Skip to content

n'-Hole Post which is a component of the Revolution… recallZ-1824-2023 · New Standard Device Inc · initiated Jul 23, 2019

Class II Open, Classified All recalls posted in June 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from New Standard Device Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The threads on posts stripping out when tensioned during application process.

FDA root cause category: Under Investigation by firm

What the company is doing

On 07/23/2019, the firm emailed a "FIELD SAFETY NOTICE" Letter to customers informing them that threads on posts were stripping out when tensioned during application process. Customers were instructed to: 1. Remove remaining POSTS from their inventory and return to Metalogix Inc. at 4848 Research Drive, San Antonio, TX 78240 Please label return package "ATTENTION: QA Dept" Metalogix Fed-Ex Account# 845 986 983 Metalogix will expedite new product back to customers For further information or questions, contact: Rachel Grimes, Quality Assurance Manager at 833-659-2019, EXT 6 [email protected]

Recall details

Recall number
Z-1824-2023
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jul 23, 2019
Posted by the FDA
Jun 9, 2023
Recalling firm
New Standard Device Inc
Quantity
222 units
Product code
KTT
FDA submissions
K181630
Distribution
U.S. Nationwide distribution in the states of AZ, CT, OH, TX, VA.

Product description

'n'-Hole Post which is a component of the Revolution External Fixation System Catalog Numbers: 100801, 100802, 100803, 100804, 100805

Code information (models, lots, serial numbers)

Catalog Numbers / Lot Numbers: 100801 / A29318R 100802 / A29318S 100803 / A29318T 100804 / A29318U 100805 / A29318V UDI Codes: Not provided/None

Other devices under the same FDA submission

6 devices cleared under K181630 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.