT605236S, RPT604764S, RPT605981S, RPT605092S, RPT605566S, RPT606016S, RPT605331S, RPT605497S, RPT606973S, RPT607062S, RPT600349S, RPT602294S, RPT604889S, RPT600388S, RPT604212S, RPT600487S, RPT601659S, RPT601336S, RPT604204S, RPT604958S, RPT605125S, RPT602215S, RPT600440S, RPT605019S, RPT605495S, RPT600837S, RPT600945S, RPT600977S, RPT601503S, RPT601646S, RPT601647S, RPT601671S, RPT603125S, RPT603803S, RPT605230S, RPT603552S, RPT605733S, RPT607300S, RPT600071S, RPT605724S, RPT602490S, RPT603321S, RPT600621S, RPT605605S, RPT603916S, RPT603917S, RPT600934S, RPT601259S, RPT602662S, RPT600338S, RPT600390S, RPT605267S, RPT606048S, RPT606522S, RPT607640S, RPT605181S, RPT606763S, RPT606793S, RPT607027S, RPT607366S, RPT607382S, RPT604314S, RPT606530S, RPT606622S, RPT606911S, RPT601579S, RPT603100S, RPT603152S, RPT604089S, RPT604090S, RPT604100S, RPT607023S, RPT605620S, RPT604041S, RPT604837S, RPT604849S, RPT606785S, RPT600745S, RPT601838S, RPT602051S, RPT602596S, RPT603914S, RPT605509S, RPT607056S, RPT604722S, RPT604835S, RPT604842S, RPT605095S, RPT601493S, RPT602948S, RPT603809S, RPT604134S, RPT604429S, RPT605094S, RPT606027S, RPT606387S, RPT606561S, RPT600482S, RPT601016S, RPT601368S, RPT601693S, RPT601758S, RPT602429S, RPT602649S, RPT602790S, RPT602913S, RPT602989S, RPT603953S, RPT604113S, RPT604263S, RPT604531S, RPT604655S, RPT604846S, RPT601137S, RPT603412S, RPT603686S, RPT604604S, RPT604606S, RPT604607S, RPT604746S, RPT604969S, RPT605102S, RPT605103S, RPT605104S, RPT605105S, RPT606587S, RPT606588S, RPT606627S, RPT606948S, RPT607013S, RPT607091S, RPT607103S, RPT607319S, RPT607675S, RPT600867S, RPT600872S, RPT601015S, RPT601060S, RPT601140S, RPT603775S, RPT604020S, RPT604460S, RPT604461S, RPT604690S, RPT604696S, RPT605152S, RPT600470S, RPT601549S, RPT601729S, RPT602257S, RPT603409S, RPT603640S, RPT603905S, RPT604903S, RPT605415S, RPT607471S, RPT600551S, RPT601637S, RPT601934S, RPT602436S, RPT602536S, RPT602767S, RPT603394S, RPT603477S, RPT603772S, RPT604170S, RPT607484S, RPT607493S, RPT603638S, RPT605044S, RPT606701S, RPT606702S, RPT601283S, RPT603164S, RPT604059S, RPT604784S, RPT604246S, RPT605020S, RPT601108S, RPT601862S, RPT600606S, RPT604572S, RPT604957S, RPT605269S, RPT605948S, RPT605951S, RPT606399S, RPT606401S, RPT606485S, RPT607550S, RPT601677S, RPT601823S, RPT602832S, RPT603388S, RPT604537S, RPT604613S, RPT604975S, RPT605033S, RPT605312S, RPT605755S, RPT606236S, RPT606491S, RPT606596S, RPT607595S, RPT601484S, RPT602076S, RPT602156S, RPT602760S, RPT602907S, RPT602975S, RPT603051S, RPT607145S, RPT604256S, RPT600648S, RPT601651S, RPT602683S, RPT602685S, RPT603401S, RPT604533S, RPT604939S, RPT603618S, RPT603797S, RPT605288S, RPT607111S, RPT601626S, RPT603372S, RPT600023S, RPT604805S, RPT605407S, RPT606880S, RPT600402S, RPT602135S, RPT602248S, RPT602507S, RPT604347S, RPT604506S, RPT606713S, RPT600094S, RPT603430S, RPT605160S, RPT606025S, RPT607459S, RPT601255S, RPT602439S, RPT602931S, RPT607539S, RPT604874S, RPT604185S, RPT607131S, RPT606192S, RPT606345S, RPT600966S, RPT601104S, RPT601162S, RPT601792S, RPT602707S, RPT603199S, RPT604340S, RPT600512S, RPT602679S, RPT602696S, RPT603346S, RPT605268S, RPT603938S, RPT600419S, RPT602097S, RPT603597S, RPT603601S, RPT605586S, RPT603976S, RPT604080S, RPT604186S, RPT604205S, RPT604220S, RPT604270S, RPT604583S, RPT602694S, RPT603531S, RPT604108S, RPT607467S, RPT600435S, RPT602393S, RPT601261S, RPT601505S, RPT602529S, RPT601316S, RPT605224S, RPT606173S, RPT606748S, RPT607250S, RPT600444S, RPT601109S, RPT602343S, RPT602710S, RPT602870S, RPT603178S, RPT603196S, RPT603207S, RPT603523S, RPT603600S, RPT603729S, RPT604045S, RPT605043S, RPT606888S, RPT600697S, RPT604003S, RPT604040S, RPT606080S, RPT606689S, RPT607211S, RPT603427S, RPT603845S, RPT604032S, RPT600734S, RPT600814S, RPT603165S, RPT601898S, RPT603726S, RPT603967S, RPT601633S, RPT602708S, RPT603244S, RPT604657S, RPT604661S, RPT604700S, RPT604727S, RPT604742S, RPT604795S, RPT605164S, RPT605456S, RPT605635S, RPT605789S, RPT605790S, RPT605939S, RPT606623S, RPT606624S, RPT606776S, RPT607148S, RPT601038S, RPT601176S, RPT602123S, RPT602179S, RPT602182S, RPT602183S, RPT602193S, RPT602666S, RPT602674S, RPT603492S, RPT603582S, RPT604345S, RPT605276S, RPT605600S, RPT604335S, RPT605471S, RPT605488S, RPT601118S, RPT601241S, RPT601287S, RPT601522S, RPT601835S, RPT603730S, RPT603906S, RPT603937S, RPT605734S, RPT605929S, RPT600769S, RPT600970S, RPT602384S, RPT602612S, RPT606667S, RPT602861S, RPT601409S, RPT604867S, RPT604871S, RPT604872S, RPT605396S, RPT605397S, RPT605978S, RPT606076S, RPT606269S, RPT606270S, RPT606610S, RPT600805S, RPT600810S, RPT600818S, RPT601114S, RPT601384S, RPT602450S, RPT602499S, RPT602623S, RPT602769S, RPT603248S, RPT603250S, RPT603482S, RPT603485S, RPT603619S, RPT607032S, RPT600630S, RPT601101S, RPT605042S, RPT605529S, RPT600188S, RPT604904S, RPT606478S, RPT606475S, RPT604776S, RPT602479S, RPT603581S, RPT600048S, RPT600634S, RPT602025S, RPT604529S, RPT606439S, RPT602206S, RPT603907S, RPT601069S, RPT603218S, RPT603666S, RPT605054S, RPT600793S, RPT601364S, RPT601628S, RPT601874S, RPT601881S, RPT604397S, RPT606560S, RPT601747S, RPT605467S, RPT605803S, RPT606514S, RPT602260S, RPT602918S, RPT603808S, RPT605151S, RPT604639S, RPT605251S, RPT605592S, RPT600592S, RPT603904S, RPT604444S, RPT604486S, RPT606079S, RPT600427S, RPT601543S, RPT602542S, RPT603855S, RPT604172S, RPT605402S, RPT600656S, RPT600848S, RPT601385S, RPT601657S, RPT601687S, RPT602416S, RPT602437S, RPT602462S, RPT604878S, RPT601785S, RPT603513S, RPT605111S, RPT605195S, RPT605272S, RPT605453S, RPT605922S, RPT606271S, RPT606298S, RPT607125S, RPT607322S, RPT601224S, RPT601325S, RPT603818S, RPT606336S, RPT604730S, RPT604738S, RPT606014S, RPT602181S, RPT602993S, RPT603292S, RPT600531S, RPT602610S, RPT604656S, RPT605825S, RPT607049S, RPT601065S, RPT601743S, RPT602919S, RPT605989S, RPT606229S, RPT606806S, RPT607218S, RPT600014S, RPT601610S, RPT602630S, RPT604755S, RPT604979S, RPT600797S, RPT604915S, RPT605874S, RPT606725S, RPT603063S, RPT604611S, RPT605008S, RPT605009S, RPT600280S, RPT600777S, RPT601897S, RPT602933S, RPT603064S, RPT603067S, RPT603077S, RPT603572S, RPT603583S, RPT603624S, RPT603741S, RPT603873S, RPT603888S, RPT605649S, RPT602997S, RPT605776S, RPT606564S, RPT600992S, RPT603056S, RPT605227S, RPT600266S, RPT601332S, RPT601708S, RPT603555S, RPT600231S, RPT600540S, RPT600762S, RPT605213S, RPT605554S, RPT601066S, RPT604920S, RPT605290S, RPT605366S, RPT606483S, RPT606636S, RPT606839S, RPT606902S, RPT607184S, RPT600833S, RPT600963S, RPT601955S, RPT602113S, RPT602577S, RPT602728S, RPT602868S, RPT603019S, RPT603097S, RPT604078S, RPT605689S, RPT606754S, RPT606925S, RPT602681S, RPT603253S, RPT604966S, RPT605959S, RPT606632S, RPT607207S, RPT603890S, RPT603086S, RPT603322S, RPT603880S, RPT604064S, RPT604378S, RPT600227S, RPT600457S, RPT605348S, RPT605349S, RPT600288S, RPT603735S, RPT604857S, RPT604865S, RPT605908S, RPT606114S, RPT606577S, RPT606992S, RPT601090S, RPT601839S, RPT601840S, RPT601990S, RPT602855S, RPT604033S, RPT604472S, RPT604852S, RPT604853S, RPT604855S, RPT605074S, RPT605075S, RPT605173S, RPT605350S, RPT605351S, RPT606428S, RPT606506S, RPT603039S, RPT603042S, RPT603047S, RPT603048S, RPT603524S, RPT603525S, RPT603925S, RPT605671S, RPT601638S, RPT605283S, RPT604046S, RPT600514S, RPT602549S, RPT602039S, RPT605912S, RPT605148S, RPT605565S, RPT601674S, RPT602357S, RPT603653S, RPT603654S, RPT603699S, RPT605225S, RPT601244S, RPT601705S, RPT601739S, RPT603628S, RPT604521S, RPT605518S, RPT605526S, RPT600431S, RPT600433S, RPT602864S, RPT603758S, RPT603839S, RPT603840S, RPT604190S, RPT604398S, RPT605708S, RPT605802S, RPT601344S, RPT601450S, RPT601787S, RPT602191S, RPT603648S, RPT603751S, RPT604119S, RPT605166S, RPT600787S, RPT601328S, RPT602892S, RPT604873S, RPT606057S, RPT606151S, RPT606657S, RPT600251S, RPT600342S, RPT601976S, RPT602127S, RPT602235S, RPT602355S, RPT602413S, RPT602438S, RPT602875S, RPT603080S, RPT603318S, RPT603731S, RPT603846S, RPT604526S, RPT603040S, RPT600911S, RPT601315S, RPT601370S, RPT602850S, RPT603285S, RPT607022S, RPT607710S, RPT604560S, RPT600430S, RPT600854S, RPT603257S, RPT604701S, RPT604976S, RPT605613S, RPT606012S, RPT600438S, RPT600655S, RPT601546S, RPT601713S, RPT602840S, RPT603445S, RPT603807S, RPT604336S, RPT604456S, RPT605735S, RPT605858S, RPT603228S, RPT604618S, RPT605624S, RPT605829S, RPT606512S, RPT606513S, RPT606653S, RPT606686S, RPT606811S, RPT606896S, RPT606897S, RPT606931S, RPT607004S, RPT607065S, RPT607275S, RPT600524S, RPT600780S, RPT601053S, RPT601249S, RPT601717S, RPT601865S, RPT602046S, RPT602291S, RPT603034S, RPT603102S, RPT603383S, RPT603384S, RPT603773S, RPT603774S, RPT604098S, RPT604160S, RPT604178S, RPT604229S, RPT604342S, RPT607543S, RPT606643S, RPT601089S, RPT601759S, RPT602382S, RPT602775S, RPT604420S, RPT600699S, RPT607674S, RPT606195S, RPT606241S, RPT606245S, RPT603134S, RPT604357S, RPT605157S, RPT606641S, RPT606864S, RPT601507S, RPT604980S, RPT600544S, RPT601187S, RPT605221S, RPT600989S, RPT601197S, RPT602815S, RPT604517S, RPT605544S, RPT604917S, RPT604924S, RPT604931S, RPT604936S, RPT604937S, RPT601055S, RPT601795S, RPT601798S, RPT601799S, RPT601804S, RPT603544S, RPT603545S, RPT603549S, RPT603559S, RPT603564S, RPT604015S, RPT604016S, RPT604271S, RPT604466S, RPT604467S, RPT604468S, RPT604469S, RPT604925S, RPT604935S, RPT606111S, RPT606112S, RPT606113S, RPT606115S, RPT606304S, RPT606559S, RPT606575S, RPT600021S, RPT600031S, RPT600038S, RPT601031S, RPT601154S, RPT602094S, RPT602642S, RPT602689S, RPT602782S, RPT602912S, RPT602961S, RPT603110S, RPT603139S, RPT603140S, RPT603144S, RPT603436S, RPT604210S, RPT604211S, RPT604213S, RPT606026S, RPT606036S, RPT607432S, RPT600975S, RPT603106S, RPT600201S, RPT601350S, RPT604221S, RPT606124S, RPT604702S, RPT605016S, RPT605795S, RPT600480S, RPT601562S, RPT603995S, RPT604359S, RPT602779S, RPT603852S, RPT600498S, RPT601431S, RPT601920S, RPT604446S, RPT600612S, RPT605682S, RPT601309S, RPT601632S, RPT603280S, RPT603605S, RPT606950S, RPT606683S, RPT600403S, RPT602714S, RPT602803S, RPT603659S, RPT605973S, RPT606182S, RPT600941S, RPT600979S, RPT601453S, RPT601461S, RPT602731S, RPT602749S, RPT602820S, RPT602985S, RPT603584S, RPT603870S, RPT603871S, RPT603872S, RPT603883S, RPT605106S, RPT605045S, RPT603411S, RPT605232S, RPT605639S, RPT602190S, RPT602608S, RPT604066S, RPT606153S, RPT606599S, RPT606699S, RPT606331S, RPT606609S, RPT602807S, RPT602905S, RPT603722S, RPT605826S, RPT600035S, RPT601168S, RPT601444S, RPT603745S, RPT604094S, RPT606853S, RPT600371S, RPT600373S, RPT604161S, RPT605237S, RPT602105S, RPT605434S, RPT602044S, RPT602517S, RPT603429S, RPT604240S, RPT600052S, RPT604504S, RPT605473S, RPT606005S, RPT602074S, RPT602865S, RPT603692S, RPT604396S, RPT604464S, RPT605601S, RPT606606S, RPT607377S, RPT602849S, RPT600054S, RPT600469S, RPT603999S, RPT605534S, RPT605194S, RPT605321S, RPT606247S, RPT604356S, RPT606103S, RPT600420S, RPT603241S, RPT603289S, RPT604034S, RPT605123S, RPT605968S, RPT600088S, RPT600228S, RPT602207S, RPT607117S, RPT606467S, RPT606520S, RPT606928S, RPT601454S, RPT605523S, RPT605436S, RPT605936S, RPT606043S, RPT606120S, RPT606255S, RPT606500S, RPT606565S, RPT606786S, RPT606892S, RPT601188S, RPT602514S, RPT604243S, RPT604789S, RPT604794S, RPT605209S, RPT605482S, RPT606268S, RPT606372S, RPT607053S, RPT601071S, RPT601361S, RPT601518S, RPT601656S, RPT602157S, RPT602688S, RPT603238S, RPT603499S, RPT603500S, RPT603932S, RPT603956S, RPT604152S, RPT607646S, RPT604757S, RPT605097S, RPT600374S, RPT605027S, RPT606233S, RPT606340S, RPT606529S, RPT601231S, RPT603079S, RPT603643S, RPT603788S, RPT605107S, RPT605563S, RPT606844S, RPT601180S, RPT601662S, RPT607639S, RPT604758S, RPT605066S, RPT605177S, RPT605368S, RPT605543S, RPT605599S, RPT605713S, RPT606077S, RPT606297S, RPT606745S, RPT606800S, RPT606901S, RPT606903S, RPT606994S, RPT607580S, RPT600459S, RPT601725S, RPT601788S, RPT601858S, RPT602245S, RPT602387S, RPT602589S, RPT602831S, RPT603626S, RPT603700S, RPT603763S, RPT603771S, RPT603793S, RPT603885S, RPT604081S, RPT604097S, RPT604760S, RPT604819S, RPT604911S, RPT605003S, RPT605067S, RPT605344S, RPT605783S, RPT607592S, RPT607629S, RPT601545S, RPT601616S, RPT601774S, RPT602083S, RPT603200S, RPT603311S, RPT603982S, RPT604021S, RPT604044S, RPT604215S, RPT604520S, RPT604580S, RPT604902S, RPT604963S, RPT605257S, RPT606553S, RPT600537S, RPT601613S, RPT601745S, RPT602199S, RPT606777S, RPT603530S, RPT600919S, RPT604348S, RPT604561S, RPT602854S, RPT602341S, RPT603155S, RPT603364S, RPT600097S, RPT607328S, RPT600252S, RPT600571S, RPT604525S, RPT604085S, RPT604812S, RPT604938S, RPT604956S, RPT603176S, RPT607488S, RPT604929S, RPT602194S, RPT600650S, RPT600982S, RPT606921S, RPT603203S, RPT603298S, RPT603983S, RPT604540S, RPT607536S, RPT605120S, RPT602452S, RPT603035S, RPT603946S, RPT606631S, RPT603518S, RPT606456S, RPT605438S, RPT600046S, RPT600832S, RPT602740S, RPT606895S, RPT606615S, RPT604262S, RPT603838S, RPT604635S, RPT605837S, RPT604132S, RPT607362S, RPT600662S, RPT601817S, RPT601992S, RPT602128S, RPT600253S, RPT602369S, RPT604498S, RPT606047S, RPT605370S, RPT601286S, RPT601805S, RPT602497S, RPT602873S, RPT602983S, RPT603116S, RPT603172S, RPT603313S, RPT605489S, RPT606984S, RPT600269S, RPT603258S, RPT603259S, RPT603615S, RPT604991S, RPT601524S, RPT603246S, RPT605988S, RPT606039S, RPT607003S, RPT602528S, RPT603447S, RPT603527S, RPT604982S, RPT600684S, RPT604555S, RPT603053S, RPT605420S, RPT605718S, RPT605723S, RPT603753S, RPT604164S, RPT604165S, RPT605417S, RPT605425S, RPT602184S, RPT606181S, RPT607311S, RPT605291S, RPT606329S, RPT604796S, RPT600809S, RPT602315S, RPT602712S, RPT603181S, RPT603242S, RPT603969S, RPT604123S, RPT604153S, RPT604206S, RPT605707S, RPT605967S, RPT600057S, RPT600376S, RPT601326S, RPT602009S, RPT603794S, RPT604863S, RPT605345S, RPT605850S, RPT605905S, RPT605909S, RPT601346S, RPT601347S, RPT601349S, RPT602857S, RPT603206S, RPT603489S, RPT604124S, RPT604126S, RPT604127S, RPT604128S, RPT607468S, RPT607482S, RPT601091S, RPT601092S, RPT601093S, RPT605421S, RPT605720S, RPT606495S, RPT603528S, RPT601912S, RPT607265S, RPT604800S, RPT604811S, RPT605369S, RPT605498S, RPT605753S, RPT606104S, RPT606418S, RPT606542S, RPT606543S, RPT606544S, RPT607078S, RPT600235S, RPT601067S, RPT601634S, RPT601635S, RPT601670S, RPT602088S, RPT602759S, RPT603000S, RPT603115S, RPT607260S, RPT603055S, RPT603062S, RPT603836S, RPT605090S, RPT607420S, RPT605441S, RPT606175S, RPT600032S, RPT600789S, RPT600950S, RPT602211S, RPT602852S, RPT603472S, RPT603651S, RPT603765S, RPT604285S, RPT604389S, RPT604438S, RPT604536S, RPT603229S, RPT603551S, RPT604004S, RPT604076S, RPT604384S, RPT601501S, RPT601198S, RPT602401S, RPT602402S, RPT600770S, RPT604612S, RPT604637S, RPT604640S, RPT604644S, RPT604645S, RPT604648S, RPT604649S, RPT605856S, RPT605863S, RPT605870S, RPT605871S, RPT605879S, RPT600299S, RPT600300S, RPT600303S, RPT600304S, RPT600306S, RPT600307S, RPT600308S, RPT601558S, RPT601559S, RPT601560S, RPT601561S, RPT601564S, RPT601565S, RPT601566S, RPT601568S, RPT601569S, RPT601570S, RPT601571S, RPT601572S, RPT601574S, RPT602141S, RPT602142S, RPT602453S, RPT602455S, RPT602456S, RPT602457S, RPT602463S, RPT602466S, RPT602467S, RPT602544S, RPT602885S, RPT602891S, RPT602893S, RPT602895S, RPT602896S, RPT602903S, RPT602911S, RPT602940S, RPT602966S, RPT602967S, RPT602970S, RPT602971S, RPT602972S, RPT602973S, RPT602974S, RPT602976S, RPT602977S, RPT602978S, RPT602979S, RPT602981S, RPT605854S, RPT605855S, RPT605857S, RPT605859S, RPT605860S, RPT605862S, RPT605864S, RPT605866S, RPT605867S, RPT605869S, RPT605872S, RPT605876S, RPT605877S, RPT605878S, RPT605880S, RPT605882S, RPT605883S, RPT606348S, RPT606349S, RPT606350S, RPT606351S, RPT606352S, RPT606354S, RPT606355S, RPT606356S, RPT606357S, RPT606358S, RPT606359S, RPT606360S, RPT606361S, RPT606362S, RPT606363S, RPT606364S, RPT606365S, RPT606366S, RPT606368S, RPT606369S, RPT604680S, RPT604750S, RPT604751S, RPT604762S, RPT604766S, RPT604775S, RPT604791S, RPT604998S, RPT605271S, RPT605319S, RPT605353S, RPT605672S, RPT606011S, RPT606099S, RPT606123S, RPT600147S, RPT600282S, RPT600301S, RPT600515S, RPT600973S, RPT601182S, RPT601196S, RPT601464S, RPT601466S, RPT601467S, RPT601468S, RPT601563S, RPT602200S, RPT602464S, RPT602465S, RPT602715S, RPT602825S, RPT604376S, RPT604748S, RPT606860S, RPT606861S, RPT606862S, RPT607138S, RPT607140S, RPT607141S, RPT607289S, RPT600915S, RPT601046S, RPT601710S, RPT604830S, RPT602718S, RPT602725S, RPT604074S, RPT605925S, RPT600816S, RPT601582S, RPT602252S, RPT602752S, RPT603506S, RPT605261S, RPT604921S, RPT604992S, RPT605460S, RPT606469S, RPT602396S, RPT602626S, RPT602732S, RPT603283S, RPT603843S, RPT603868S, RPT605172S, RPT605004S, RPT605376S, RPT600716S, RPT602203S, RPT606521S, RPT605144S, RPT600486S, RPT601418S, RPT601704S, RPT602271S, RPT602676S, RPT602801S, RPT603879S, RPT604224S, RPT604981S, RPT605439S, RPT602204S, RPT603369S, RPT607440S, RPT600279S, RPT600593S, RPT601047S, RPT601403S, RPT602888S, RPT603590S, RPT604417S, RPT604810S, RPT601098S, RPT602579S, RPT602480S, RPT600082S, RPT600083S, RPT600124S, RPT604851S, RPT605384S, RPT602398S, RPT607082S, RPT603004S, RPT603271S, RPT603539S, RPT604393S, RPT603011S, RPT604166S, RPT600172S, RPT600225S, RPT600234S, RPT601410S, RPT603358S, RPT606374S, RPT606376S, RPT607157S, RPT601150S, RPT602219S, RPT604424S, RPT603740S, RPT603343S, RPT606460S, RPT606552S, RPT606558S, RPT607340S, RPT600843S, RPT601324S, RPT601822S, RPT602269S, RPT605356S, RPT605688S, RPT605945S, RPT602188S, RPT601853S, RPT602282S, RPT603260S, RPT603261S, RPT604770S, RPT605301S, RPT606023S, RPT601460S, RPT604382S, RPT606732S, RPT606924S, RPT602186S, RPT605076S, RPT604737S, RPT606421S, RPT604717S, RPT605440S, RPT605842S, RPT606744S, RPT604227S, RPT604358S, RPT606859S, RPT606822S, RPT605140S, RPT602739S, RPT600029S, RPT600869S, RPT604582S, RPT604686S, RPT605545S, RPT605838S, RPT606532S, RPT606807S, RPT606900S, RPT607077S, RPT600452S, RPT600645S, RPT601339S, RPT603665S, RPT603717S, RPT603973S, RPT604244S, RPT607591S, RPT600575S, RPT603119S, RPT603475S, RPT605286S, RPT605494S, RPT606159S, RPT606177S, RPT606219S, RPT606554S, RPT606629S, RPT600579S, RPT600580S, RPT600747S, RPT600826S, RPT602722S, RPT602932S, RPT606444S, RPT606881S, RPT607200S, RPT607272S, RPT600577S, RPT600617S, RPT600921S, RPT600971S, RPT601144S, RPT601622S, RPT601695S, RPT602001S, RPT602148S, RPT603126S, RPT603297S, RPT604546S, RPT604988S, RPT605136S, RPT606316S, RPT606731S, RPT600352S, RPT601291S, RPT602871S, RPT603013S, RPT603015S, RPT604317S, RPT605703S, RPT600265S, RPT605787S, RPT605695S, RPT601808S, RPT605011S, RPT606637S, RPT603945S, RPT605069S, RPT605238S, RPT601295S, RPT602747S, RPT602949S, RPT604677S, RPT604694S, RPT604781S, RPT605091S, RPT605612S, RPT606061S, RPT606064S, RPT606488S, RPT606717S, RPT607015S, RPT600411S, RPT601146S, RPT602533S, RPT602741S, RPT603620S, RPT603687S, RPT604147S, RPT604241S, RPT604297S, RPT604556S, RPT606932S, RPT601323S, RPT603148S, RPT605252S, RPT606820S, RPT606982S, RPT604075S, RPT604949S, RPT605476S, RPT606277S, RPT606468S, RPT606808S, RPT606987S, RPT607446S, RPT607454S, RPT600633S, RPT601220S, RPT601631S, RPT601680S, RPT601750S, RPT602468S, RPT602483S, RPT603240S, RPT606608S, RPT602110S, RPT603146S, RPT600812S, RPT603402S, RPT604184S, RPT604664S, RPT606834S, RPT602672S, RPT602947S, RPT603392S, RPT603661S, RPT607562S, RPT607374S, RPT605711S, RPT600850S, RPT605255S, RPT603529S, RPT601160S, RPT603263S, RPT605642S, RPT606093S, RPT600229S, RPT600434S, RPT601744S, RPT601126S, RPT606257S, RPT604754S, RPT604901S, RPT602208S, RPT603045S, RPT602510S, RPT606877S, RPT606878S, RPT603820S, RPT602065S, RPT604876S, RPT604955S, RPT605226S, RPT605901S, RPT606288S, RPT606887S, RPT602461S, RPT602691S, RPT603156S, RPT604552S, RPT605334S, RPT605765S, RPT605996S, RPT600445S, RPT605207S, RPT601544S, RPT605437S, RPT606691S, RPT605744S, RPT603696S, RPT604684S, RPT605685S, RPT600788S, RPT601136S, RPT602955S, RPT602956S, RPT604707S, RPT605499S, RPT606063S, RPT607393S, RPT600492S, RPT600874S, RPT601551S, RPT603681S, RPT603709S, RPT604031S, RPT604399S, RPT604449S, RPT600401S, RPT602018S, RPT603437S, RPT600825S, RPT603425S, RPT600407S, RPT600589S, RPT605940S, RPT602027S, RPT605983S, RPT605815S, RPT607007S, RPT601362S, RPT602026S, RPT602058S, RPT602650S, RPT602692S, RPT602758S, RPT602917S, RPT603382S, RPT601724S, RPT602100S, RPT603456S, RPT605330S, RPT607426S, RPT600151S, RPT600164S, RPT605749S, RPT600170S, RPT600289S, RPT600363S, RPT600784S, RPT601018S, RPT602434S, RPT602937S, RPT603262S, RPT603329S, RPT605896S, RPT601553S, RPT602350S, RPT603996S, RPT604223S, RPT606243S, RPT607369S, RPT605546S, RPT605834S, RPT605836S, RPT603968S, RPT600738S, RPT601883S, RPT602474S, RPT603111S, RPT604141S, RPT606166S, RPT603623S, RPT606645S, RPT607323S, RPT603415S, RPT603847S, RPT604143S, RPT605165S, RPT607026S, RPT600268S, RPT602953S, RPT603002S, RPT603875S, RPT604364S, RPT600741S, RPT600753S, RPT600803S, RPT601312S, RPT601719S, RPT600067S, RPT600774S, RPT600243S, RPT601932S, RPT604071S, RPT600009S, RPT601977S, RPT601978S, RPT602963S, RPT602964S, RPT602965S, RPT603592S, RPT600638S, RPT601097S, RPT601974S, RPT602040S, RPT602280S, RPT602984S, RPT603806S, RPT604367S, RPT601533S, RPT603554S, RPT604833S, RPT601754S, RPT605500S, RPT607253S, RPT606261S, RPT607070S, RPT607167S, RPT603041S, RPT604900S, RPT605200S, RPT606248S, RPT602152S, RPT603081S, RPT602101S, RPT603954S, RPT603955S, RPT600917S, RPT605608S, RPT606154S, RPT601794S, RPT602011S, RPT602151S, RPT602742S, RPT604565S, RPT604633S, RPT604634S, RPT604740S, RPT604951S, RPT604999S, RPT605035S, RPT605101S, RPT605952S, RPT606161S, RPT606309S, RPT600439S, RPT600458S, RPT600538S, RPT600657S, RPT600675S, RPT600713S, RPT600731S, RPT600786S, RPT601123S, RPT601676S, RPT602406S, RPT602509S, RPT602534S, RPT602702S, RPT602704S, RPT603497S, RPT603533S, RPT603565S, RPT603566S, RPT603567S, RPT600615S, RPT600988S, RPT601782S, RPT602178S, RPT602289S, RPT602411S, RPT602607S, RPT602634S, RPT602656S, RPT602910S, RPT602935S, RPT603175S, RPT603290S, RPT603330S, RPT607655S, RPT604505S, RPT605970S, RPT601998S, RPT603690S, RPT607112S, RPT607113S, RPT607114S, RPT607115S, RPT607116S, RPT607409S, RPT607410S, RPT600192S, RPT600642S, RPT601318S, RPT601531S, RPT602299S, RPT602711S, RPT603639S, RPT604479S, RPT604616S, RPT604617S, RPT604697S, RPT605904S, RPT606022S, RPT606213S, RPT600370S, RPT600400S, RPT600717S, RPT600961S, RPT601449S, RPT601621S, RPT601726S, RPT601731S, RPT601995S, RPT602052S, RPT602348S, RPT602444S, RPT602824S, RPT602884S, RPT603089S, RPT603129S, RPT603130S, RPT603131S, RPT603135S, RPT604030S, RPT604513S, RPT604470S, RPT605918S, RPT604887S, RPT605231S, RPT605647S, RPT600808S, RPT601814S, RPT601855S, RPT602071S, RPT603933S, RPT604402S, RPT604480S, RPT600842S, RPT600857S, RPT600860S, RPT600863S, RPT600864S, RPT600868S, RPT600878S, RPT600910S, RPT600912S, RPT602108S, RPT602494S, RPT602946S, RPT602951S, RPT604650S, RPT606797S, RPT600542S, RPT601500S, RPT601641S, RPT601841S, RPT601882S, RPT603031S, RPT603032S, RPT603837S, RPT602279S, RPT607380S, RPT601772S, RPT602201S, RPT605128S, RPT605693S, RPT601661S, RPT604344S, RPT602133S, RPT604765S, RPT602588S, RPT604294S, RPT605403S, RPT605736S, RPT606442S, RPT603237S, RPT606191S, RPT601587S, RPT601723S, RPT601864S, RPT602047S, RPT602080S, RPT602469S, RPT602606S, RPT604968S, RPT604953S, RPT605184S, RPT605479S, RPT605923S, RPT605987S, RPT606484S, RPT606593S, RPT606644S, RPT606737S, RPT600504S, RPT602586S, RPT602880S, RPT602882S, RPT603166S, RPT603252S, RPT603264S, RPT603266S, RPT603558S, RPT604183S, RPT605583S, RPT606879S, RPT602210S, RPT602587S, RPT603344S, RPT605634S, RPT606798S, RPT600329S, RPT606789S, RPT601619S, RPT602004S, RPT602356S, RPT602874S, RPT601441S, RPT606856S, RPT604139S, RPT606427S, RPT605113S, RPT607434S, RPT604564S, RPT604695S, RPT604772S, RPT604817S, RPT604928S, RPT604983S, RPT605617S, RPT605717S, RPT605818S, RPT606019S, RPT606119S, RPT606326S, RPT606614S, RPT607063S, RPT607313S, RPT607436S, RPT600311S, RPT600450S, RPT600771S, RPT601358S, RPT601375S, RPT601530S, RPT602295S, RPT602562S, RPT602844S, RPT603025S, RPT603057S, RPT603192S, RPT603211S, RPT603251S, RPT603304S, RPT603307S, RPT603448S, RPT603894S, RPT604093S, RPT604198S, RPT604392S, RPT607490S, RPT606960S, RPT606275S, RPT602548S, RPT604499S, RPT604500S, RPT606955S, RPT601815S, RPT603342S, RPT606189S, RPT602883S, RPT602006S, RPT600260S, RPT600348S, RPT606602S, RPT606165S, RPT603108S, RPT603109S, RPT604318S, RPT602898S, RPT604001S, RPT602526S, RPT601382S, RPT601636S, RPT602225S, RPT603217S, RPT600356S, RPT601391S, RPT602440S, RPT605041S, RPT604777S, RPT604778S, RPT604947S, RPT605018S, RPT605133S, RPT605134S, RPT605328S, RPT605329S, RPT605615S, RPT605616S, RPT605627S, RPT605824S, RPT606222S, RPT606223S, RPT606842S, RPT606975S, RPT607019S, RPT607126S, RPT607127S, RPT607128S, RPT607129S, RPT607201S, RPT607460S, RPT607461S, RPT600262S, RPT600264S, RPT600292S, RPT600456S, RPT600466S, RPT600726S, RPT601161S, RPT601181S, RPT601706S, RPT601720S, RPT601769S, RPT601806S, RPT601908S, RPT602015S, RPT602236S, RPT602478S, RPT602628S, RPT602651S, RPT602653S, RPT602697S, RPT602764S, RPT602784S, RPT602785S, RPT602847S, RPT603006S, RPT603007S, RPT603071S, RPT603075S, RPT603083S, RPT603173S, RPT603219S, RPT603226S, RPT603227S, RPT603403S, RPT603517S, RPT603984S, RPT604037S, RPT604105S, RPT604295S, RPT604296S, RPT604325S, RPT604426S, RPT604458S, RPT604459S, RPT607626S, RPT607645S, RPT607647S, RPT606044S, RPT605964S, RPT603980S, RPT606228S, RPT605430S, RPT600799S, RPT601869S, RPT604102S, RPT603748S, RPT604741S, RPT605010S, RPT605077S, RPT605275S, RPT605327S, RPT605337S, RPT606633S, RPT606762S, RPT600748S, RPT601664S, RPT601665S, RPT603033S, RPT603450S, RPT603744S, RPT605137S, RPT606601S, RPT602423S, RPT603038S, RPT604474S, RPT601356S, RPT604926S, RPT605414S, RPT604101S, RPT603631S, RPT604391S, RPT605064S, RPT601896S, RPT606555S, RPT603975S, RPT605178S, RPT603557S, RPT603635S, RPT604828S, RPT605278S, RPT604026S, RPT605037S, RPT605180S, RPT605625S, RPT606680S, RPT600222S, RPT602748S, RPT603719S, RPT603737S, RPT605696S, RPT606263S, RPT607347S, RPT606985S, RPT604815S, RPT604816S, RPT606757S, RPT606816S, RPT601029S, RPT601238S, RPT603027S, RPT603254S, RPT603255S, RPT603795S, RPT600507S, RPT606778S, RPT605079S, RPT601203S, RPT603351S, RPT604365S, RPT606649S, RPT606651S, RPT602988S, RPT602454S, RPT602690S, RPT602729S, RPT600687S, RPT600688S, RPT604734S, RPT601916S, RPT604736S, RPT603919S, RPT604250S, RPT606499S, RPT603550S, RPT600824S, RPT605168S, RPT605626S, RPT605746S, RPT606592S, RPT603454S, RPT603679S, RPT604276S, RPT602165S, RPT603757S, RPT601156S, RPT606464S, RPT603398S, RPT605481S, RPT605484S, RPT606663S, RPT606685S, RPT600484S, RPT600536S, RPT600980S, RPT601209S, RPT602062S, RPT602927S, RPT602929S, RPT603143S, RPT601236S, RPT606579S, RPT606662S, RPT607209S, RPT607088S, RPT606758S, RPT606759S, RPT600143S, RPT600408S, RPT602021S, RPT603974S, RPT604024S, RPT605363S, RPT605769S, RPT605817S, RPT605832S, RPT606813S, RPT600834S, RPT601165S, RPT601993S, RPT602010S, RPT602268S, RPT602283S, RPT603338S, RPT603347S, RPT603348S, RPT603354S, RPT604062S, RPT604419S, RPT604788S, RPT601390S, RPT607524S, RPT604845S, RPT604698S, RPT602259S, RPT602795S, RPT605325S, RPT601149S, RPT603862S, RPT603929S, RPT604442S, RPT606239S, RPT605496S, RPT600490S, RPT602095S, RPT604077S, RPT604859S, RPT605487S, RPT601122S, RPT601424S, RPT602856S, RPT603091S, RPT605884S, RPT603256S, RPT605684S, RPT604554S, RPT604110S, RPT604343S, RPT604801S, RPT600068S, RPT601036S, RPT603070S, RPT603201S, RPT605083S, RPT605804S, RPT606078S, RPT606481S, RPT600976S, RPT604258S, RPT606096S, RPT602572S, RPT602777S, RPT604331S, RPT602787S, RPT605549S, RPT605060S, RPT600441S, RPT601133S, RPT606619S, RPT602908S, RPT601644S, RPT604329S, RPT605728S, RPT606031S, RPT603136S, RPT601844S, RPT605311S, RPT601609S, RPT602508S, RPT603431S, RPT604882S, RPT600974S, RPT604961S, RPT605360S, RPT600180S, RPT601948S, RPT603046S, RPT604321S, RPT606186S, RPT600802S, RPT600920S, RPT602428S, RPT604923S, RPT604333S, RPT601583S, RPT602495S, RPT604083S, RPT602043S, RPT603733S, RPT600500S, RPT604038S, RPT605794S, RPT603160S, RPT605464S, RPT604239S, RPT600742S, RPT602522S, RPT602597S, RPT606013S, RPT606301S, RPT601171S, RPT602925S, RPT607336S, RPT600942S, RPT604803S, RPT600987S, RPT601222S, RPT604674S, RPT604769S, RPT605779S, RPT606134S, RPT606525S, RPT606526S, RPT603867S, RPT603950S, RPT604028S, RPT606752S, RPT606780S, RPT603150S, RPT605129S, RPT605576S, RPT601294S, RPT601860S, RPT602267S, RPT602721S, RPT601771S, RPT607160S, RPT606505S, RPT601694S, RPT602145S, RPT602663S, RPT603761S, RPT604403S, RPT606922S, RPT604155S, RPT605961S, RPT605969S, RPT600727S, RPT601007S, RPT607020S, RPT601740S, RPT601397S, RPT600239S, RPT600743S, RPT603439S, RPT606205S, RPT602162S, RPT604829S, RPT601306S, RPT602366S, RPT603830S, RPT606453S, RPT602433S, RPT607029S, RPT600925S, RPT601653S, RPT604688S, RPT605250S, RPT605648S, RPT606550S, RPT600056S, RPT600556S, RPT600644S, RPT601009S, RPT604268S, RPT607566S, RPT601973S, RPT601979S, RPT605061S, RPT605793S, RPT605910S, RPT600184S, RPT600739S, RPT601547S, RPT601960S, RPT602580S, RPT602611S, RPT603314S, RPT603328S, RPT604162S, RPT604551S, RPT606029S, RPT606030S, RPT604569S, RPT604850S, RPT602821S, RPT602866S, RPT603012S, RPT603182S, RPT603279S, RPT603442S, RPT604095S, RPT604121S, RPT604298S, RPT605511S, RPT605726S, RPT606054S, RPT606125S, RPT606264S, RPT607021S, RPT602615S, RPT605917S, RPT600320S, RPT600322S, RPT600323S, RPT600324S, RPT600325S, RPT600668S, RPT601099S, RPT604785S, RPT607108S, RPT602568S, RPT602809S, RPT604544S, RPT606015S, RPT606942S, RPT607217S, RPT600010S, RPT602378S, RPT602171S, RPT602377S, RPT600632S, RPT604952S, RPT606041S, RPT606323S, RPT605781S, RPT600778S, RPT601452S, RPT602242S, RPT603598S, RPT606210S, RPT602523S, RPT605210S, RPT605211S, RPT605324S, RPT606122S, RPT606858S, RPT600804S, RPT603024S, RPT603612S, RPT603613S, RPT606914S, RPT601943S, RPT605531S, RPT606569S, RPT604600S, RPT606957S, RPT602605S, RPT602221S, RPT606227S, RPT606720S, RPT601941S, RPT602243S, RPT603668S, RPT606710S, RPT603434S, RPT603568S, RPT603958S, RPT604946S, RPT605299S, RPT606200S, RPT606654S, RPT606991S, RPT607237S, RPT600285S, RPT600294S, RPT600686S, RPT602195S, RPT602661S, RPT602682S, RPT603169S, RPT603649S, RPT603943S, RPT601396S, RPT600720S, RPT600983S, RPT601157S, RPT602375S, RPT600261S, RPT600550S, RPT600599S, RPT601061S, RPT601276S, RPT601302S, RPT600565S, RPT600663S, RPT604747S, RPT606184S, RPT606187S, RPT606209S, RPT606419S, RPT600330S, RPT600332S, RPT600334S, RPT600379S, RPT600511S, RPT600672S, RPT601333S, RPT601848S, RPT601942S, RPT602404S, RPT602405S, RPT606183S, RPT606207S, RPT606208S, RPT606411S, RPT606413S, RPT606414S, RPT606415S, RPT606417S, RPT606420S, RPT606422S, RPT606424S, RPT606425S, RPT606426S, RPT607242S, RPT607243S, RPT607244S, RPT607245S, RPT607246S, RPT605570S, RPT605700S, RPT600951S, RPT600952S, RPT604006S, RPT600583S, RPT602229S, RPT603579S, RPT606760S, RPT600990S, RPT601415S, RPT601499S, RPT601629S, RPT602286S, RPT600391S, RPT600562S, RPT607079S, RPT604689S, RPT605243S, RPT606576S, RPT600999S, RPT601000S, RPT601462S, RPT601784S, RPT601991S, RPT602109S, RPT602389S, RPT602560S, RPT602701S, RPT602506S, RPT603515S, RPT604744S, RPT602566S, RPT603577S, RPT605821S, RPT601164S, RPT602297S, RPT602684S, RPT603363S, RPT605598S, RPT604585S, RPT604587S, RPT604588S, RPT604589S, RPT604590S, RPT604591S, RPT604593S, RPT604636S, RPT605443S, RPT605712S, RPT602136S, RPT602635S, RPT602818S, RPT603486S, RPT603498S, RPT604870S, RPT605813S, RPT606167S, RPT602556S, RPT604719S, RPT605270S, RPT604541S, RPT601020S, RPT601922S, RPT600831S, RPT603481S, RPT600144S, RPT601875S, RPT601987S, RPT601989S, RPT605521S, RPT600396S, RPT606790S, RPT600683S, RPT604913S, RPT605715S, RPT605757S, RPT606000S, RPT601279S, RPT601408S, RPT603174S, RPT604492S, RPT600108S, RPT600110S, RPT600111S, RPT600114S, RPT600115S, RPT600116S, RPT600117S, RPT601395S, RPT601399S, RPT601429S, RPT601469S, RPT601473S, RPT601475S, RPT601477S, RPT601529S, RPT601532S, RPT601573S, RPT604818S, RPT600840S, RPT602675S, RPT604668S, RPT603923S, RPT604603S, RPT604844S, RPT605277S, RPT602538S, RPT603537S, RPT604724S, RPT601105S, RPT601426S, RPT603438S, RPT604879S, RPT600905S, RPT602005S, RPT607565S, RPT600194S, RPT600195S, RPT600198S, RPT601436S, RPT601443S, RPT601506S, RPT601508S, RPT601511S, RPT601513S, RPT602325S, RPT602330S, RPT602331S, RPT602332S, RPT602351S, RPT602352S, RPT603026S, RPT600171S, RPT600196S, RPT600199S, RPT600205S, RPT601048S, RPT601343S, RPT601489S, RPT601509S, RPT602160S, RPT602410S, RPT603088S, RPT603535S, RPT604530S, RPT605132S, RPT600204S, RPT600206S, RPT600207S, RPT600214S, RPT600215S, RPT604216S, RPT604218S, RPT604228S, RPT604886S, RPT600203S, RPT600208S, RPT600281S, RPT602085S, RPT602412S, RPT603030S, RPT600386S, RPT601435S, RPT601437S, RPT602539S, RPT602311S, RPT604594S, RPT604597S, RPT604601S, RPT604813S, RPT604822S, RPT604825S, RPT605536S, RPT605538S, RPT605737S, RPT605739S, RPT605742S, RPT606333S, RPT600595S, RPT600596S, RPT601733S, RPT601885S, RPT601886S, RPT601889S, RPT601926S, RPT601935S, RPT601949S, RPT601951S, RPT601952S, RPT601953S, RPT601954S, RPT602435S, RPT602576S, RPT602687S, RPT602900S, RPT603010S, RPT603272S, RPT603390S, RPT603466S, RPT603467S, RPT603468S, RPT603471S, RPT603977S, RPT603979S, RPT604112S, RPT604201S, RPT604202S, RPT604203S, RPT604422S, RPT604423S, RPT605738S, RPT606334S, RPT606335S, RPT607414S, RPT607415S, RPT607416S, RPT606367S, RPT601135S, RPT601139S, RPT601142S, RPT602581S, RPT602743S, RPT602744S, RPT606394S, RPT606395S, RPT600173S, RPT600174S, RPT600176S, RPT600177S, RPT600178S, RPT600179S, RPT600216S, RPT600520S, RPT600522S, RPT600523S, RPT600643S, RPT600693S, RPT600694S, RPT600698S, RPT600701S, RPT600702S, RPT600703S, RPT600940S, RPT601172S, RPT601173S, RPT601217S, RPT601246S, RPT601377S, RPT601427S, RPT601430S, RPT601432S, RPT601438S, RPT601442S, RPT601479S, RPT601483S, RPT601535S, RPT601537S, RPT601591S, RPT601592S, RPT601594S, RPT601595S, RPT601597S, RPT601598S, RPT601603S, RPT601702S, RPT601703S, RPT601829S, RPT601832S, RPT601836S, RPT601928S, RPT601930S, RPT601931S, RPT602102S, RPT602103S, RPT602104S, RPT602106S, RPT602112S, RPT602114S, RPT602116S, RPT602118S, RPT602119S, RPT602120S, RPT602122S, RPT602124S, RPT602417S, RPT605191S, RPT600302S, RPT600443S, RPT602425S, RPT604749S, RPT606286S, RPT600381S, RPT601474S, RPT601581S, RPT603423S, RPT605287S, RPT601329S, RPT604199S, RPT604768S, RPT604180S, RPT600460S, RPT606397S, RPT600248S, RPT607312S, RPT601169S, RPT600746S, RPT602992S, RPT602414S, RPT600305S, RPT604586S, RPT601681S, RPT603461S, RPT607696S, RPT606412S, RPT602877S, RPT606150S, RPT606317S, RPT603213S, RPT604713S, RPT604715S, RPT604906S, RPT604909S, RPT605196S, RPT605298S, RPT605342S, RPT605558S, RPT605677S, RPT600274S, RPT600275S, RPT601292S, RPT601293S, RPT601520S, RPT601521S, RPT601523S, RPT601648S, RPT601888S, RPT601903S, RPT601904S, RPT601967S, RPT602301S, RPT602302S, RPT603021S, RPT603022S, RPT603359S, RPT603366S, RPT603541S, RPT603542S, RPT603543S, RPT603560S, RPT603561S, RPT603899S, RPT603900S, RPT604453S, RPT604454S, RPT605678S, RPT605679S, RPT605975S, RPT605976S, RPT605977S, RPT606321S, RPT606322S, RPT607427S, RPT607428S, RPT607429S, RPT601428S, RPT604679S, RPT604708S, RPT605001S, RPT605504S, RPT605596S, RPT605935S, RPT606201S, RPT606448S, RPT606580S, RPT606687S, RPT600287S, RPT600378S, RPT601262S, RPT601863S, RPT602159S, RPT602364S, RPT603224S, RPT603418S, RPT603496S, RPT604411S, RPT604494S, RPT606449S, RPT607124S, RPT601707S, RPT601767S, RPT601925S, RPT601113S, RPT601939S, RPT602863S, RPT601684S, RPT604274S, RPT601232S, RPT606502S, RPT607378S, RPT606628S, RPT603462S, RPT605174S, RPT605175S, RPT607090S, RPT604234S, RPT604235S, RPT604236S, RPT607089S, RPT605189S, RPT607058S, RPT603188S, RPT605314S, RPT605652S, RPT605653S, RPT605654S, RPT600357S, RPT600360S, RPT600361S, RPT600659S, RPT600876S, RPT600877S, RPT600881S, RPT600882S, RPT600884S, RPT600886S, RPT600887S, RPT600894S, RPT600895S, RPT600896S, RPT600938S, RPT601299S, RPT601369S, RPT601790S, RPT601944S, RPT601965S, RPT601970S, RPT601984S, RPT602087S, RPT602089S, RPT602138S, RPT602139S, RPT602140S, RPT602147S, RPT602312S, RPT602347S, RPT602365S, RPT602390S, RPT604286S, RPT604289S, RPT604292S, RPT605315S, RPT605659S, RPT605668S, RPT606539S, RPT600362S, RPT600611S, RPT600661S, RPT600898S, RPT602316S, RPT602321S, RPT604287S, RPT605669S, RPT605650S, RPT605657S, RPT600893S, RPT600927S, RPT600933S, RPT600935S, RPT601698S, RPT601791S, RPT601966S, RPT602278S, RPT602322S, RPT602334S, RPT602337S, RPT602349S, RPT602368S, RPT605661S, RPT605651S, RPT605655S, RPT606145S, RPT606722S, RPT600358S, RPT600890S, RPT601854S, RPT602272S, RPT602281S, RPT602284S, RPT605662S, RPT601851S, RPT602323S, RPT600278S, RPT600897S, RPT600926S, RPT600929S, RPT601969S, RPT601982S, RPT602358S, RPT606724S, RPT605664S, RPT606343S, RPT600932S, RPT602346S, RPT600879S, RPT602086S, RPT600856S, RPT600891S, RPT601981S, RPT605665S, RPT605666S, RPT605667S, RPT607248S, RPT600276S, RPT600549S, RPT600883S, RPT600888S, RPT601013S, RPT601300S, RPT602314S, RPT602318S, RPT602339S, RPT602371S, RPT604284S, RPT605663S, RPT600889S, RPT601697S, RPT602298S, RPT602319S, RPT605656S, RPT605660S, RPT600885S, RPT601813S, RPT601980S, RPT602137S, RPT602336S, RPT602353S, RPT602391S, RPT604282S, RPT604283S, RPT604288S, RPT604290S, RPT604291S, RPT604360S, RPT605897S, RPT607084S, RPT600930S, RPT601045S, RPT606330S, RPT607137S
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Claria MRI™ CRT-D SureScan™ DTMA1D1Medtronic, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)1 recall7 deaths415 reports
- Claria MRI™ CRT-D SureScan™ DTMA1D1Medtronic, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)1 recall10 deaths269 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92328).
- Z-1704-2023 · Open, ClassifiedICD-DR DDBB1D1 EVERA XT IS1/DF1 US, Model Number DDBB1D1; Implantable Cardioverter Defibrillators
- Z-1705-2023 · Open, ClassifiedICD-DR DDBB1D4 EVERA XT IS1/DF4 US, Model Number DDBB1D4; Implantable Cardioverter Defibrillators
- Z-1709-2023 · Open, ClassifiedICD-DR DDBC3D1 EVERA S IS1/DF1 GLOB, Model Number DDBC3D1; Implantable Cardioverter Defibrillators
- Z-1710-2023 · Open, ClassifiedICD-DR DDBC3D4 EVERA S IS1/DF4 GLOB, Model Number DDBC3D4; Implantable Cardioverter Defibrillators
- Z-1711-2023 · Open, ClassifiedICD-DR DDMB1D1 EVERA MRI XT US DF1, Model Number DDMB1D1; Implantable Cardioverter Defibrillators
- Z-1712-2023 · Open, ClassifiedICD-DR DDMB1D4 EVERA MRI XT IS-1/DF4 US, Model Number DDMB1D4; Implantable Cardioverter Defibrillators
- Z-1715-2023 · Open, ClassifiedICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
- Z-1716-2023 · Open, ClassifiedICD-DR DDMC3D4 EVERA MRI S IS-1/DF4 GLOB, Model Number DDMC3D4; Implantable Cardioverter Defibrillators
- Z-1717-2023 · Open, ClassifiedICD-DR DDMD3D1 PRIMO MRI, Model Number DDMD3D1; Implantable Cardioverter Defibrillators
- Z-1718-2023 · Open, ClassifiedICD-DR DDMD3D4 PRIMO MRI, Model Number DDMD3D4; Implantable Cardioverter Defibrillators
- Z-1721-2023 · Open, ClassifiedICD COBALT XT DR MRI IS1 DF1, Model Number DDPA2D1; Implantable Cardioverter Defibrillators
- Z-1723-2023 · Open, ClassifiedICD COBALT XT DR MRI IS1 DF4, Model Number DDPA2D4; Implantable Cardioverter Defibrillators
- Z-1725-2023 · Open, ClassifiedICD COBALT DR MRI IS1 DF1, Model Number DDPB3D1; Implantable Cardioverter Defibrillators
- Z-1726-2023 · Open, ClassifiedICD COBALT DR MRI IS1 DF4, Model Number DDPB3D4; Implantable Cardioverter Defibrillators
- Z-1727-2023 · Open, ClassifiedICD CROME DR MRI IS1 DF1, Model Number DDPC3D1; Implantable Cardioverter Defibrillators
- Z-1728-2023 · Open, ClassifiedICD CROME DR MRI IS1 DF4, Model Number DDPC3D4; Implantable Cardioverter Defibrillators
- Z-1729-2023 · Open, ClassifiedCRT-D DTBA1D1 VIVA XT IS1/DF1 US, Model Number DTBA1D1; Implantable Cardioverter Defibrillators
- Z-1731-2023 · Open, ClassifiedCRT-D DTBA1D4 VIVA XT IS1/DF4 US, Model Number DTBA1D4; Implantable Cardioverter Defibrillators
- Z-1732-2023 · Open, ClassifiedCRT-D DTBA1Q1 VIVA QUAD XT IS4/DF1 US, Model Number DTBA1Q1; Implantable Cardioverter Defibrillators
- Z-1733-2023 · Open, ClassifiedCRT-D DTBA1QQ VIVA QUAD XT IS4/DF4 US, Model Number DTBA1QQ; Implantable Cardioverter Defibrillators
- Z-1741-2023 · Open, ClassifiedCRT-D DTBB1D1 VIVA S IS1/DF1 US, Model Number DTBB1D1; Implantable Cardioverter Defibrillators
- Z-1742-2023 · Open, ClassifiedCRT-D DTBB1D4 VIVA S IS1/DF4 US, Model Number DTBB1D4; Implantable Cardioverter Defibrillators
- Z-1743-2023 · Open, ClassifiedCRT-D DTBB1Q1 VIVA QUAD S IS4/DF1 US, Model Number DTBB1Q1; Implantable Cardioverter Defibrillators
- Z-1744-2023 · Open, ClassifiedCRT-D DTBB1QQ VIVA QUAD S IS4/DF4 US, Model Number DTBB1QQ; Implantable Cardioverter Defibrillators
- Z-1754-2023 · Open, ClassifiedCRTD DTMA1D4 CLARIA MRI US DF4, Model Number DTMA1D4; Implantable Cardioverter Defibrillators
- Z-1755-2023 · Open, ClassifiedCRTD DTMA1Q1 CLARIA MRI QUAD US DF1, Model Number DTMA1Q1; Implantable Cardioverter Defibrillators
- Z-1756-2023 · Open, ClassifiedCRTD DTMA1QQ CLARIA MRI QUAD US DF4, Model Number DTMA1QQ; Implantable Cardioverter Defibrillators
- Z-1761-2023 · Open, ClassifiedCRTD DTMB1D1 AMPLIA MRI US DF1, Model Number DTMB1D1; Implantable Cardioverter Defibrillators
- Z-1762-2023 · Open, ClassifiedCRTD DTMB1D4 AMPLIA MRI US DF4, Model Number DTMB1D4; Implantable Cardioverter Defibrillators
- Z-1763-2023 · Open, ClassifiedCRTD DTMB1Q1 AMPLIA MRI QUAD US DF1, Model Number DTMB1Q1; Implantable Cardioverter Defibrillators
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.